缝隙腐蚀使全膝关节置换术后的干骺端袖套降解:一项取回研究
本研究分析17个股骨和15个胫骨植入物-袖套组件,发现股骨适配器82%、股骨袖套100%、胫骨托73%、胫骨袖套86%存在腐蚀;股骨钴铬钼适配器轴向腐蚀较胫骨托更重(P=0.04);主要损伤模式为点蚀、蚀刻和氧化物碎屑。
要点
- 股骨袖套腐蚀率100%,胫骨袖套86%,股骨适配器82%,胫骨托73%。
- 股骨钴铬钼适配器轴向腐蚀较胫骨托更重(P=0.04)。
- 股骨适配器主要损伤模式为点蚀、蚀刻和氧化物碎屑。
核心结论
- 机械辅助缝隙腐蚀促使干骺端植入物降解。
- 对干骺端袖套且不明原因疼痛或疑似金属释放的患者,在严格排除其他诊断后可考虑测量金属浓度。
研究设计与样本
- 取回研究,分析17个股骨和15个胫骨植入物-袖套组件。
- 采用Goldberg评分分类腐蚀,以钴铬钼植入物轴向腐蚀(ΔCmax)量化腐蚀。
- 对中度和重度腐蚀设备行扫描电子显微镜和能量色散X射线光谱成像。
- 使用非参数统计评估钴铬钼胫骨和股骨植入物腐蚀差异。
主要发现
- 股骨适配器腐蚀率82%,股骨袖套100%,胫骨托73%,胫骨袖套86%。
- 钴铬钼股骨适配器轴向腐蚀大于钴铬钼胫骨托(P = 0.04)。
- 股骨适配器主要损伤模式包括点蚀、蚀刻和氧化物碎屑。
- 钛合金股骨袖套可见抛光效应,显露底层微观结构。
临床怎么用
- 对干骺端袖套且不明原因疼痛或疑似金属释放的患者,在严格排除其他诊断后可考虑测量金属浓度。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Surgeons use metaphyseal sleeves in revision total knee arthroplasty (TKA) to compensate for bone loss and improve stability. However, a clinical case series documented corrosion on these devices and associated the subsequent metal release with adverse local tissue reactions. In this study, we investigated corrosion at the mixed-alloy, metaphyseal sleeve-implant interface. We asked the following: (1) How common is corrosion? (2) Does more damage occur at the femoral cobalt chrome alloy (CoCrMo) taper? (3) What damage modes occur? METHODS: We analyzed 17 femoral and 15 tibial implant-sleeve components following revision surgery. First, corrosion was classified using the Goldberg score and quantified using the axial corrosion (ΔCmax) on CoCrMo implants. Then, moderately and severely corroded devices were imaged using scanning electron microscopy and energy dispersive X-ray spectroscopy. Differences in corrosion on CoCrMo tibial and femoral implants were assessed using nonparametric statistics. RESULTS: We observed corrosion on 82% of femoral adapters, 100% of femoral sleeves, 73% of tibial trays, and 86% of tibial sleeves. On CoCrMo femoral adapters, we measured more axial corrosion than on CoCrMo tibial trays (P = 0.04). The predominant damage modes on femoral adapters included pitting, etching, and oxide debris. On titanium alloy femoral sleeves, we documented a burnishing effect, revealing the underlying microstructure. CONCLUSIONS: Our findings indicate that mechanically assisted crevice corrosion promotes metaphyseal implant degradation. Should surgeons encounter patients who have metaphyseal sleeves and unexplained pain or suspected metal release, consider measuring metal concentrations after rigorous rule out of competing diagnoses.
引用
- Aslani S, Kurtz MA, Pelt CE, et al. Crevice Corrosion Degrades Metaphyseal Sleeves Following Total Knee Arthroplasty: A Retrieval Study. J Arthroplasty. 2025. PMID: 40480330. doi:10.1016/j.arth.2025.05.118
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。