基于计算机断层扫描的机器人辅助全膝关节置换术在实现矢状面力线目标方面比手动器械更准确

Computed Tomography-Based Robotics Are More Accurate than Manual Instruments in Achieving Sagittal Alignment Targets in Total Knee Arthroplasty.
PubMed 文献 · J Arthroplasty · 2025-02-15

本研究比较了201例机器人辅助全膝关节置换术(RA-TKA)与365例手动器械全膝关节置换术(M-TKA)在实现预定冠状面和矢状面力线目标方面的准确性。结果显示RA-TKA在全部四个矢状面力线目标和三个冠状面目标中的两个上更准确,影像学异常值比例更低。

要点

  • RA-TKA在股骨远端角、胫骨近端角、股骨矢状角、胫骨矢状角、前髁偏移和后髁偏移比方面异常值均少于M-TKA。
  • RA-TKA组无影像学异常值的患者比例更高(58.2% vs 35.2%),出现≥2个异常值的比例更低(5.5% vs 25%)。
  • RA-TKA在实现冠状面和矢状面力线方面比M-TKA更有效,可能改善手术结果。

核心结论

  • RA-TKA在实现冠状面和矢状面力线方面比M-TKA更有效,可能改善手术结果。

研究设计与样本

  • 回顾性比较研究,纳入201例RA-TKA(2021年6月至2022年6月由三位高手术量、经过专科培训的外科医生完成)与365例M-TKA(2013年至2017年由同一组外科医生完成)作为历史对照。

主要发现

  • RA-TKA组相比M-TKA组,股骨远端角异常值更少(0% vs 2.5%;P=0.03)。
  • RA-TKA组相比M-TKA组,胫骨近端角异常值更少(1% vs 10.1%;P<0.001)。
  • RA-TKA组相比M-TKA组,股骨矢状角异常值更少(7.0% vs 15.6%;P<0.01)。
  • RA-TKA组相比M-TKA组,胫骨矢状角异常值更少(5.0% vs 14.3%;P<0.01)。
  • RA-TKA组相比M-TKA组,前髁偏移异常值更少(8.5% vs 30.6%;P<0.01),后髁偏移比异常值更少(1.5% vs 9.5%;P<0.01),无影像学异常值的患者比例更高(58.2% vs 35.2%;P<0.001),出现≥2个异常值的比例更低(5.5% vs 25%;P<0.001)。

关键数值与推荐

  • 正常范围和异常值范围根据既往研究确定。
  • 比较的测量指标包括股骨胫骨角、股骨远端内侧角、胫骨近端角、股骨矢状角、胫骨矢状角、前髁偏移和后髁偏移比。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Implant malalignment may predispose patients to implant failure or pain following total knee arthroplasty (TKA). Previous studies indicate that robotically-assisted total knee arthroplasty (RA-TKA) can achieve coronal alignment targets more accurately and precisely than manually instrumented total knee arthroplasty (M-TKA). The purpose of this study was to evaluate the accuracy of RA-TKA versus M-TKA in achieving predetermined coronal and sagittal alignment targets in TKA. METHODS: A total of 201 RA-TKAs performed by three high-volume, fellowship-trained surgeons between June 2021 and June 2022 were compared to a historical control of 365 M-TKAs performed between 2013 and 2017 by the same surgeons. Coronal and sagittal component alignments were assessed using standing antero-posterior and lateral radiographs. Included measurements were femoro-tibial angle (FTA), medial distal femoral angle, proximal tibial angle, femoral sagittal angle, tibial sagittal angle, anterior condylar offset, and posterior condylar offset ratio. Normal and outlier ranges were determined from prior studies. The proportions of outliers were compared using univariate analyses. RESULTS: The RA-TKA was more accurate than M-TKA in achieving all four sagittal alignment targets and two of the three coronal alignment targets. The RA-TKA group exhibited fewer radiographic outliers for distal femoral angle (zero versus 2.5%; P = 0.03), proximal tibial angle (one versus 10.1%; P < 0.001), femoral sagittal angle (7.0 versus 15.6%; P < 0.01), tibial sagittal angle (5.0 versus 14.3%; P < 0.01), anterior condylar offset (8.5 versus 30.6%; P < 0.01), and posterior condylar offset ratio (1.5 versus 9.5%; P < 0.01). Patients in the RA-TKA group had a higher proportion with no radiographic outliers (58.2 versus 35.2%; P < 0.001) and a lower incidence of ≥ two outliers (5.5 versus 25%; P < 0.001) compared to the M-TKA group. CONCLUSIONS: The RA-TKA is more effective than M-TKA in achieving coronal and sagittal alignment, potentially enhancing surgical outcomes.