文献速递骨质疏松与代谢性骨病

在乳腺X线摄影环境中通过数字腕部断层合成进行骨质疏松筛查

Osteoporosis screening in the mammography setting via digital wrist tomosynthesis.
PubMed 文献 · Osteoporos Int · 2025-05-09

在150名女性中,使用3D乳腺摄影系统进行数字腕部断层合成(DWT)测量骨密度,并与髋部、脊柱和前臂双能X线吸收测定法(DXA)比较。DWT骨密度与DXA超远端桡骨骨密度强相关(R²最高0.814),诊断骨质疏松的AUC最高0.978,诊断骨量减少的AUC最高0.938,体内精密度误差低(0.91-2.3%)。结论认为DWT可在乳腺摄影环境中实施,诊断准确性与DXA相似,有潜力提高骨质疏松筛查的依从性。

要点

  • DWT骨密度与DXA超远端桡骨骨密度强相关,R²最高0.814。
  • DWT诊断骨质疏松的AUC最高0.978,诊断骨量减少的AUC最高0.938。
  • DWT体内精密度误差低,为0.91-2.3%。

核心结论

  • DWT可在乳腺摄影环境中实施,诊断准确性与DXA相似,有潜力提高骨质疏松筛查建议的依从性,并在高度可及的乳腺筛查环境中提供便捷的骨密度测量手段。

研究设计与样本

  • 在150名女性中,使用3D乳腺摄影系统进行数字腕部断层合成(DWT)成像,并从3D断层合成和合成2D图像测量骨密度;同时进行髋部、脊柱和前臂部位的标准DXA骨密度测量。

主要发现

  • DWT骨密度与DXA衍生的超远端桡骨骨密度强相关,R²最高0.814。
  • DWT通过超远端T值判别骨质疏松的AUC最高0.978,判别骨量减少的AUC最高0.938,体内精密度误差低(0.91-2.3%)。
  • 由3D DWT得出的骨密度在诊断由髋部和脊柱DXA确定的骨量减少(AUC最高0.916)和骨质疏松(AUC最高0.946)方面与前臂DXA骨密度表现相当。

关键数值与推荐

  • DWT骨密度与DXA超远端桡骨骨密度相关R²最高0.814。
  • DWT判别骨质疏松AUC最高0.978,判别骨量减少AUC最高0.938。
  • DWT体内精密度误差0.91-2.3%。
  • 3D DWT骨密度诊断骨量减少AUC最高0.916,诊断骨质疏松AUC最高0.946。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

UNLABELLED: Adherence to osteoporosis screening guidelines could be considerably higher if offered at the time of routine mammography using the same imaging modality. We found that forearm density measurements using a breast imaging system provides density information with excellent diagnostic capability for osteoporosis and osteopenia status determined by hip and spine DXA. PURPOSE: Adherence to osteoporosis screening guidelines via bone mineral density (BMD) measurements with dual-energy x-ray absorptiometry (DXA) is low. Since adherence to breast cancer screening is quite high, it was suggested that the rate of osteoporosis screening can be improved if wrist imaging were performed at the time of breast screening using the very same equipment. METHODS: Digital wrist tomosynthesis (DWT) imaging was performed in 150 women using a 3D mammography system and BMD was measured from both 3D tomosynthesis and synthesized 2D images. In addition, standard DXA based BMD measurements were performed at the hip, spine, and forearm sites. We examined the extent to which DWT-derived ultradistal radius BMD correlates with DXA based BMD measurements, evaluated DWT measurement precision errors, and determined the accuracy of DWT in diagnosing low bone mass and osteoporosis in vivo. RESULTS: DWT BMD strongly correlated with DXA-derived ultradistal radius BMD (R2 up to 0.814) and discriminated osteoporosis (AUC up to 0.978) and osteopenia (AUC up to 0.938) by ultradistal T-score with low in vivo precision errors (0.91-2.3%). BMD derived from 3D DWT BMD performed comparably to forearm DXA BMD in the diagnosis of osteopenia (AUC up to 0.916) and osteoporosis (AUC up to 0.946) determined by hip and spine DXA. CONCLUSIONS: DWT can be readily implemented in mammography settings with similar diagnostic accuracy to DXA, has the potential to increase adherence to osteoporosis screening recommendations, and offers a convenient means to measure bone density within the highly accessible breast screening environment.

来源:Osteoporos Int,查看原文。版权归原作者所有。

引用

  1. Oravec D, Yadav RN, Cushman T, et al. Osteoporosis screening in the mammography setting via digital wrist tomosynthesis. Osteoporos Int. 2025. PMID: 40341965. doi:10.1007/s00198-025-07516-y

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。