机器人技术与手工技术在保留交叉韧带及内侧稳定型假体全膝关节置换术中实现矢状面目标的影响
本研究基于U-Net深度学习模型在侧位膝关节X线片上计算股骨后髁偏移(PCO)与胫骨后倾(TS),对比280例采用同一内侧稳定型假体的机器人辅助全膝关节置换术(raTKA,n=148)与手工全膝关节置换术(mTKA,n=132)实现术者术前矢状面目标的情况。结果显示raTKA术后TS与目标值的平均差异小于mTKA(0.3° vs 1.3°,P=0.03),但mTKA在恢复原生PCO方面更优(-1.7 mm vs 3.3 mm,P<0.001);raTKA术后数据变异性更小,提示精确性更高。作者认为这些差异是否具有临床相关性尚需未来研究确定。
要点
- raTKA术后TS与目标值平均差异0.3°,小于mTKA的1.3°(P=0.03)。
- mTKA恢复原生PCO优于raTKA(-1.7 mm vs 3.3 mm,P<0.001)。
- raTKA术后数据变异性更小,提示精确性更高。
核心结论
- 作者开发了基于U-Net架构的深度学习算法在侧位膝关节X线片上计算PCO与TS。
- raTKA与mTKA在实现矢状面目标上存在显著差异,raTKA比mTKA更精确。
- 作者认为这些差异是否具有临床相关性尚需未来研究确定。
研究设计与样本
- 基于U-Net架构训练深度学习模型,用于在侧位膝关节X线片上计算PCO与TS。
- 将该模型应用于连续队列280例患者,其中mTKA组132例、raTKA组148例,均在同一三级转诊中心采用同一内侧稳定型膝关节假体。
- mTKA采用测量截骨技术,raTKA采用卡尺运动学对线或间隙平衡技术。
主要发现
- 算法与人工测量之间的平均绝对误差为TS 1.3±1.6°、PCO 1.7±1.4 mm,分别小于人工标注者之间的差异(2.0±1.9°和2.2±2.6 mm)。
- 术后TS与目标值的平均差异raTKA小于mTKA(0.3° vs 1.3°,P=0.03)。
- 恢复原生PCO方面mTKA优于raTKA(-1.7 mm vs 3.3 mm,P<0.001)。
- 尽管raTKA对线理念更多样且非骨水泥固定比例更高,其术后数据变异性更小,提示精确性更高。
局限性
- 作者结论指出未来需要研究以确定这些差异是否具有临床相关性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Most of the focus regarding total knee arthroplasty (TKA) implant positioning and alignment has been centered on the coronal plane. Posterior condylar offset (PCO) and tibial slope (TS) are sagittal parameters that are measured on radiographs, managed intraoperatively, and are crucial to a stable TKA. We sought to compare whether robotic-assisted TKA (raTKA) versus manual TKA (mTKA) are different with regard to achieving a surgeon's preoperative sagittal targets. METHODS: We trained a deep learning model based on a U-Net architecture that calculates PCO and TS on lateral knee radiographs. We deployed this model on a consecutive cohort of 280 patients who underwent either mTKA (n = 132) or raTKA (n = 148), with the same medial stabilized knee implant at a tertiary referral center. Measured resection was the technique for mTKA and either calipered kinematic alignment or gap balancing for raTKA. RESULTS: Mean absolute error between the algorithm and human measurements was 1.3 ± 1.6° for TS and 1.7 ± 1.4 mm for PCO, which was less than the difference between the human annotators (2.0 ± 1.9° and 2.2 ± 2.6 mm, respectively). Mean difference between goal and postoperative TS was less in raTKA than mTKA (0.3 versus 1.3°; P = 0.03). However, the opposite was observed regarding restoration of native PCO, favoring mTKA (-1.7 versus 3.3 mm; P < 0.001). Overall, despite increased diversity in alignment philosophies and proportion of cementless fixation, there was less variability in raTKA postoperative data, suggesting increased precision. CONCLUSIONS: We developed a deep learning algorithm to calculate PCO and TS on lateral knee radiographs. We observed significant differences between raTKA and mTKA in achieving sagittal plane targets, with raTKA being more precise than mTKA. Future studies are warranted to determine whether these differences are clinically relevant.
引用
- Girod MM, Mulford KL, Kaji ES, et al. Influence of Robotic Versus Manual Technology for Achieving Sagittal Targets in Total Knee Arthroplasty Using a Cruciate-Retaining and Medial-Stabilized Implant. J Arthroplasty. 2025. PMID: 40398583. doi:10.1016/j.arth.2025.05.052