机器人辅助手术并未减少全髋关节置换术中的假体撞击:一项取回物分析
该取回物分析比较了18例机器人辅助与11例手动放置的全髋关节置换术髋臼内衬的撞击发生率和严重程度。机器人组撞击率为61%,对照组为45%(P=0.14),两组间无显著差异;髋臼倾斜度和前倾角的均值及方差在两组间亦无显著差异。结论提示,采用现代大直径股骨头的机器人辅助技术并未减少翻修全髋关节置换术取回髋臼内衬的假体撞击发生率或严重程度。
要点
- 机器人辅助全髋关节置换术髋臼内衬撞击率为61%,手动对照组为45%(P=0.14),无显著差异。
- 机器人组髋臼倾斜度方差较低(41.7°±3.9 vs 42.8°±7.1,P=0.64),前倾角方差较低(22.8°±2.8 vs 20.8°±7.9,P=0.43),但均值无差异。
- 撞击存在与严重程度与股骨头尺寸、植入时长或其他人口学变量无关。
核心结论
- 本研究提示,采用现代大直径股骨头的机器人辅助技术并未减少翻修全髋关节置换术取回髋臼内衬的假体撞击发生率或严重程度。
研究设计与样本
- 取回物分析,纳入18例机器人辅助全髋关节置换术髋臼内衬和11例非机器人手动放置的对照内衬,评估撞击存在与严重程度,并使用翻修前站立位X线片评估髋臼倾斜度和前倾角。
主要发现
- 机器人组61%的髋臼内衬显示撞击,非机器人对照组为45%(P=0.14)。
- 机器人组髋臼倾斜度方差较低(机器人组41.7°±3.9 vs 对照组42.8°±7.1,P=0.64),前倾角方差较低(机器人组22.8°±2.8 vs 对照组20.8°±7.9,P=0.43),但两组均值无差异。
- 撞击存在与严重程度与股骨头尺寸、植入时长或其他人口学变量无关。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Prosthetic impingement after total hip arthroplasty (THA) has been associated with instability and may be a cause of accelerated polyethylene wear and pain. Previous retrieval studies report a high prevalence of impingement in acetabular liners. Robotic technology has the potential to reduce THA instability as it enables technical precision and optimizes implant positioning. However, whether robotics can improve impingement prevalence is unknown. Thus, the objectives of this study were to: (1) determine the prevalence and severity of acetabular liner impingement with robotic navigation; and (2) compare impingement prevalence with a control cohort of manually placed THA liners. METHODS: There were 18 robotic-assisted liners and 11 non-robotic controls scored for the presence and severity of impingement. Radiographic measurements of acetabular inclination and anteversion were assessed using prerevision standing radiographs. Femoral head size (36 being the most common), length of implantation, revision indication, age, gender, and body mass index were recorded. RESULTS: Of the robotic liners, 61% showed impingement, while 45% of the non-robotic liners showed impingement (P = 0.14). The robotic group demonstrated a lower variance of inclination (robotic: 41.7° ± 3.9, control: 42.8° ± 7.1, P = 0.64) and lower anteversion variance (robotic: 22.8° ± 2.8, control: 20.8° ± 7.9, P = 0.43), but mean values did not differ between the groups. Impingement presence and severity were not related to head size, length of implantation, or other demographic variables. CONCLUSIONS: This study suggests that the use of robotic-assisted technology employing modern-day implants with larger diameter heads does not reduce the prevalence or severity of prosthetic impingement in retrieved acetabular liners from revision THA.
引用
- Sacher SE, A O'Donnell J, Wright TM, et al. Robotic-Assisted Surgery Does Not Decrease Prosthetic Impingement in Total Hip Arthroplasty: A Retrieval Analysis. J Arthroplasty. 2025. PMID: 40057210. doi:10.1016/j.arth.2025.02.077
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。