全髋关节置换术中第三代与第二代中度交联聚乙烯内衬的体内磨损分析
本研究比较了初次全髋关节置换术中第二代与第三代中度交联聚乙烯内衬的体内磨损及骨溶解风险。共分析58个第三代和51个第二代内衬,第二代内衬中位线性磨损为0.068 mm/年而第三代为零,但内衬类型与磨损>0.1或>0.05 mm/年无关联;第二代内衬、股骨头尺寸增大和髋臼外展角增大与磨损增加相关。
要点
- 第三代内衬中位线性磨损为零,第二代内衬为0.068 mm/年。
- 20个(39%)第二代和13个(22%)第三代内衬磨损>0.1 mm/年。
- 第二代内衬中3个出现影像学骨溶解,第三代内衬中无骨溶解。
核心结论
- 第三代内衬在至少3年随访时表现出比第二代内衬更低的磨损,但与更低的骨溶解率或超过0.1或0.05 mm/年的磨损率无关。
- 两种交联聚乙烯内衬中磨损大于0.1 mm/年的比例均出乎意料地高。
- 使用当代交联聚乙烯时,部件位置似乎比内衬材料更重要。
研究设计与样本
- 回顾性分析接受初次全髋关节置换术并植入第二代或第三代交联聚乙烯内衬的患者,排除影像学随访不足3年者,共分析58个第三代和51个第二代内衬。
- 聚乙烯磨损通过骨盆正位X线片的半自动边缘检测软件分析估算,并手动检查X线片评估骨溶解。
- 采用多变量分析估算和比较磨损率,计算比值比以评估发生磨损>0.1或>0.05 mm/年的相对几率。
主要发现
- 第二代内衬中位线性磨损为0.068 mm/年,第三代内衬中位线性磨损为零mm/年。
- 20个(39%)第二代内衬和13个(22%)第三代内衬磨损>0.1 mm/年。
- 影像学观察到3个第二代内衬出现骨溶解,第三代内衬中无骨溶解。
- 多变量分析发现第二代内衬、股骨头尺寸增大和髋臼外展角增大与磨损增加相关。
- 内衬类型与磨损>0.1或>0.05 mm/年无关联;倾斜角增大在所有分析中均与更高磨损水平一致相关。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Prior studies have reported low rates of wear associated with crosslinked polyethylene (XLPE) liners used in total hip arthroplasty, but no studies have compared the in vivo performance of second-generation and third-generation XLPE liners. The purpose of this study was to compare the in vivo wear profile and corresponding odds of developing osteolysis associated with second-generation and third-generation liners. METHODS: Patients receiving a second-generation or third-generation XLPE liner during primary total hip arthroplasty were reviewed. Patients who had < three years of radiographic follow-up were excluded. A total of 58 third-generation and 51 second-generation liners were analyzed. Polyethylene wear was estimated using a semiautomated, edge-detection software analysis of anterioposterior pelvis radiographs. Radiographs were examined manually for osteolysis. Multivariable analysis was performed to estimate and compare the rates of wear. Odds ratios were calculated to estimate the relative odds of developing wear > 0.1 or > 0.05 mm/y. RESULTS: The median linear wear of second-generation liners was 0.068 mm/year, and that of third-generation liners was zero mm/year. There were 20 (39%) second-generation and 13 (22%) third-generation liners that demonstrated wear > 0.1 mm/year. Osteolysis was radiographically observed in three second-generation liners and no third-generation liners. Multivariable analysis found second-generation liners, increased femoral head size, and increased acetabular abduction angle to be associated with increased wear. Liner type was not associated with wear > 0.1 or > 0.05 mm/year. Increased inclination angle was consistently associated with higher levels of wear across all analyses. CONCLUSIONS: Third-generation liners exhibited lower wear than second-generation liners at a minimum 3-year follow-up, but were not associated with lower rates of osteolysis or wear exceeding 0.1 or 0.05 mm/year. Both XLPEs had surprisingly high proportions of liners with wear greater than > 0.1 mm/year. Component positioning appears to be more important than liner material when using contemporary XLPEs.
引用
- Blackburn CW, Jarrett CM, Sattar A, et al. In Vivo Wear Analysis of Third- versus Second-Generation Moderately Crosslinked Polyethylene Liners in Total Hip Arthroplasty. J Arthroplasty. 2025. PMID: 40840658. doi:10.1016/j.arth.2025.08.032