无骨水泥股骨固定后温哥华B型假体周围骨折:单中心经验
该单中心研究纳入2009至2023年12400例初次全髋关节置换术,评估无骨水泥股骨柄术后温哥华B型假体周围骨折的发生率与结局。温哥华B型骨折总体发生率为0.3%,手术时年龄>75岁患者发生率高于18-75岁患者(0.6% vs 0.2%,P<0.001),C3型三锥度带领钛柄骨折率低于A型平锥柄(0.1% vs 0.7%,P<0.001)。
要点
- 温哥华B型假体周围骨折总体发生率为0.3%(34/12400),中位发生时间为术后33天。
- 手术时年龄>75岁患者温哥华B型骨折率高于18-75岁患者(0.6% vs 0.2%,P<0.001)。
- C3型三锥度带领钛柄骨折率低于A型平锥柄(0.1% vs 0.7%,P<0.001)。
核心结论
- 温哥华B型假体周围骨折在手术时年龄>75岁患者中发生率更高,与本研究中心既往报道一致。
- 本研究中心发现C3型三锥度带领钛柄在所有年龄组(包括手术时>75岁)均降低了假体周围骨折率。
- 二次再手术对患者仍是高度致病的。
研究设计与样本
- 单中心数据库回顾性识别2009年至2023年期间12400例初次全髋关节置换术。
- 手术时平均年龄64岁,57%的全髋关节置换术在女性中实施。
- 识别所有术后假体周围骨折,对分类为温哥华B型的骨折评估治疗方法和二次再手术。
主要发现
- 72例术后假体周围骨折中34例分类为温哥华B型,总体发生率为0.3%(34/12400)。
- 从全髋关节置换术到温哥华B型假体周围骨折的中位时间为33天。
- 手术时年龄>75岁患者温哥华B型假体周围骨折率高于18至75岁患者(0.6%,13/2094 vs 0.2%,21/10306,P<0.001)。
- C3型三锥度带领钛柄骨折率低于A型平锥柄(0.1%,4/4748 vs 0.7%,23/3105,P<0.001)。
- 手术时>75岁患者中,C3型柄骨折率低于A型柄(0.3%,3/923 vs 1.5%,7/472,P=0.04)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Cementless stems have been the predominant type of femoral component used for total hip arthroplasty (THA) in the United States for several decades. However, recent literature has reported an increased periprosthetic fracture (PPFx) rate and complications associated with these components, particularly among older patients. This study evaluated the incidence and outcome of Vancouver B PPFx after primary THA using cementless stems. METHODS: A single institution's database was used to identify 12,400 primary THAs performed from 2009 through 2023. The mean age at surgery was 64 years, and 57% of THAs were performed among women. All postoperative PPFx were identified, and those classified as Vancouver B were evaluated for treatment method and secondary reoperations. RESULTS: Among 72 postoperative PPFx, 34 were classified as Vancouver B for an overall rate of 0.3% (34 of 12,400). The median time from THA to Vancouver B PPFx was 33 days. Patients aged > 75 years at surgery had a higher rate of Vancouver B PPFx (0.6%, 13 of 2,094) compared to patients aged 18 to 75 years at surgery (0.2%, 21 of 10,306, P < 0.001). Type C3 triple-taper collared titanium stems had a lower fracture rate (0.1%, four of 4,748) compared to Type A flat taper stems (0.7%, 23 of 3,105, P < 0.001). In patients over 75 at surgery, Type C3 stems had a lower fracture rate (0.3%, three of 923) compared to Type A stems (1.5%, seven of 472, P = 0.04). There were 28 patients who had Vancouver B fractures (82%) who underwent reoperation. Secondary reoperations were performed among 25% (seven of 28) of THAs, and three of these involved periprosthetic joint infection. CONCLUSIONS: Consistent with other reports, the Vancouver B PPFx rate was higher in patients aged > 75 years at surgery. At our institution, Type C3 triple-taper collared titanium stems lowered the PPFx fracture rate in all age groups, including those over age 75 years at surgery. Secondary reoperations remain highly morbid to patients.
引用
- Mosher ZA, Strait AV, Olson NR, et al. Vancouver B Fractures After Using Cementless Femoral Fixation: A Single Center Experience. J Arthroplasty. 2025. PMID: 40107572. doi:10.1016/j.arth.2025.03.024
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。