干扰螺钉、U形钉与缝线锚钉均可为改良Lemaire外侧关节外腱固定术提供足够的失效载荷:一项结构特性的生物力学研究

Adequate Failure Loads for Modified Lemaire Lateral Extra-articular Tenodesis Are Achieved With an Interference Screw, Staple, and Suture Anchor: A Biomechanical Study of Structural Properties.
PubMed 文献 · Am J Sports Med · 2025-01-06

研究将18例新鲜冷冻尸体膝关节随机分配至干扰螺钉、U形钉或缝线锚钉股骨固定组,行改良Lemaire外侧关节外腱固定术并以20 mm/min进行载荷至失败测试。干扰螺钉平均失效载荷最高(252.7±131.2 N),与缝线锚钉(105.7±16.4 N)差异有统计学意义(P=.015),与U形钉(151.8±34.1 N)无显著差异;各固定方法刚度无显著差异(P=.089)。

要点

  • 干扰螺钉失效载荷252.7±131.2 N,与缝线锚钉105.7±16.4 N差异有统计学意义(P=.015)。
  • U形钉151.8±34.1 N与干扰螺钉(P=.101)、缝线锚钉(P=.577)失效载荷均无显著差异。
  • 三种股骨固定方法刚度无显著差异(P=.089),均达到足够失效载荷。

核心结论

  • 三种股骨固定方法均达到足够的失效载荷,干扰螺钉失效载荷大于缝线锚钉,U形钉与两者均无显著差异。
  • 各固定方法间移植物刚度无显著差异。

研究设计与样本

  • 控制性实验室研究,采用18例新鲜冷冻尸体膝关节。
  • 标本经随机分组生成器随机分配至干扰螺钉、U形钉或缝线锚钉股骨固定组,行改良Lemaire外侧关节外腱固定术。
  • 以20 mm/min进行载荷至失败测试直至移植物失败,记录最大失效载荷、刚度和失败模式。

主要发现

  • 平均失效载荷以干扰螺钉最高(252.7±131.2 N),其次为U形钉(151.8±34.1 N)和缝线锚钉(105.7±16.4 N)。
  • 干扰螺钉与缝线锚钉间失效载荷差异有统计学意义(P=.015)。
  • U形钉与干扰螺钉(P=.101)及缝线锚钉(P=.577)间失效载荷均无显著差异。
  • 各固定方法间移植物刚度均无显著差异(P=.089)。

关键数值与推荐

  • 干扰螺钉平均失效载荷252.7±131.2 N,U形钉151.8±34.1 N,缝线锚钉105.7±16.4 N。
  • 干扰螺钉与缝线锚钉失效载荷比较P=.015;U形钉与干扰螺钉P=.101;U形钉与缝线锚钉P=.577。
  • 移植物刚度组间比较P=.089。

临床怎么用

  • 作者指出干扰螺钉股骨固定时最大失效载荷最高,但因隧道尺寸,该植入物相较直径更小的缝线锚钉钻孔可能面临更高的前交叉韧带重建隧道碰撞风险。
  • 作者指出缝线锚钉失效载荷最低,但基于文献报道的外侧关节外腱固定术移植物所承受的力,其似乎足以实现稳定固定。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: A lateral extra-articular tenodesis (LET) is increasingly being utilized to augment an anterior cruciate ligament reconstruction because it has been shown to reduce the risk of postreconstruction graft failure or recurrent rotatory instability. Various femoral fixation techniques are available, including the use of an interference screw, staple, or suture anchor. PURPOSE: To determine and compare the biomechanical properties of an LET graft when using an interference screw, staple, or suture anchor for the femoral fixation for a modified Lemaire LET. STUDY DESIGN: Controlled laboratory study. METHODS: Eighteen fresh-frozen cadaveric knees were obtained and randomly assigned via a random group generator to undergo a modified Lemaire LET using either an interference screw, a staple, or a suture anchor for femoral fixation. The specimen underwent load-to-failure testing at 20 mm/min until graft failure. The maximum failure load, stiffness, and failure mode for each specimen were recorded. RESULTS: The mean failure load was highest for the interference screw (252.7 ± 131.2 N), followed by the staple (151.8 ± 34.1 N) and the suture anchor (105.7 ± 16.4 N). There was a significant difference in failure load between the interference screw and the suture anchor (P = .015). There was no significant difference between the staple and the interference screw (P = .101) or the suture anchor (P = .577). There was no significant difference in graft stiffness across all fixation methods (P = .089). CONCLUSION: All 3 femoral fixation methods achieved adequate failure loads, although the interference screw had a greater failure load than the suture anchor and there was no significant difference between these implants and the staple. There were no significant differences in stiffness between the fixation methods. CLINICAL RELEVANCE: The maximum failure load occurred with an interference screw for femoral fixation of a modified Lemaire LET; however, because of socket size, this implant may be at greater risk of anterior cruciate ligament reconstruction tunnel collision compared to a smaller-diameter suture anchor drill hole. The failure load of the suture anchor was the lowest; however, it appears sufficient for stable fixation based on the force experienced by an LET graft reported in the literature.

来源:Am J Sports Med,查看原文。版权归原作者所有。

引用

  1. Tollefson LV, Shoemaker EP, Slette EL, et al. Adequate Failure Loads for Modified Lemaire Lateral Extra-articular Tenodesis Are Achieved With an Interference Screw, Staple, and Suture Anchor: A Biomechanical Study of Structural Properties. Am J Sports Med. 2025. PMID: 39760525. doi:10.1177/03635465241305739