初次金属对聚乙烯全髋关节置换术后因锥度腐蚀行翻修术的临床结局与再翻修危险因素

Clinical Outcomes and Risk Factors for Re-Revision Due to Trunnion Corrosion in Primary Metal-on-Polyethylene Total Hip Arthroplasty.
PubMed 文献 · J Arthroplasty · 2024-02-08

回顾性分析80髋(79例)金属对聚乙烯全髋关节置换术后因锥度腐蚀行翻修术的患者,平均随访4.6年。再翻修率26%,2年和5年无再翻修生存率分别为75.0%和73.2%。未行髋臼杯翻修与初次置换至锥度翻修时间较短是再翻修的独立危险因素。

要点

  • 因锥度腐蚀行翻修术后再翻修率高(26%),最常见原因为不稳和感染。
  • 未行髋臼杯翻修者再翻修风险增高(OR 4.5,95%CI 1.03-19.7)。
  • 初次全髋置换至锥度翻修时间较短与再翻修风险相关(OR 0.77,95%CI 0.62-0.97)。

核心结论

  • 因锥度腐蚀行翻修术后早期再翻修风险高(26%),主要原因为感染和不稳,功能结局尚可;未行髋臼杯翻修及初次置换至锥度翻修时间较短是再翻修的独立危险因素。

研究设计与样本

  • 回顾性分析80髋(79例)因金属对聚乙烯全髋关节置换术后锥度腐蚀行翻修术的患者,平均随访4.6年(范围2.0-9.4年),采用Kaplan-Meier生存分析和多因素logistic回归。

主要发现

  • 21髋(26%)在锥度翻修术后平均8.0个月(范围0.03-36.3个月)行再翻修,最常见原因为不稳和感染;2年和5年无再翻修生存率分别为75.0%和73.2%;平均牛津髋关节评分33.7(范围11-48);76%满意,24%不满意。

关键数值与推荐

  • 多因素分析显示未行髋臼杯翻修者再翻修风险增高(OR 4.5,95%CI 1.03-19.7),初次置换至锥度翻修时间较长者再翻修风险降低(OR 0.77,95%CI 0.62-0.97)。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: There is a paucity of literature regarding the mid-term (greater than 2 years) outcomes of revision for adverse local tissue reaction to metal debris due to corrosion at the head-neck junction (trunnionosis) in metal-on-polyethylene total hip arthroplasty (THA), and risk factors for re-revision remain largely unknown. We aimed to report the re-revision-free survival and functional outcomes for this patient population and to identify risk factors for re-revision. METHODS: A total of 80 hips (79 patients) with a metal-on-polyethylene THA who had undergone revision for trunnionosis at our institution were included. The mean study follow-up from index trunnionosis revision was 4.6 years (range, 2.0 to 9.4). Kaplan-Meier survival analysis was performed with all-cause re-revision as the end point, and multivariate logistic regression was used to identify risk factors for re-revision. RESULTS: We saw that twenty-one hips (26%) underwent re-revision at a mean of 8.0 months (range, 0.03 to 36.3) after the index trunnionosis revision, most commonly for instability and infection. The two- and five-year all-cause re-revision-free survival rates were 75.0 and 73.2%, respectively. The mean Oxford Hip Score was 33.7 (range, 11 to 48); 76% were satisfied, and 24% were dissatisfied with their hip. Multivariate analysis identified not undergoing a cup revision (odds ratio: 4.5; 95% confidence interval: 1.03 to 19.7) and time from primary THA to the index trunnionosis revision (odds ratio: 0.77; 95% confidence interval: 0.62 to 0.97) as risk factors for undergoing re-revision. CONCLUSIONS: The risk of early re-revision for these patients is high (26%), mostly due to infection and instability, and functional outcomes are fair. Not performing a cup revision appears to be a risk factor for re-revision, as is the shorter time from primary THA to trunnionosis revision. LEVEL OF EVIDENCE: III.