文献速递关节外科

第二代高交联聚乙烯胫骨衬垫的氧化与损伤机制

Oxidation and Damage Mechanisms of Second-Generation Highly Cross-Linked Polyethylene Tibial Inserts.
PubMed 文献 · J Arthroplasty · 2024-06-19

多中心翻修取出研究比较植入超过5年的74例胫骨衬垫(46例顺序退火高交联聚乙烯与28例常规伽马惰性灭菌超高分子量聚乙烯),两组翻修原因和表面损伤机制相当,但顺序退火高交联聚乙烯在关节面与前后面氧化更高,作者认为需进一步研究评估其临床意义。

要点

  • 两组最常见翻修原因均为松动(伽马惰性组17/28,顺序退火高交联组15/46)和不稳定(伽马惰性组6/28,顺序退火高交联组15/46)。
  • 最常见表面损伤机制为抛光、点蚀和划痕,伽马惰性组股骨髁抛光评分更高(P = .022)。
  • 顺序退火高交联聚乙烯在关节面(P = .002)与前后面(P = .023)平均氧化更高,背面未见差异(P = .060)。

核心结论

  • 伽马惰性灭菌与顺序退火高交联聚乙烯两组翻修原因和表面损伤机制相当。
  • 作者认为需进一步研究以持续评估顺序退火高交联聚乙烯在更长植入时间下氧化的体内损伤及可能的临床意义,尤其关注植入超过10年的假体。

研究设计与样本

  • 多中心取出研究,纳入74例植入超过5年(平均7±2年)的全膝关节置换胫骨衬垫,其中顺序退火高交联聚乙烯46例、常规伽马惰性灭菌超高分子量聚乙烯28例。
  • 后交叉韧带保留型假体在顺序退火高交联聚乙烯组占44%,在伽马惰性组占14%。
  • 从翻修手术记录中收集患者因素与翻修原因,采用半定量评分法评估表面损伤机制。
  • 按美国材料与试验协会2102标准,用傅里叶变换红外显微镜测量氧化。
  • 采用曼-惠特尼U检验比较两组差异。

主要发现

  • 松动为伽马惰性组17/28、顺序退火高交联组15/46,不稳定为伽马惰性组6/28、顺序退火高交联组15/46,为两组最常见翻修原因。
  • 最常见表面损伤机制为抛光、点蚀和划痕,伽马惰性组股骨髁抛光高于顺序退火高交联组(P = .022)。
  • 顺序退火高交联组平均氧化在关节面(P = .002)和前后面(P = .023)均高于伽马惰性组,背面未见差异(P = .060)。

关键数值与推荐

  • 松动:伽马惰性组17/28 vs 顺序退火高交联组15/46。
  • 不稳定:伽马惰性组6/28 vs 顺序退火高交联组15/46。
  • 股骨髁抛光评分伽马惰性组高于顺序退火高交联组(P = .022)。
  • 平均氧化顺序退火高交联组高于伽马惰性组:关节面P = .002,前后面P = .023,背面P = .060。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 作者指出需进一步研究以持续评估顺序退火高交联聚乙烯在更长植入时间下氧化的体内损伤及可能的临床意义,尤其关注植入超过10年的假体。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: After clinical introduction in 2005, sequentially annealed, highly cross-linked polyethylene (SA HXLPE) was studied for retrievals with short implantation times; however, long-term follow-ups are lacking. The objective of this study was to examine and compare the revision reasons, damage mechanisms, and oxidation indices of SA HXLPE and conventional gamma inert-sterilized (Gamma Inert) ultra-high-molecular-weight polyethylene tibial inserts implanted for >5 years. METHODS: There were 74 total knee arthroplasty tibial inserts (46 SA HXLPEs, 28 Gamma Inerts) implanted for >5 years (mean 7 ± 2 years) retrieved as part of a multicenter retrieval program. Cruciate-retaining implants comprised 44% of the SA HXLPEs and 14% of the Gamma Inerts. Patient factors and revision reasons were collected from revision operating notes. A semiquantitative scoring method was used to assess surface damage mechanisms. Oxidation was measured using Fourier transform infrared microscopy according to American Society for Testing and Materials 2102. Differences between cohorts were assessed with Mann-Whitney U-tests. RESULTS: Loosening (Gamma Inert: 17 of 28, SA HXLPE: 15 of 46) and instability (Gamma Inert: 6 of 28, SA HXLPE: 15 of 46) were the most common revision reasons for both cohorts. The most prevalent surface damage mechanisms were burnishing, pitting, and scratching, with burnishing of the condyles being higher in Gamma Inert components (P = .022). Mean oxidation was higher in the SA HXLPE inserts at the articulating surface (P = .002) and anterior-posterior faces (P = .023). No difference was observed at the backside surface (P = .060). CONCLUSIONS: Revision reasons and surface damage mechanisms were comparable in the Gamma Inert and SA cohorts. Further studies are needed to continue to assess the in vivo damage and clinical relevance, if any, of oxidation in SA HXLPE over longer implantation times, particularly for implants implanted for more than 10 years.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Derr T, MacDonald DW, Malkani AL, et al. Oxidation and Damage Mechanisms of Second-Generation Highly Cross-Linked Polyethylene Tibial Inserts. J Arthroplasty. 2024. PMID: 38906197. doi:10.1016/j.arth.2024.06.032

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。