全髋关节置换术中高交联聚乙烯与常规聚乙烯比较的随机前瞻性双盲试验:至少20年随访结果

A randomized prospective double-blind trial comparing highly cross-linked with conventional polyethylene in total hip arthroplasty : results at a minimum follow-up of 20 years.
PubMed 文献 · Bone Joint J · 2025-05-01

这项前瞻性双盲随机对照试验将122例患者随机分配至常规聚乙烯(Enduron)或高交联聚乙烯(Marathon)内衬组,其他假体组件相同。至少20年随访时,高交联聚乙烯组磨损率显著低于常规聚乙烯组,且翻修率更低。

要点

  • 至少20年随访时,常规聚乙烯组磨损率为0.181 mm/年(95%CI 0.175至0.188),高交联聚乙烯组为0.034 mm/年(95%CI 0.028至0.039)。
  • 37例需翻修的患者中,常规聚乙烯内衬组28例,高交联聚乙烯内衬组9例,常规聚乙烯组翻修率是高交联聚乙烯组的三倍。
  • 这是随访时间最长的随机对照试验,显示高交联聚乙烯作为初次全髋关节置换术的承重面可显著改善临床和影像学结果。

核心结论

  • 这是随访时间最长的随机对照试验,显示高交联聚乙烯作为初次全髋关节置换术的承重面可显著改善临床和影像学结果。

研究设计与样本

  • 前瞻性双盲随机对照试验,共122例患者随机接受常规聚乙烯内衬(Enduron)或高交联聚乙烯内衬(Marathon),其他组件相同:Duraloc 300金属壳、28 mm直径钴铬股骨头、骨水泥型Charnley Elite股骨柄。

主要发现

  • 至少20年随访时,47例患者死亡(其中5例曾接受翻修),另32例存活但已接受翻修,43例存活且未翻修(Enduron组15例,Marathon组28例)。
  • 无患者失访,但2例因痴呆在最终随访时无法获得X线片,41例(Enduron组15例,Marathon组26例)可用于聚乙烯磨损测量。
  • 磨合期后,Enduron组磨损率为0.181 mm/年(95%CI 0.175至0.188),Marathon组为0.034 mm/年(95%CI 0.028至0.039)。
  • 20年随访时,37例患者需翻修,常规聚乙烯内衬组翻修率(28/37)是高交联聚乙烯内衬组(9/37)的三倍。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIMS: Highly cross-linked polyethylene (XLPE) was developed in an attempt to address the problems of wear and osteolysis causing failure in patients who undergo total hip arthroplasty (THA). The aim of this study was to compare the rates of PE wear and revision, and the clinical outcomes of conventional polyethylene (CPE) and XLPE, in a prospective double-blind randomized controlled trial (RCT) with a minimum follow-up of 20 years. METHODS: A total of 122 patients were randomized to be treated with either a CPE (Enduron) or an XLPE liner (Marathon; both DePuy Synthes, USA). Other than the liners, identical components were used: a Duraloc 300 metal shell, a 28 mm diameter cobalt-chromium femoral head, and a cemented Charnley Elite femoral component. All patients were followed with anteroposterior (AP) and lateral radiographs at two days, six weeks, three and six months, and one, two, three, four, five, ten, and 20 years. PE wear was measured using PolyMig, which has a phantom validated accuracy of approximately 0.09 mm. The outcomes at ten years' follow-up in these patients have previously been reported. RESULTS: At a minimum follow-up of 20 years, 47 patients had died, five of whom had undergone a revision. Another 32 patients had undergone a revision and were alive, leaving 43 patients who had not undergone a revision and were alive (15 in the Enduron group and 28 in the Marathon group). No patients were lost to follow-up, but radiographs could not be obtained at final follow-up in two due to dementia, leaving 41 patients (15 in the Enduron group and 26 in the Marathon group) available for the measurement of PE wear. After the bedding-in period, the rate of wear was 0.181 mm/yr (95% CI 0.175 to 0.188) in the Enduron group and 0.034 mm/yr (95% CI 0.028 to 0.039) in the Marathon group. At 20-year follow-up, 37 patients had required revision. Those with CPE liners had a revision rate three times higher (28/37) than those with XLPE liners (9/37). CONCLUSION: This is the longest-term RCT showing substantially improved clinical and radiological results when XLPE is used as the bearing surface in patients who undergo primary THA.