移除断裂的组配式干骺端固定股骨柄需行扩大转子截骨术
本研究回顾了2002年9月至2023年5月连续取出的131例组配式干骺端固定(MME)股骨柄,其中10例(7.6%)因模块套筒近端柄断裂失败,断裂柄取出时9例行扩大转子截骨术(ETO)、1例行皮质开窗并发生术中股骨近端骨折;作者认为翻修断裂MME柄时应行ETO且截骨段需足够长以进入远端。
要点
- 131例取出MME柄中10例(7.6%)为模块套筒近端柄断裂。
- 断裂柄取出时9例行ETO,1例行皮质开窗并发生术中股骨近端骨折。
- 作者认为翻修断裂MME柄时应行ETO,截骨段需足够长以进入远端。
核心结论
- 断裂的MME股骨柄是翻修全髋关节置换术(THA)中骨科医生面临的挑战。
- 当柄在已骨长入的套筒近端发生断裂时,远端花键可能与骨发生骨性交锁(clothespin)。
- 作者认为翻修断裂MME柄时应行ETO,且截骨段应足够长以允许进入远端。
研究设计与样本
- 回顾性分析2002年9月至2023年5月连续取出的MME股骨柄,计算相关变量的描述性统计。
主要发现
- 共取出131例MME股骨柄,手术时平均年龄59岁(范围28至75岁),女性占49%,平均体重指数29.4(范围20.7至33.3),平均美国麻醉医师协会评分2.4±0.5。
- 131例中102例(78%)为设计A柄,29例(22%)为设计B柄。
- 131例中10例(7.6%)因模块套筒近端柄断裂而失败;设计A柄102例中4例(4%)断裂,设计B柄29例中6例(21%)断裂。
- 所有断裂柄在翻修THA取出时均需额外干预:9例行ETO,1例行皮质开窗并发生术中股骨近端骨折。
临床怎么用
- 作者认为翻修断裂MME柄时应行ETO,且截骨段应足够长以允许进入远端。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Modular metaphyseal engaging (MME) femoral components in total hip arthroplasty (THA) allow optimized femoral length, offset, and anteversion and are useful in patients with unusual proximal femoral anatomy. Fretting, corrosion, and stem fractures above the modular sleeve are complications associated with these implants. The purpose of this study was to identify failure mechanisms of retrieved MME femoral components at our institution, identify all broken stem cases, and evaluate how often an extended trochanteric osteotomy (ETO) was required for removal. METHODS: All consecutively retrieved MME femoral components from September 2002 to May 2023 were reviewed. Patient demographics, procedure information, component specifications, indications for removal, and requirements for further revision surgery were reviewed. Descriptive statistics were calculated for variables of interest. RESULTS: There were 131 retrieved MME components. The mean age at surgery was 59 years (range, 28 to 75), 49% were women, mean body mass index was 29.4 (range, 20.7 to 33.3), and mean American Society of Anesthesiologists score was 2.4 ± 0.5. There were 102 (78%) stems of 1 design (stem A), and 29 (22%) stems of a different design (stem B). Of 131 components, 10 (7.6%) failed secondary to stem fracture proximal to the modular sleeve. Four of 102 (4%) of stem A and 6 of 29 (21% of stem B) fractured. All broken stems required additional intervention for removal during revision THA, using an ETO (N = 9) or cortical window (N = 1) in which an intraoperative proximal femoral fracture occurred. CONCLUSIONS: Broken MME stems present a challenge for orthopaedic surgeons during revision THA. When a stem fracture occurs above the ingrown sleeve, the distal splines may have osseous interdigitation into the clothespin. Thus, when revising a broken MME stem, an ETO should be performed, and the segment should be long enough to allow distal access.
引用
- Ricotti RG, Schiller NC, Selkridge IK, et al. An Extended Trochanteric Osteotomy Is Necessary for the Removal of Broken Modular Metaphyseal Engaging Femoral Stems. J Arthroplasty. 2024. PMID: 38897261. doi:10.1016/j.arth.2024.06.018
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。