常规与Attune S+胫骨底板全膝关节置换术后放射透亮线与翻修风险的比较
这项多中心回顾性观察研究纳入780例骨水泥型全膝关节置换患者,比较常规Attune底板与新型Attune S+底板术后放射透亮线(RLL)的发生率及翻修风险。新型底板独立关联较低的胫骨和股骨RLL发生率,但短期随访中底板设计与RLL对翻修风险无差异。
要点
- 新型Attune S+底板较常规底板独立关联更低的胫骨和股骨放射透亮线发生率。
- 中位随访14个月,胫骨RLL发生率为29%(228/777),股骨为15%(116/776)。
- 底板设计或RLL存在与否在短期随访中与翻修风险无差异。
核心结论
- 新型Attune胫骨底板较常规设计的整体、胫骨和股骨放射透亮线发生率更低,但仍高于前代设计和其他常用全膝关节置换系统。
研究设计与样本
- 多中心回顾性观察研究,共纳入780例在五个中心接受骨水泥型Attune膝关节系统全膝关节置换的患者,39%为男性,中位年龄70.7岁(IQR 62.0至77.2),最新放射影像用于评估术后6至36个月的放射透亮线。
主要发现
- 常规底板用于349例(45%),新型底板用于431例(55%)。
- 中位随访14个月(IQR 11至25),胫骨组件放射透亮线发生率为29%(228/777),股骨组件为15%(116/776),常规底板较新型底板更常见。
- 新型底板独立关联较低的胫骨和股骨放射透亮线发生率(均不受年龄、性别、BMI和影像时间影响)。
- 一年和三年翻修风险分别为1%(95%CI 0.4%至1.9%)和6%(95%CI 2.6%至13.2%)。
- 短期随访中底板设计与放射透亮线存在与否对翻修风险无差异。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIMS: This multicentre retrospective observational study's aims were to investigate whether there are differences in the occurrence of radiolucent lines (RLLs) following total knee arthroplasty (TKA) between the conventional Attune baseplate and its successor, the novel Attune S+, independent from other potentially influencing factors; and whether tibial baseplate design and presence of RLLs are associated with differing risk of revision. METHODS: A total of 780 patients (39% male; median age 70.7 years (IQR 62.0 to 77.2)) underwent cemented TKA using the Attune Knee System) at five centres, and with the latest radiograph available for the evaluation of RLL at between six and 36 months from surgery. Univariate and multivariate logistic regression models were performed to assess associations between patient and implant-associated factors on the presence of tibial and femoral RLLs. Differences in revision risk depending on RLLs and tibial baseplate design were investigated with the log-rank test. RESULTS: The conventional and novel Attune baseplates were used in 349 (45%) and 431 (55%) patients, respectively. At a median follow-up of 14 months (IQR 11 to 25), RLLs were present in 29% (n = 228/777) and 15% (n = 116/776) of the tibial and femoral components, respectively, and were more common in the conventional compared to the novel baseplate. The novel baseplate was independently associated with a lower incidence of tibial and femoral RLLs (both regardless of age, sex, BMI, and time to radiograph). One- and three-year revision risk was 1% (95% CI 0.4% to 1.9%)and 6% (95% CI 2.6% to 13.2%), respectively. There was no difference between baseplate design and the presence of RLLs on the the risk of revision at short-term follow-up. CONCLUSION: The overall incidence of RLLs, as well as the incidence of tibial and femoral RLLs, was lower with the novel compared to the conventional tibial Attune baseplate design, but higher than in the predecessor design and other commonly used TKA systems.
引用
- Smolle MA, Keintzel M, Staats K, et al. Radiolucent lines and revision risk in total knee arthroplasty using the conventional versus the Attune S+ tibial baseplate. Bone Joint J. 2024. PMID: 39481434. doi:10.1302/0301-620X.106B11.BJJ-2024-0084.R3
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。