翻修髋关节柄Prevision模块化锥部骨折分析及原设计与现行设计的比较
该回顾性长期分析利用制造商警戒数据库,调查2004年1月至2022年12月Prevision模块化翻修髋关节柄的机械失败模式。识别出远端柄骨折(失败率0.07%)和模块化锥部骨折(原设计1.74%,现行设计0.013%)。结论认为两类骨折风险均较低,且设计改进显著降低了模块化骨折风险。
要点
- Prevision翻修柄远端柄骨折失败率为0.07%。
- 模块化锥部骨折率原设计为1.74%,现行设计为0.013%。
- 设计改进显著降低了模块化锥部骨折风险。
核心结论
- 与翻修髋关节置换术其他已知并发症相比,两种骨折类型的机械失败风险均较低。
- 数据表明设计改进显著降低了模块化锥部骨折的风险。
研究设计与样本
- 回顾性长期分析,利用制造商警戒数据库调查2004年1月至2022年12月Prevision模块化髋关节柄的机械失败模式。
- 2008年引入了模块化锥部的设计改进,数据用于比较原设计与现行设计的模块化连接。
主要发现
- 识别出两种机械失败模式:远端开槽区域骨折(远端柄骨折)和模块化锥部失败(模块化骨折)。
- 远端柄骨折的失败率为0.07%。
- 模块化骨折率原设计为1.74%,现行设计为0.013%。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIMS: The risk of mechanical failure of modular revision hip stems is frequently mentioned in the literature, but little is currently known about the actual clinical failure rates of this type of prosthesis. The current retrospective long-term analysis examines the distal and modular failure patterns of the Prevision hip stem from 18 years of clinical use. A design improvement of the modular taper was introduced in 2008, and the data could also be used to compare the original and the current design of the modular connection. METHODS: We performed an analysis of the Prevision modular hip stem using the manufacturer's vigilance database and investigated different mechanical failure patterns of the hip stem from January 2004 to December 2022. RESULTS: Two mechanical failure patterns were identified: fractures in the area of the distal fluted profile (distal stem fracture) and failure of the modular taper (modular fracture). A failure rate of 0.07% was observed for distal stem fracture, and modular fracture rates of 1.74% for the original and 0.013% for the current taper design. CONCLUSION: A low risk of mechanical failure for both fracture types was observed compared to other known complications in revision hip arthroplasty. In addition, the data show that a design change did significantly reduce the risk of a modular fracture.
引用
- Dreyer L, Bader C, Flörkemeier T, et al. Analysis of modular taper fractures of the revision hip stem Prevision and comparison of the original and current taper design. Bone Joint J. 2024. PMID: 38295827. doi:10.1302/0301-620X.106B2.BJJ-2023-0605.R1
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。