薄型维生素E掺杂交联聚乙烯内衬中使用大号金属球头不增加全髋关节置换术中的聚乙烯磨损:随机对照试验5年结果

The Use of Large Metal Heads in Thin Vitamin E-Doped Cross-Linked Polyethylene Inserts Does Not Increase Polyethylene Wear in Total Hip Arthroplasty: 5-Year Results From a Randomized Controlled Trial.
PubMed 文献 · J Arthroplasty · 2024-02-11

该随机对照试验将96例非骨水泥全髋关节置换术候选者随机分配至最大可适配薄型维生素E掺杂交联聚乙烯(VEPE)内衬的金属球头(36至44 mm,干预组)或标准32 mm金属球头(对照组),5年中期结果显示两组近端球头穿透量中位数无显著差异(P=0.691),髋臼周围透亮线发生率无显著差异(P=0.197),患者报告结局无差异但干预组活动水平更高(P=0.020)。

要点

  • 薄型VEPE内衬中使用大号金属球头5年未增加内衬磨损。
  • 两组髋臼周围透亮线发生率无显著差异(P=0.197)。
  • 干预组University of California活动水平评分中位数7分高于对照组6分(P=0.020)。

核心结论

  • 薄型VEPE内衬中使用大号金属球头未增加内衬磨损,且术后5年与内衬失败无关。

研究设计与样本

  • 随机对照试验,96例非骨水泥全髋关节置换术候选者随机分配至干预组(最大可适配薄型VEPE内衬的金属球头,36至44 mm)或对照组(标准32 mm金属球头),报告5年中期结果。

主要发现

  • 近端球头总穿透量中位数(四分位距)干预组为-0.04 mm(-0.12至0.02),对照组为-0.03 mm(-0.14至0.05),P=0.691。
  • 髋臼周围透亮线发生率干预组为44例中1例,对照组为42例中4例,P=0.197。
  • 患者报告的髋关节功能和健康相关生活质量两组间无差异,但干预组报告更高的活动水平(University of California活动水平评分中位数7分 vs 6分,P=0.020)。
  • 共有5例翻修,原因分别为脱位(2例)、假体周围骨折(1例)、股骨柄下沉(1例)或髂腰肌撞击(1例)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Vitamin E-doped cross-linked polyethylene (VEPE) has encouraged the use of larger heads in thinner liners in total hip arthroplasty (THA). However, there are concerns about wear and mechanical failure of the thin liner, especially when metal heads are used. The aim of this randomized controlled trial was to investigate if the use of a large metal head in thin VEPE liner would increase polyethylene wear compared with a standard 32-mm metal head and to compare periacetabular radiolucencies and patient-reported outcomes in THA. METHODS: There were 96 candidates for uncemented THA who were randomly allocated to either the largest possible metal head (36 to 44 mm) that could be fitted in the thinnest available VEPE liner (intervention group) or a standard 32-mm metal head (control group). The primary outcome was proximal head penetration, measured with a model-based radiostereometric analysis. Secondary outcomes were periacetabular radiolucencies and patient-reported outcomes. The midterm results of the trial at 5 years are presented. RESULTS: The median total proximal head penetration (interquartile range) was -0.04 mm (-0.12 to 0.02) in the intervention group and -0.03 mm (-0.14 to 0.05) in the control group (P = .691). The rates of periacetabular radiolucencies were 1 of 44 and 4 of 42 (P = .197), respectively. Patient-reported hip function and health-related quality of life did not differ between the groups, but participants in the intervention group reported a higher level of activity (median University of California Level of Activity score 7 versus 6, P = .020). There were 5 revisions caused by dislocations (2), periprosthetic fracture (1), stem subsidence (1), or iliopsoas impingement (1). CONCLUSIONS: Large metal heads in thin VEPE liners did not increase liner wear and were not associated with liner failure 5 years after THA.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Tsikandylakis G, Mortensen KRL, Gromov K, et al. The Use of Large Metal Heads in Thin Vitamin E-Doped Cross-Linked Polyethylene Inserts Does Not Increase Polyethylene Wear in Total Hip Arthroplasty: 5-Year Results From a Randomized Controlled Trial. J Arthroplasty. 2024. PMID: 38350518. doi:10.1016/j.arth.2024.02.008