开发计算机视觉模型以自动定量测量全髋与全膝关节置换患者的髋-膝-踝角

Developing a Computer Vision Model to Automate Quantitative Measurement of Hip-Knee-Ankle Angle in Total Hip and Knee Arthroplasty Patients.
PubMed 文献 · J Arthroplasty · 2024-04-27

本研究回顾性纳入学术医疗中心内计划行全髋或全膝关节置换患者的1379张全长下肢X线片,其中1221张用于开发基于检测的深度学习模型,158张因股骨头难以辨别被视为「困难」片。模型在测试集平均精度达0.985,手术侧与非手术侧髋-膝-踝角与人工测量的组内相关系数均为0.987,提示深度学习算法计算的髋-膝-踝角与人工测量相当。

要点

  • 基于检测的深度学习模型可自动计算全髋与全膝关节置换患者的髋-膝-踝角。
  • 测试集上模型平均精度为0.985。
  • 手术侧与非手术侧髋-膝-踝角人工与深度学习测量的组内相关系数均为0.987。

核心结论

  • 基于检测的深度学习算法可在全长下肢X线片上计算髋-膝-踝角,且与人工测量结果相当。
  • 该算法能高精度检测双侧股骨头、膝关节与踝关节,即使在股骨头难以辨认的患者中亦然。

研究设计与样本

  • 回顾性纳入学术医疗中心内计划行全髋或全膝关节置换患者的1379张全长下肢X线片,其中1221张用于开发模型,随机分为70%训练集、20%验证集与10%留出测试集。
  • 158张全长下肢X线片因股骨头难以与周围组织区分被视为「困难」片。
  • 由2名评估者对双下肢髋-膝-踝角进行标注,并在测试集中计算组内相关系数以比较深度学习推导的髋-膝-踝角与人工测量。

主要发现

  • 深度学习模型在测试集上平均精度为0.985。
  • 手术侧髋-膝-踝角平均为173.05 ± 4.54°,非手术侧为175.55 ± 3.56°。
  • 手术侧人工与深度学习测量髋-膝-踝角的组内相关系数(2,k)为0.987 [0.96, 0.99],非手术侧为0.987 [0.98, 0.99]。
  • 深度学习推导髋-膝-踝角在手术侧与非手术侧的测量标准误分别为0.515°与0.403°。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Increasing deformity of the lower extremities, as measured by the hip-knee-ankle angle (HKAA), is associated with poor patient outcomes after total hip and knee arthroplasty (THA, TKA). Automated calculation of HKAA is imperative to reduce the burden on orthopaedic surgeons. We proposed a detection-based deep learning (DL) model to calculate HKAA in THA and TKA patients and assessed the agreement between DL-derived HKAAs and manual measurement. METHODS: We retrospectively identified 1,379 long-leg radiographs (LLRs) from patients scheduled for THA or TKA within an academic medical center. There were 1,221 LLRs used to develop the model (randomly split into 70% training, 20% validation, and 10% held-out test sets); 158 LLRs were considered "difficult," as the femoral head was difficult to distinguish from surrounding tissue. There were 2 raters who annotated the HKAA of both lower extremities, and inter-rater reliability was calculated to compare the DL-derived HKAAs with manual measurement within the test set. RESULTS: The DL model achieved a mean average precision of 0.985 on the test set. The average HKAA of the operative leg was 173.05 ± 4.54°; the nonoperative leg was 175.55 ± 3.56°. The inter-rater reliability between manual and DL-derived HKAA measurements on the operative leg and nonoperative leg indicated excellent reliability (intraclass correlation (2,k) = 0.987 [0.96, 0.99], intraclass correlation (2, k) = 0.987 [0.98, 0.99, respectively]). The standard error of measurement for the DL-derived HKAA for the operative and nonoperative legs was 0.515° and 0.403°, respectively. CONCLUSIONS: A detection-based DL algorithm can calculate the HKAA in LLRs and is comparable to that calculated by manual measurement. The algorithm can detect the bilateral femoral head, knee, and ankle joints with high precision, even in patients where the femoral head is difficult to visualize.