直接前入路全髋关节置换术后撞击风险的纵向评估

Longitudinal Assessment of Impingement Risk Following Total Hip Arthroplasty Through the Direct Anterior Approach.
PubMed 文献 · J Arthroplasty · 2025-09-17

本研究对24例直接前入路全髋关节置换术参与者进行放射立体测量分析,在术后5年内多个时间点计算股骨柄与聚乙烯内衬间最近距离以评估组件间撞击风险。结果显示仰卧位距离保持稳定,而站立位平均距离显著更低(P < 0.001);32 mm股骨头距离低于36 mm(P < 0.0001),带offset柄的颈-内衬距离显著更大(P值0.016至0.046),带领与无领柄间无差异。结论提示站立位功能性负重评估对评估术后撞击风险至关重要。

要点

  • 仰卧位撞击风险保持稳定,站立位平均距离显著更低(P < 0.001)。
  • 32 mm股骨头距离低于36 mm股骨头(P < 0.0001)。
  • 带offset柄的颈-内衬距离显著更大(P值0.016至0.046),带领与无领柄间无差异。

核心结论

  • 组件间撞击风险在仰卧位保持稳定,但在站立位增加,提示功能性负重评估对评估术后撞击风险至关重要。

研究设计与样本

  • 对24例直接前入路全髋关节置换术参与者,在手术当天、2周、4周、6周、3个月、6个月、1年、2年和5年多个时间点进行放射立体测量分析,使用定制计算工具基于三维姿态测量股骨柄与聚乙烯内衬间最近距离,以1 mm为阈值分类撞击。

主要发现

  • 仰卧位距离保持稳定,站立位测量显示平均距离显著更低(P < 0.001)。
  • 32 mm股骨头距离低于36 mm股骨头(P < 0.0001)。
  • 带offset柄的颈-内衬距离显著更大(P值0.016至0.046)。
  • 带领与无领柄之间无差异。

关键数值与推荐

  • 以1 mm为阈值分类撞击,距离接近1 mm被视为更高撞击风险。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Component-on-component impingement following total hip arthroplasty (THA) can lead to postoperative complications, but its prevalence in nonfailed hips remains unclear due to dynamic factors and limited measurement methodologies. Mechanisms of impingement are mostly thought to be due to component malalignment, such as acetabular cups placed outside of the Lewinnek safe zones, or component sizing, such as the use of large head sizes. However, it is known that early migration of implant components does occur following THA, which may potentially impact postoperative impingement risk. The objective of this study was therefore to evaluate component-on-component impingement risk in patients who have undergone THA over a 5-year postoperative timeline and to evaluate positional impact by comparing impingement risk in the supine and standing positions. METHODS: There were 24 anterior-approach THA participants who underwent radiostereometric analysis at multiple time points: on the day of surgery, at two weeks, four weeks, six weeks, three months, six months, one year, two years, and five years. Impingement risk was calculated as per a custom computational tool, which measures the closest distance between the femoral stem and the polyethylene liner based on the 3-dimensional poses extracted from radiostereometric analysis. A threshold of one mm was defined to classify impingement, with distances approaching one mm being considered at higher impingement risk. RESULTS: Supine distances remained stable, though standing measurements showed significantly lower mean distances (P < 0.001). Smaller femoral heads (32 mm) had lower distances than larger heads (36 mm; P < 0.0001). Offset stems had significantly greater (P values = 0.016 to 0.046) neck-liner distances. There were no differences observed between collared and collarless stems. CONCLUSIONS: Our findings indicate that component-on-component impingement risk remains stable in the supine position, but increases in the standing position, suggesting that functional weight-bearing evaluations are critical for assessing postoperative impingement risk.