钴铬钼股骨膝假体损伤与假体周围金属浓度升高相关
研究前瞻性收集51例钴铬钼股骨膝关节假体及滑膜样本,评估假体损伤评分与假体周围组织钴、铬、钼、钛浓度的关联。结果显示视觉损伤与钴、铬、钼浓度升高相关(P=0.0008、0.029、0.007),严重损伤假体周围组织中位钴、铬、钼浓度分别为7.81、5.26、0.713 μg/mL。51例中14%组织样本钴和铬浓度>10 μg/mL。
要点
- 假体视觉损伤与假体周围组织钴、铬、钼浓度升高相关。
- 严重损伤假体周围组织中位钴7.81、铬5.26、钼0.713 μg/mL。
- 51例中14%组织样本钴和铬浓度>10 μg/mL。
核心结论
- 研究显示全膝关节置换患者假体周围组织金属浓度升高。
- 研究建立了假体损伤严重程度与后续金属释放之间的正相关。
- 测得的组织金属浓度接近金属对金属全髋关节置换后报道的水平。
研究设计与样本
- 研究前瞻性收集51个钴铬钼股骨假体及滑膜样本,来源于机构审查委员会豁免的回收项目。
- 假体损伤评分从轻微(1分)到严重(4分)。
- 使用电感耦合等离子体质谱法量化组织钴、铬、钼、钛浓度。
主要发现
- 视觉损伤与假体周围组织钴、铬、钼浓度升高相关(P=0.0008、0.029、0.007)。
- 严重损伤假体周围组织中位钴、铬、钼浓度分别为7.81、5.26、0.713 μg/mL。
- 轻微损伤假体中位钴、铬、钼浓度分别为0.111、1.80、0.179 μg/mL。
- 51例中14%组织样本钴和铬浓度>10 μg/mL。
- 钛胫骨托假体周围组织钛浓度升高(P=0.0052),部分源于关节置换过程中胫股部件接触。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Total knee arthroplasty (TKA) device systems frequently include cast cobalt chrome alloy (CoCrMo) femoral components. However, compared to total hip arthroplasty (THA), gaps persist in our understanding of the correlations between femoral component damage, local metal release, and potential biological effects. Additionally, it remains unclear how TKA metal release affects clinical success or patient satisfaction. In this study, we investigated the associations between implant damage and metal release in the periprosthetic tissues following TKA. We asked: (1) does damage severity correlate with increased metal concentrations within the periprosthetic tissue? and (2) does the magnitude of metal released from CoCrMo femoral components merit clinical concern? METHODS: There were 51 CoCrMo femoral components and synovial samples that were prospectively collected by an institutional review board-exempt retrieval program. Devices received damage scores ranging from minimal (1) to severe (4). Tissue metal concentrations for cobalt (Co), chromium (Cr), molybdenum (Mo), and titanium (Ti) were quantified using inductively coupled plasma mass spectroscopy. RESULTS: Visual damage correlated with increases in Co, Cr, and Mo concentrations in the periprosthetic tissue (P = 0.0008, 0.029, and 0.007, respectively). Within the tissue adjacent to severely damaged implants, we measured median Co, Cr, and Mo concentrations of 7.81, 5.26, and 0.713 μg/mL, respectively. For minimally damaged implants, we report median Co, Cr, and Mo concentrations of 0.111, 1.80, and 0.179 μg/mL, respectively. In several of the 51 (14%) tissue samples, we measured Co and Cr concentrations > 10 μg/mL. Additionally, within periprosthetic tissues of devices with Ti tibial trays, Ti concentrations increased (P = 0.0052) arising, in part, from tibial-femoral component contact during arthroplasty. CONCLUSIONS: We (1) showed elevated periprosthetic tissue metal concentrations in TKA patients and (2) established a positive correlation between damage severity and subsequent metal release. Measured tissue metal concentrations approached the magnitudes reported following metal-on-metal THA.
引用
- Kurtz PW, Aslani S, Kurtz MA, et al. Cobalt-Chromium-Molybdenum Femoral Knee Implant Damage Correlates With Elevated Periprosthetic Metal Concentrations. J Arthroplasty. 2025. PMID: 40054540. doi:10.1016/j.arth.2025.02.075
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。