新型非骨水泥羟基磷灰石涂层钛臼杯的迁移与临床结局:采用放射立体测量分析的随机对照试验两年随访
本研究将87例全髋关节置换术患者随机分配至Trident II HA或Trident HA非骨水泥臼杯组,采用放射立体测量分析(RSA)评估术后3、12、24个月假体迁移。两组近端迁移在24个月时均为0.14 mm(p=0.875),其他方向平移和旋转无显著差异,临床评分和患者报告结局可比,仅Trident HA组术后早期髋关节残疾和骨关节炎结局症状评分改善更明显(p=0.033)。结论认为Trident II HA臼杯迁移与前一版本相当,提示长期无菌性松动风险相似。
要点
- 24个月近端迁移:Trident II HA组0.14 mm vs Trident HA组0.14 mm(p=0.875)。
- 其他方向平移与旋转两组无显著差异。
- 仅Trident HA组术后早期髋关节残疾和骨关节炎结局症状评分改善更明显(p=0.033)。
核心结论
- Trident II clusterhole HA臼杯与前一版本Trident半球形HA cluster臼杯的迁移相当,提示长期无菌性松动风险相似。
研究设计与样本
- 随机对照试验,共87例接受全髋关节置换术的患者随机分配至Trident II HA或Trident HA非骨水泥臼杯组,两组臼杯均带cluster孔和HA涂层,均联合骨水泥Exeter V40股骨柄。
- 采用放射立体测量分析(RSA)测量假体迁移,于术后两天内(基线)及术后3、12、24个月拍摄X线片,主要结局为髋臼假体近端迁移。
主要发现
- 术后3、12、24个月近端迁移:Trident II HA组分别为0.08 mm(95%CI 0.03至0.14)、0.11 mm(95%CI 0.06至0.16)、0.14 mm(95%CI 0.09至0.20),Trident HA组分别为0.11 mm(95%CI 0.06至0.16)、0.12 mm(95%CI 0.07至0.17)、0.14 mm(95%CI 0.09至0.19)(p=0.875)。
- 任何其他方向的平移或旋转两组均无显著差异。
- 临床评分和患者报告结局两组可比,仅Trident HA组术后早期髋关节残疾和骨关节炎结局症状评分改善更明显(p=0.033)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIMS: The objective of this study was to compare the two-year migration and clinical outcomes of a new cementless hydroxyapatite (HA)-coated titanium acetabular shell with its previous version, which shared the same geometrical design but a different manufacturing process for applying the titanium surface. METHODS: Overall, 87 patients undergoing total hip arthroplasty (THA) were randomized to either a Trident II HA or Trident HA shell, each cementless with clusterholes and HA-coating. All components were used in combination with a cemented Exeter V40 femoral stem. Implant migration was measured using radiostereometric analysis (RSA), with radiographs taken within two days of surgery (baseline), and at three, 12, and 24 months postoperatively. Proximal acetabular component migration was the primary outcome measure. Clinical scores and patient-reported outcome measures (PROMs) were collected at each follow-up. RESULTS: Mean proximal migrations at three, 12, and 24 months were 0.08 mm (95% confidence interval (CI) 0.03 to 0.14), 0.11 mm (95% CI 0.06 to 0.16), and 0.14 mm (95% CI 0.09 to 0.20), respectively, in the Trident II HA group, versus 0.11 mm (95% CI 0.06 to 0.16), 0.12 mm (95% CI 0.07 to 0.17), and 0.14 mm (95% CI 0.09 to 0.19) in the Trident HA group (p = 0.875). No significant differences in translations or rotations between the two designs were found in any other direction. Clinical scores and PROMs were comparable between groups, except for an initially greater postoperative improvement in Hip disability and Osteoarthritis Outcome Symptoms score in the Trident HA group (p = 0.033). CONCLUSION: The Trident II clusterhole HA shell has comparable migration with its predecessor, the Trident hemispherical HA cluster shell, suggesting a similar risk of long-term aseptic loosening.
引用
- van der Lelij TJN, Marang-van de Mheen PJ, Kaptein BL, et al. Migration and clinical outcomes of a novel cementless hydroxyapatite-coated titanium acetabular shell: two-year follow-up of a randomized controlled trial using radiostereometric analysis. Bone Joint J. 2024. PMID: 38295835. doi:10.1302/0301-620X.106B2.BJJ-2023-0862.R1
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。