预攻丝在腰椎脊柱内固定中不能有效预防椎弓根骨折:一项生物力学尸体研究
该生物力学尸体研究利用4具尸体的20个腰椎椎体,交替采用预攻丝与非预攻丝技术置入椎弓根螺钉,评估预攻丝是否降低椎弓根骨折风险。结果显示两组最大弯曲失效载荷无显著差异(预攻丝组103.9±47.5 Nm vs 非预攻丝组98.1±43.4 Nm,P=0.321),预攻丝显著降低最大置入扭矩(1.30±1.05 vs 2.91±2.65 Nm,P=0.025),椎弓根填充率是弯曲力的显著预测因子(β=222.29 Nm,P=0.035)。结论认为预攻丝不影响生物力学稳定性或骨折风险,优化椎弓根填充率比预攻丝更重要。
要点
- 预攻丝与非预攻丝组最大弯曲失效载荷无显著差异(103.9±47.5 vs 98.1±43.4 Nm,P=0.321)。
- 预攻丝显著降低最大置入扭矩(1.30±1.05 vs 2.91±2.65 Nm,P=0.025)。
- 椎弓根填充率是弯曲力的显著预测因子(β=222.29 Nm,P=0.035),骨密度和椎弓根直径不是。
核心结论
- 预攻丝不显著影响椎弓根螺钉的生物力学稳定性或屈曲力矩下的椎弓根骨折风险。
- 椎弓根填充率是螺钉固定强度的关键因素。
- 这些发现提示优化椎弓根填充率比预攻丝更能增强构造稳定性。
研究设计与样本
- 生物力学尸体研究,取4具尸体的20个腰椎椎体(L1-L5)。
- 术前CT扫描用于规划螺钉轨迹和评估骨密度(BMD)。
- 每个椎体交替采用预攻丝和非预攻丝技术置入螺钉。
- 生物力学测试测量屈曲力矩下的最大弯曲失效载荷。
- 根据CT影像对椎弓根骨折进行分类,并进行统计分析以评估影响骨折风险的因素。
主要发现
- 预攻丝组最大弯曲失效载荷为103.9±47.5 Nm,非预攻丝组为98.1±43.4 Nm,两组间无显著差异(P=0.321)。
- 预攻丝组最大置入扭矩为1.30±1.05 Nm,非预攻丝组为2.91±2.65 Nm,预攻丝显著降低最大置入扭矩(P=0.025)。
- 椎弓根填充率是弯曲力的显著预测因子(β=222.29 Nm,P=0.035),骨密度和椎弓根直径不是显著预测因子。
- 两组间骨折模式无显著差异(P=0.384)。
临床怎么用
- 原文结论未据此给出临床推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
STUDY DESIGN: Biomechanical cadaveric study. OBJECTIVE: To evaluate whether pretapping reduces the risk of pedicle fractures following pedicle screw instrumentation. SUMMARY OF BACKGROUND DATA: Pedicle screw instrumentation is the standard for rigid fixation of the thoracolumbar spine. While complications such as screw loosening and misplacement are well-documented, pedicle fractures are rare and poorly understood. Pretapping, a technique used during screw insertion, has been proposed to enhance fixation strength and potentially reduce the risk of fractures. However, its biomechanical effects on pedicle integrity remain unclear. METHODS: Twenty lumbar vertebrae (L1-L5) were harvested from four cadavers. Preoperative CT scans were used to plan screw trajectories and assess bone mineral density (BMD). Each vertebra was alternately instrumented with screws using pretapping and non-pretapping techniques. Biomechanical testing measured maximum bending force to failure under a flexion moment. Pedicle fractures were classified based on CT imaging, and statistical analysis was performed to evaluate factors influencing fracture risk. RESULTS: The maximum bending force to failure showed no significant difference between pretapping (103.9±47.5 Nm) and non-pretapping (98.1±43.4 Nm) groups ( P =0.321). Pretapping significantly reduced maximum insertion torque (1.30±1.05 vs. 2.91±2.65 Nm, P =0.025). Pedicle fill was a significant predictor of bending force (β=222.29 Nm, P =0.035), while BMD and pedicle diameter were not. Fracture patterns did not differ significantly between groups ( P =0.384). CONCLUSIONS: Pre-tapping does not significantly influence the biomechanical stability of pedicle screws or the risk of pedicle fractures under flexion moments. However, pedicle fill is a critical factor in screw fixation strength. These findings suggest that optimizing pedicle fill is more relevant than pre-tapping in enhancing construct stability.
引用
- Bauer DE, Mari Major L, Wigger O, et al. Pretapping Is Not Effective in Preventing Pedicle Fractures in Lumbar Spinal Instrumentation: A Biomechanical Cadaveric Study. Spine (Phila Pa 1976). 2025. PMID: 40325503. doi:10.1097/BRS.0000000000005384