全膝关节置换术中氧化锆股骨假体:胫骨垫片的取回研究

Oxidized Zirconium Femoral Components in Total Knee Arthroplasty: A Retrieval Study of the Tibial Bearing.
PubMed 文献 · J Arthroplasty · 2025-01-30

本研究通过取回数据库分析137例氧化锆股骨全膝关节置换假体,其中75例胫骨垫片为从未辐照聚乙烯(PE),62例为交联聚乙烯(XL)。两组尺寸变化率和体内氧化率均无统计学差异,XL未表现出优于PE。

要点

  • 氧化锆股骨假体对胫骨垫片的尺寸变化率和氧化率均较低。
  • PE组与XL组的尺寸变化率和氧化率无统计学差异。
  • 仅1例PE假体超过氧化阈值,提示力学性能下降。

核心结论

  • 氧化锆股骨假体对应的胫骨垫片尺寸变化率和氧化率较低,与既往文献报道值一致。
  • 在本分析中,交联聚乙烯未表现出优于聚乙烯的优越性。

研究设计与样本

  • 经机构审查委员会批准的取回数据库检索,纳入2004年至2023年间137例氧化锆股骨全膝关节置换假体。
  • 75例胫骨垫片为从未辐照聚乙烯(PE),62例为交联聚乙烯(XL),所有假体均使用抛光胫骨托。
  • PE组平均植入时间58个月,XL组平均植入时间34个月。
  • 采用短植入期(<16个月)和从未植入的假体作为参考,通过测量聚乙烯垫片厚度得出尺寸变化(蠕变加磨损)。
  • 取回后<3个月,通过傅里叶变换红外光谱测量聚乙烯内侧髁横截面的最大酮氧化指数。

主要发现

  • PE组平均尺寸变化率为0.018±0.024 mm/年,XL组为0.021±0.030 mm/年(P=0.64)。
  • PE组平均氧化率为0.051±0.052任意单位(a.u.)/年,XL组为0.064±0.055 a.u./年(P=0.19)。
  • 仅1例假体(PE)超过氧化阈值,标记力学性能下降。
  • XL组最常见的取回原因为不稳定和感染,PE组最常见的取回原因为不稳定。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Advanced materials for total knee arthroplasty, including crosslinked bearings and cobalt-chrome-molybdenum-alternative components, show promise for reducing wear rates and improving outcomes. The objective of the current study was to test whether an alternative femoral-bearing material achieved these goals. METHODS: An institutional review board-approved retrieval database was searched to identify 137 oxidized zirconium femoral total knee arthroplastycomponents from 2004 to 2023; 75 had a tibial bearing of never-irradiated polyethylene (PE) and 62 had a bearing of cross-linked polyethylene (XL). All had a polished tibial tray. The mean duration of implantation was 58 months for the PE group and 34 months for the XL group. The most common reasons for retrieval were instability and infection in the XL group and instability in the PE group. Dimensional change (creep plus wear) was measured from the thickness of the PE insert using short-duration (< 16 months) and never-implanted devices as a reference. Oxidation (maximum ketone oxidation index) was measured via fourier transform infraredusing a cross-section of the medial condyle of the PE at < three months postexplant. Statistical analyses were conducted using groups matched for in vivo duration. A P < 0.05 was considered significant. RESULTS: Dimensional change rate and in vivo oxidation rate were statistically indistinguishable between the two bearing types. The mean dimensional change rate for the PE group was 0.018 ± 0.024 mm/year and 0.021 ± 0.030 mm/year for the XL group (P = 0.64). The mean oxidation rates were also not statistically different, with the mean rate for the PE group being 0.051 ± 0.052 arbitrary units (a.u.)/year and the mean value for the XL group being 0.064 ± 0.055 a.u./year (P = 0.19). Only one device (PE) exceeded the oxidation threshold, marking a decrease in mechanical properties. CONCLUSIONS: The dimensional change rate and oxidation rate for tibial inserts against oxidized zirconium femoral components were low and consistent with previously reported literature values. The XL did not demonstrate superiority to PE in this analysis.