文献速递关节外科

干骺端袖套与多孔锥对翻修全膝关节置换术风险降低的影响:美国整合医疗系统中无菌性再翻修的队列研究

Impact of Metaphyseal Sleeves and Porous Cones on Risk Reduction for Revision Total Knee Arthroplasty: Aseptic Re-Revision in a United States Integrated Health Care System.
PubMed 文献 · J Arthroplasty · 2025-02-24

基于美国整合医疗系统关节置换登记处的队列研究,纳入2008至2022年2882例无菌性翻修全膝关节置换术,比较仅使用带柄组件(2264例)与柄+锥/袖套(618例)的再翻修风险。经倾向评分加权Cox回归,8年随访时两组无菌性再翻修风险无显著差异(HR=0.76,95%CI 0.42-1.37),无菌性松动再翻修亦无显著差异(HR=0.62,95%CI 0.27-1.43)。作者认为加用干骺端袖套或多孔锥未改变再翻修风险,需进一步大规模长期随访研究证实。

要点

  • 8年随访时,柄+锥/袖套组与仅柄组无菌性再翻修风险无显著差异(HR=0.76,95%CI 0.42-1.37)。
  • 无菌性松动再翻修风险两组亦无显著差异(HR=0.62,95%CI 0.27-1.43)。
  • 作者认为加用干骺端袖套或多孔锥未改变无菌性再翻修风险。

核心结论

  • 在超过2500例翻修全膝关节置换术的队列中,加用干骺端袖套或多孔锥未改变无菌性再翻修总体风险或无菌性松动再翻修风险。
  • 作者指出需进一步大规模长期随访研究证实这些结果。

研究设计与样本

  • 美国整合医疗系统关节置换登记处的队列研究,纳入2008至2022年2882例无菌性翻修全膝关节置换术。
  • 仅带柄组件组2264例,柄+锥/袖套组618例。
  • 采用倾向评分加权Cox比例风险回归评估无菌性再翻修及无菌性松动再翻修风险。

主要发现

  • 8年随访时,经倾向评分加权,柄+锥/袖套组与仅柄组无菌性再翻修风险无显著差异(HR=0.76,95%CI 0.42-1.37)。
  • 8年随访时,经倾向评分加权,柄+锥/袖套组与仅柄组无菌性松动再翻修风险无显著差异(HR=0.62,95%CI 0.27-1.43)。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 作者指出需进一步大规模长期随访研究证实这些结果。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Bone loss in revision total knee arthroplasty (rTKA) makes achieving a well-fixed construct challenging. Metaphyseal sleeves and porous cones were designed to enhance fixation during rTKA and are typically used in combination with modular stems. However, there are limited outcome data comparing rTKA constructs with stems and sleeves or cones to those with stems alone. Therefore, we sought to compare re-revision rates in patients undergoing rTKA with cones/sleeves + stems to those with just stems. METHODS: We conducted a cohort study from a United States health care system's arthroplasty registry. The study sample included 2,882 aseptic rTKAs that included a stemmed component only (N = 2,264) or a stem and a cone/sleeve (N = 618) (2008 to 2022). Propensity score-weighted Cox proportional hazard regression was used to evaluate the risk of aseptic re-revision and re-revision due to loosening specifically. RESULTS: At 8-year follow-up, after propensity score weighting, no difference was observed between stem + cone/sleeve compared with stem alone for aseptic re-revision (hazard ratio = 0.76, 95% confidence interval = 0.42 to 1.37) or aseptic loosening specifically (hazard ratio = 0.62, 95% confidence interval = 0.27 to 1.43). CONCLUSIONS: In a cohort of over 2,500 rTKAs, we found the addition of metaphyseal sleeves or porous cones did not change the risk of aseptic re-revision overall or re-revision due to loosening specifically. Further large-scale studies with long-term follow-up are warranted to confirm these results.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Hinman AD, Chang RN, Kelly MP, et al. Impact of Metaphyseal Sleeves and Porous Cones on Risk Reduction for Revision Total Knee Arthroplasty: Aseptic Re-Revision in a United States Integrated Health Care System. J Arthroplasty. 2025. PMID: 40010446. doi:10.1016/j.arth.2025.02.050

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。