双动髋系统的监测:损伤模式与临床数据分析

Surveillance of Dual-Mobility Hip Systems: Damage Mode and Clinical Data Analysis.
PubMed 文献 · J Arthroplasty · 2025-01-27

研究分析了51例双动全髋关节置换系统的取回植入物,评估宏观损伤和微观微动腐蚀。翻修指征包括机械并发症(27%)、感染(22%)和脱位(16%);平均植入时间12个月,与股骨头锥度腐蚀总和显著相关(P=0.044);螺钉-内衬腐蚀可能与骨溶解相关。

要点

  • 翻修指征以机械并发症(27%)、感染(22%)和脱位(16%)为主。
  • 平均植入12个月,股骨头锥度腐蚀总和与更长植入时间显著相关(P=0.044)。
  • 螺钉-内衬腐蚀存在时髋臼和股骨骨溶解发生率增加(P=0.019和P=0.022)。

核心结论

  • 本系列显示双动全髋关节置换组件在短期至中期(零至五年)植入后出现体内损伤,总体微动和腐蚀评分为轻度至中度。
  • 股骨头锥度腐蚀与更长植入时间相关,感染情况下观察到锥度腐蚀加重。
  • 螺钉-内衬腐蚀可能与髋臼和股骨骨溶解相关。
  • 本研究提示的趋势值得进一步评估。

研究设计与样本

  • 在植入物取回方案下,分析了51例双动全髋关节置换系统。
  • 对每个组件进行宏观损伤检查,并对锥度界面的微动腐蚀进行微观分级。
  • 从医疗记录中收集临床相关数据。

主要发现

  • 翻修指征包括机械并发症(n=15,27%)、感染(n=12,22%)和脱位(n=9,16%)。
  • 平均植入时间为12个月,与股骨头锥度腐蚀总和显著相关(P=0.044)。
  • 股骨头微动和腐蚀评分总和平均分别为2.9和3.0(两个区域,总和评分可能为2至8),锥度分别为7.6和7.1(四个象限,总和评分可能为4至16)。
  • 髋臼杯关节面最常见的损伤模式为划痕(n=26,65.4%),金属内衬为擦伤(n=29,55.2%),聚乙烯内衬为边缘变形(n=22,45.5%)。
  • 在可用的髋臼内衬中,10.3%(n=3)注意到螺钉-内衬腐蚀;当存在这种损伤模式时,髋臼和股骨骨溶解的发生率增加(分别为P=0.019和P=0.022)。

局限性

  • 本研究提示的趋势值得进一步评估。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Hip instability following total hip arthroplasty (THA) is among the most common indications for revision surgery. The implantation of dual-mobility (DM) systems, designed to improve stability, continues to rise, and thus, characterizing in vivo implant damage modes is paramount. METHODS: Under an implant retrieval protocol, 51 DM THA systems were analyzed. Each component was examined for macroscopic damage and microscopically graded for fretting corrosion at the trunnion interface. Clinically relevant data were collected from medical records. RESULTS: Revision indications included mechanical complications (n = 15, 27%), infection (n = 12, 22%), and dislocation (n = 9, 16%). The average duration of implantation was 12 months, which significantly correlated with summed femoral head taper corrosion (P = 0.044). Average summed fretting and corrosion scores were 2.9 and 3.0 for heads (two regions, 2 to 8 summed scores possible) and 7.6 and 7.1 for trunnions (four quadrants, 4 to 16 summed scores possible), respectively. Scratching (n = 26, 65.4%) was the most common damage mode on articular surfaces of acetabular cups, burnishing (n = 29, 55.2%) on metal liners, and edge deformation (n = 22, 45.5%) on polyethylene liners. Screw-liner corrosion was noted on 10.3% (n = 3) of available acetabular liners; an increased incidence of acetabular and femoral osteolysis was noted when this damage mode was present (P = 0.019 and P = 0.022, respectively). CONCLUSIONS: This series demonstrated in vivo damage of DM THA components following short-term-to-midterm (zero to five years) implantation, with overall mild-to-moderate fretting and corrosion scores. Femoral head taper corrosion correlated with longer implantation, and amplified trunnion corrosion was observed in the setting of infection. Screw-liner corrosion may be related to acetabular and femoral osteolysis. This study indicates trends that merit further evaluation.