非骨水泥翻修髋臼组件中骨水泥固定双动性假体的应用
本研究回顾性分析了64例(65髋)在翻修全髋关节置换术中,将单一设计骨水泥固定双动性杯骨水泥固定于非骨水泥髋臼壳内的患者,平均随访2年。结果显示无菌性双动性假体翻修和脱位的2年生存率均为92%,术后脱位5例均发生于既往脱位或外展肌缺陷患者,仅1例出现骨水泥界面失败。
要点
- 在非骨水泥髋臼壳内骨水泥固定双动性杯,2年无菌性翻修生存率92%。
- 术后脱位5例,均发生于既往脱位或外展肌缺陷患者。
- 影像学评估仅1例骨水泥界面出现失败。
核心结论
- 该技术在此复杂翻修全髋关节置换系列中未能消除脱位,但允许最大化股骨头直径并在骨水泥固定时优化髋臼组件有效位置。
- 值得注意的是,仅1例出现骨水泥界面失败。
研究设计与样本
- 回顾性识别2018年至2020年在本机构翻修全髋关节置换术中,将单一设计骨水泥固定双动性杯骨水泥固定于非骨水泥髋臼壳的64例患者(65髋)。
- 骨水泥固定于为内衬骨水泥固定设计的非骨水泥杯(n=48;74%)或保留的髋臼组件(n=17;26%)。
- 中位外头直径42 mm;平均年龄69岁(SD 11),平均BMI 32 kg/m2(SD 8),女性占52%(n=34)。
- 采用Kaplan-Meier法评估生存率,平均随访2年(SD 0.97)。
主要发现
- 共9例骨水泥固定双动性杯翻修:3例假体周围感染,3例髋臼无菌性松动(来自骨),2例脱位,1例杯-笼架构造断裂。
- 无菌性双动性假体翻修和脱位的2年生存率均为92%。
- 术后脱位5例,均发生于既往脱位或外展肌缺陷患者。
- 影像学复查显示,除1例外,所有病例双动性杯在骨水泥界面保持良好固定。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIMS: Dislocation remains a leading cause of failure following revision total hip arthroplasty (THA). While dual-mobility (DM) bearings have been shown to mitigate this risk, options are limited when retaining or implanting an uncemented shell without modular DM options. In these circumstances, a monoblock DM cup, designed for cementing, can be cemented into an uncemented acetabular shell. The goal of this study was to describe the implant survival, complications, and radiological outcomes of this construct. METHODS: We identified 64 patients (65 hips) who had a single-design cemented DM cup cemented into an uncemented acetabular shell during revision THA between 2018 and 2020 at our institution. Cups were cemented into either uncemented cups designed for liner cementing (n = 48; 74%) or retained (n = 17; 26%) acetabular components. Median outer head diameter was 42 mm. Mean age was 69 years (SD 11), mean BMI was 32 kg/m2 (SD 8), and 52% (n = 34) were female. Survival was assessed using Kaplan-Meier methods. Mean follow-up was two years (SD 0.97). RESULTS: There were nine cemented DM cup revisions: three for periprosthetic joint infection, three for acetabular aseptic loosening from bone, two for dislocation, and one for a broken cup-cage construct. The two-year survivals free of aseptic DM revision and dislocation were both 92%. There were five postoperative dislocations, all in patients with prior dislocation or abductor deficiency. On radiological review, the DM cup remained well-fixed at the cemented interface in all but one case. CONCLUSION: While dislocation was not eliminated in this series of complex revision THAs, this technique allowed for maximization of femoral head diameter and optimization of effective acetabular component position during cementing. Of note, there was only one failure at the cemented interface.
引用
- Wilson JM, Trousdale RT, Bedard NA, et al. Cemented dual-mobility constructs in uncemented revision acetabular components. Bone Joint J. 2024. PMID: 38555941. doi:10.1302/0301-620X.106B4.BJJ-2023-1061.R1
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。