文献速递关节外科

评估基于计算机断层扫描的混合现实导航工具在全髋关节置换术中髋臼假体定位的准确性

Evaluating the Accuracy of a Computed Tomography-Based Mixed-Reality Navigation Tool for Acetabular Component Positioning in Total Hip Arthroplasty.
PubMed 文献 · J Arthroplasty · 2025-02-13

本研究回顾性纳入79例由单一术者使用新型混合现实(MR)导航系统行全髋关节置换术(THA)的患者,最终39例符合纳入标准。术后6周三维EOS影像测量显示,髋臼假体倾斜度与前倾角的术后测量值与术前计划值无显著差异,平均绝对误差分别为1.8度和2.0度,倾斜度与前倾角均无超出计划目标±10度的离群值。作者认为THA术中使用MR导航可获得相对术前计划的准确髋臼假体倾斜度与前倾角放置,且无离群值。

要点

  • 39例THA患者术后倾斜度平均绝对误差1.8度、前倾角2.0度。
  • 术后倾斜度与前倾角相对术前计划值均无显著差异(P分别为0.40和0.11)。
  • 倾斜度与前倾角均无超出计划目标±10度的离群值。

核心结论

  • 作者认为THA术中使用MR导航可获得相对术前计划的准确髋臼假体倾斜度与前倾角放置,且无离群值。

研究设计与样本

  • 回顾性综述单术者使用新型MR导航系统完成的79例THA患者,最终39例符合纳入标准;术前收集计划髋臼假体手术倾斜度(OI)和倾斜校正手术前倾角(OA),术后6周行三维EOS影像并用经验证的测量软件测量假体位置。
  • 双侧THA或三维影像缺乏可用于测量的关键骨性标志者被排除。

主要发现

  • 每例患者平均计划OI为40.8度(范围40至41度),术后平均测量OI为40.4度(范围37至46度)(P=0.40),平均绝对误差1.8度(范围0至5度)。
  • 平均计划OA为30.2度(范围25至35度),术后平均测量OA为31.1度(范围27至42度)(P=0.11),平均绝对误差2.0度(范围0至6度)。
  • OI与OA均无离群值(离群值定义为假体位置超出术前计划目标±10度范围)。

关键数值与推荐

  • 离群值定义为任何超出术前计划目标±10度范围的假体位置。
  • 平均计划OI为40.8度(范围40至41度),平均测量术后OI为40.4度(范围37至46度),P=0.40。
  • 平均计划OA为30.2度(范围25至35度),平均测量术后OA为31.1度(范围27至42度),P=0.11。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Robotics and navigation systems have improved the accuracy and precision of acetabular component placement in total hip arthroplasty (THA). The purpose of this study was to determine if the intraoperative use of a novel mixed-reality (MR) navigation system resulted in accurate acetabular component placement with minimal outliers. METHODS: A series of 79 patients who underwent THA performed by a single surgeon using a novel MR navigation system were retrospectively reviewed. The preoperative planned acetabular component operative inclination (OI) and tilt-adjusted operative anteversion (OA) were collected for each case. At the six-week follow-up, each patient underwent three-dimensional EOS imaging, which was used to measure the acetabular component placement via validated measuring software. Patients who had bilateral THAs or three-dimensional images that did not have critical bony landmarks visible for measurement were excluded from the study. Absolute error was calculated for both inclination and anteversion. Outliers were defined as any cup position that was outside a range of ± 10 degrees from the preoperatively planned target. A total of 39 patients met the inclusion criteria. RESULTS: The mean planned OI for each patient was 40.8 (range, 40 to 41) degrees, and the mean measured postoperative OI was 40.4 (range, 37 to 46) degrees (P = 0.40), resulting in a mean absolute error of 1.8 (range, zero to five) degrees. The mean planned OA was 30.2 (range, 25 to 35) degrees, and the mean measured postoperative OA was 31.1 (range, 27 to 42) degrees (P = 0.11), resulting in a mean absolute error of 2.0 (range, zero to six) degrees. There were no outliers in either OI or OA. CONCLUSIONS: The results of this study suggest that the use of MR navigation during THA results in accurate postoperative acetabular component inclination and anteversion relative to the preoperative plan without outliers.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Dilbone ES, Heimann AF, Leal J, et al. Evaluating the Accuracy of a Computed Tomography-Based Mixed-Reality Navigation Tool for Acetabular Component Positioning in Total Hip Arthroplasty. J Arthroplasty. 2025. PMID: 39954749. doi:10.1016/j.arth.2025.02.003

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。