带花键锥形开槽柄的设计是否影响其初期稳定性?一项生物力学尸体研究
研究纳入6对尸体股骨,制作扩展转子截骨并钢缆固定后,分别植入两种商用带花键锥形开槽柄,激光扫描测量柄骨接触,并施加1000次循环生物力学加载测试柄骨微动。两种柄的平均下沉量分别为313±138 μm和181±113 μm(P > 0.05),最大微动均出现在第一循环,前侧柄骨接触多于后侧,平均接触面积分别为1546±533 mm2和1525±587 mm2(P > 0.05)。结论认为不同几何设计未导致柄骨接触范围和分布的显著变化,两种柄在测试载荷下初期稳定性相似。
要点
- 两种带花键锥形开槽柄的平均下沉量分别为313±138 μm和181±113 μm,差异无统计学意义(P > 0.05)。
- 两种柄的最大微动均出现在第一循环,后续循环保持恒定。
- 两种柄的前侧柄骨接触均多于后侧,平均接触面积分别为1546±533 mm2和1525±587 mm2(P > 0.05)。
核心结论
- 不同几何设计未导致柄骨接触范围和分布的显著变化,两种柄在测试载荷下初期稳定性相似。
研究设计与样本
- 研究纳入6对尸体股骨,行CT成像后制作扩展转子截骨并用3根钢缆复位固定,每对中一侧股骨植入两种商用带花键锥形开槽柄之一。
- 通过激光扫描测量扩髓器和假体的位置与方向以评估柄骨接触,随后施加1000次循环生物力学加载测试,用运动捕捉系统测量柄骨微动。
主要发现
- 两种柄加载后的平均下沉量分别为313±138 μm和181±113 μm(P > 0.05)。
- 两种柄的最大微动均出现在第一循环,后续循环保持恒定。
- 两种柄的前侧柄骨接触均多于后侧,平均接触面积分别为1546±533 mm2和1525±587 mm2(P > 0.05)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Tapered fluted stems are an excellent solution for revision hip arthroplasty following periprosthetic fractures. However, existing implant designs differ in terms of the number of splines, taper angle, and cross-sectional design, and there is a lack of knowledge on the biomechanical effect of these design variations. Therefore, the objective of this study was to compare the primary stability of two representative stem designs and the extent and distribution of stem-bone contact in a periprosthetic fracture model. It was hypothesized that a greater number of splines would lead to greater stability. METHODS: There were six pairs of cadaver femora included and underwent computed tomography imaging for templating. An extended trochanteric osteotomy was created and then reduced using three cerclage cables. Subsequently, one femur in each pair was implanted with 1 of 2 commercial designs of tapered fluted stems. During specimen preparation, the position and orientation of the reamer and the stems were measured by laser scanning, allowing measurement of stem-bone contact. Subsequently, each specimen was subjected to 1,000 cycles of biomechanical loading tests, and stem-bone motion was measured using a motion capture system. RESULTS: The average subsidence of the two stems after loading was 313 ± 138 and 181 ± 113 μm (P > 0.05). The largest micromotion during each loading-unloading cycle was observed in the first cycle for both stems, and it remained constant in the subsequent cycles. There was more anterior stem-bone contact than posterior region in both stems, and the average contact area was comparable (1,546 ± 533 and 1,525 ± 587 mm2, P > 0.05). CONCLUSIONS: Different geometric designs did not lead to significant changes in the extent and distribution of stem-bone contact, and both stems demonstrated similar primary stability under the loads tested in this study.
引用
- Han S, Mascarenhas DC, Pletka CA, et al. Does Stem Design Affect Primary Stability of Splined Tapered Fluted Stems? A Biomechanical Cadaver Study. J Arthroplasty. 2025. PMID: 40339936. doi:10.1016/j.arth.2025.04.064
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。