无交叉韧带全膝关节置换假体设计以实现解剖性股胫运动与松弛度

Total Knee Arthroplasty Design Without Cruciates to Achieve Anatomic Femoral-Tibial Motion and Laxity.
PubMed 文献 · J Arthroplasty · 2025-03-12

该研究提出一种保留内侧稳定性和外侧活动性的无交叉韧带全膝关节置换假体设计,通过中央偏移圆柱形承重面在屈曲时诱导渐进性轴向旋转和外侧「后滚」,并提供前后向和旋转松弛度。研究用三维打印模型和测试机比较该设计与四种当代无交叉韧带全膝关节置换假体,发现在剪切和扭矩力下该设计更符合设计标准。

要点

  • 无交叉韧带全膝关节置换假体设计采用内侧稳定、外侧活动的主承重面,中央偏移圆柱面诱导屈曲时渐进性轴向旋转和外侧「后滚」。
  • 该设计同时提供前后向和旋转松弛度,模拟正常膝关节。
  • 体外实验显示,带髁间圆柱面的设计较其他四种当代无交叉韧带假体更符合设计标准。

核心结论

  • 对于无交叉韧带设计,要产生更正常的运动特征,可能需要某种与内外侧髁协同作用的机械构型。

研究设计与样本

  • 研究提出一种无交叉韧带全膝关节置换假体设计,并制作三维打印模型与四种当代无交叉韧带全膝关节置换假体进行实验比较。
  • 构建测试机在多个屈曲角度施加剪切力和扭矩力,测定接触位置。

主要发现

  • 带髁间圆柱面的设计较其他设计更符合设计标准。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: A frequently stated goal of an artificial knee arthroplasties is to achieve normal kinematics. However, this is not easily defined based on variations in motions previously measured for a range of activities. For activities such as crouching up and down, a fan pattern has been measured, where the lateral femoral contact displaces progressively posteriorly with flexion, and the medial contact remains almost constant. In walking and other activities, femoral-tibial contacts vary considerably in position, and even lateral pivoting has been measured at the start of the motion cycle. Fluoroscopic studies of total knee arthroplasty (TKA) patients have shown that such kinematics is not usually achieved for most TKA designs. In recent years, there has been an increasing interest in noncruciate retaining knee arthroplasties, where both cruciate ligaments are resected. A challenge with such designs is to define the design criteria, taking account of the extensive kinematic data of normal knees, as well as clinical factors. METHODS: A TKA design was formulated where the main bearing surfaces produced medial stability and lateral mobility, but where the addition of an offset cylindrical bearing surface in the center induced progressive axial rotation and lateral "rollback" with flexion. At the same time, anterior-posterior and rotational laxity were provided, as in the normal knee. The new design was compared experimentally with four types of contemporary noncruciate total knee arthroplasties. There were three-dimensional printed models fabricated. A test machine was constructed where shear and torque forces were applied at a range of flexion angles, and contact positions were determined. RESULTS: It was found that the design with the intercondylar cylindrical surface satisfied the design criteria more closely compared with the other designs. CONCLUSIONS: For noncruciate designs to produce more normal motion characteristics, some mechanical configuration acting in concert with the lateral and medial condyles is likely to be necessary.

来源:J Arthroplasty,查看原文。版权归原作者所有。