使用超匹配胫骨衬垫的非骨水泥全膝关节置换术:至少5年随访结果
本研究在两家机构连续纳入242例使用超匹配(UC)运动学胫骨衬垫的非骨水泥全膝关节置换术(TKA),平均随访5.6年。末次随访KOOS-JR平均84.0分,Knee Society评分平均94分,屈曲平均116°;7膝(2.9%)翻修,5年无菌性松动生存率99.6%。结论认为该匹配设计非骨水泥TKA临床结果优良,UC特性似不影响假体固定耐久性。
要点
- 242例非骨水泥TKA使用UC胫骨衬垫,平均随访5.6年,KOOS-JR平均84.0分。
- 7膝(2.9%)翻修,5年无菌性松动生存率99.6%。
- CPAK类型改变与末次KOOS-JR、疼痛、UCLA活动及满意度评分无相关性。
核心结论
- 使用该匹配设计的非骨水泥TKA在5年时提供了优良的临床结果。
- 一旦获得初始部件稳定性,该超匹配关节的UC特性似乎不会对植入物固定的耐久性产生不良影响,无论CPAK类型是否改变。
研究设计与样本
- 连续系列研究,2017年至2019年在两家机构植入242例使用UC运动学胫骨衬垫的非骨水泥TKA。
- 患者56%为男性,平均年龄60岁,平均随访5.6年(范围5至7.5年)。
- 采用KOOS-JR、Knee Society临床与影像学评分随访,术前与术后X线片按CPAK类型分类。
主要发现
- 末次随访KOOS-JR评分平均84.0分,Knee Society评分平均94分,屈曲平均116°。
- 7膝(2.9%)行翻修:3膝因屈曲不稳、1膝因疼痛、1膝因股骨纤维长入、1膝因股骨远端骨折、1膝因关节纤维化。
- 另有7例(2.9%)手法操作、1例髌骨骨折,无深部感染。
- 5年无菌性松动生存率为99.6%。
- CPAK类型改变与末次KOOS-JR、疼痛、UCLA活动评分或满意度评分无相关性。
临床怎么用
- 原文未据此给出推荐。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Cementless fixation for primary total knee arthroplasty (TKA) continues to increase in the United States. However, compared to cemented TKA, reports on revision rates have been mixed. A confounding variable may include the tibial insert design. This study aimed to assess the minimum 5-year survivorship and outcomes of a cementless TKA using an ultracongruent (UC) articulation. METHODS: A consecutive series of 242 cementless TKAs were implanted at two institutions between 2017 and 2019 using an UC kinematic tibial insert. Of the patients, 56% were men. The average age was 60 years. Patients were followed using Knee Injury and Osteoarthritis Outcome Score for Joint Replacement (KOOS-JR) and Knee Society clinical and radiographic scores. Preoperative and postoperative radiographs were classified according to the Coronal Plane Alignment of the Knee (CPAK) type. The average follow-up was 5.6 years (range, five to 7.5). RESULTS: At the final follow-up, KOOS-JR scores averaged 84.0. Knee Society scores averaged 94. Flexion averaged 116°. There were seven manipulations (2.9%), one patella fracture, and no deep infections. There were seven knees (2.9%) revised (three for flexion instability, one for pain, one for femoral fibrous ingrowth, one for distal femur fracture, and one for arthrofibrosis). At 5 years, survivorship free from aseptic loosening was 99.6%. Change in CPAK type did not correlate with final KOOS-JR, pain, University of California Los Angeles activity score, or satisfaction scores. CONCLUSIONS: Cementless TKA using this conforming design has provided excellent clinical results out to 5 years. Once initial component stability is achieved, the UC nature of this articulation does not appear to adversely influence the durability of implant fixation, regardless of whether the CPAK type was changed.
引用
- Meding JB, Meneghini RM, Meding LK, et al. Cementless Total Knee Arthroplasty Using an Ultraconforming Tibial Bearing: Outcomes at Minimum 5-Year Follow-Up. J Arthroplasty. 2025. PMID: 40020944. doi:10.1016/j.arth.2025.02.054
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。