股骨假体设计与65岁及以上患者非骨水泥全髋关节置换术后假体周围股骨骨折风险相关
基于美国关节置换登记系统2012年至2020年3月的数据,分析266,040例65岁及以上骨关节炎患者初次非骨水泥全髋关节置换术。调整年龄、性别、地区、骨质疏松诊断、医院手术量和死亡竞争风险后,单楔形和双楔形股骨柄假体周围股骨骨折风险均高于渐变锥形/干骺端填充型(HR 2.9和3.0),无领假体风险高于有领假体(HR 7.8)。作者认为,若在此人群使用非骨水泥固定,可考虑渐变锥形/干骺端填充型和有领股骨柄设计,因其与较低的假体周围股骨骨折风险相关。
要点
- 在65岁及以上非骨水泥全髋关节置换患者中,单楔形股骨柄假体周围股骨骨折风险高于渐变锥形/干骺端填充型(HR 2.9)。
- 双楔形股骨柄假体周围股骨骨折风险也高于渐变锥形/干骺端填充型(HR 3.0)。
- 无领股骨柄假体周围股骨骨折风险高于有领股骨柄设计(HR 7.8)。
核心结论
- 若65岁及以上患者全髋关节置换术使用非骨水泥股骨固定,作者认为可考虑使用渐变锥形/干骺端填充型和有领股骨柄设计,因其与较低的假体周围股骨骨折风险相关。
- 未来研究应在此人群中比较渐变锥形/干骺端填充型和有领非骨水泥设计与骨水泥固定。
研究设计与样本
- 研究分析美国关节置换登记系统(American Joint Replacement Registry)2012年至2020年3月的数据,纳入266,040例65岁及以上骨关节炎患者初次非骨水泥全髋关节置换术,并与补充的联邦医疗保险和医疗补助服务中心(Centers for Medicare and Medicaid Services)数据链接。
- 主要分析按Kheir和Chen分类比较股骨柄几何形状,共232,972例初次非骨水泥全髋关节置换术纳入分析:单楔形42%(112,231/266,040)、双楔形22%(57,758/266,040)、渐变锥形/干骺端填充型24%(62,983/266,040)。
- 附加分析比较有领非骨水泥股骨柄(20% [47,376/232,972])与无领设计(80% [185,596/232,972])。
- 采用Cox比例风险回归分析并考虑死亡竞争风险,评估假体设计与骨折风险的关联,同时调整潜在混杂因素。
- 证据等级为III级,治疗性研究。
临床怎么用
- 作者认为,若65岁及以上患者全髋关节置换术使用非骨水泥股骨固定,可考虑使用渐变锥形/干骺端填充型和有领股骨柄设计,因其与较低的假体周围股骨骨折风险相关。
- 作者提出,未来研究应在此人群中比较渐变锥形/干骺端填充型和有领非骨水泥设计与骨水泥固定。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Periprosthetic femur fracture is a known complication after THA. The associated risk of cementless femoral component design for periprosthetic femur fracture in a registry population of patients older than 65 years has yet to be clearly identified. QUESTIONS/PURPOSES: (1) Is femoral stem geometry associated with the risk of periprosthetic femur fracture after cementless THA? (2) Is the presence or absence of a collar on cementless femoral implant designs associated with the risk of periprosthetic femur fracture after THA? METHODS: We analyzed American Joint Replacement Registry data from 2012 to March 2020. Unique to this registry is the high use of cementless femoral stems in patients 65 years and older. We identified 266,040 primary cementless THAs during the study period in patients with a diagnosis of osteoarthritis and surgeries linked to supplemental Centers for Medicare and Medicaid data where available. Patient demographics, procedure dates, and reoperation for periprosthetic femur fracture with revision or open reduction and internal fixation were recorded. The main analysis was performed comparing the Kheir and Chen classification: 42% (112,231 of 266,040) were single-wedge, 22% (57,758 of 266,040) were double-wedge, and 24% (62,983 of 266,040) were gradual taper/metadiaphyseal-filling cementless femoral components, which yielded a total of 232,972 primary cementless THAs. An additional analysis compared cementless stems with collars (20% [47,376 of 232,972]) with those with collarless designs (80% [185,596 of 232,972]). A Cox proportional hazard regression analysis with the competing risk of death was used to evaluate the association of design and fracture risk while adjusting for potential confounders. RESULTS: After controlling for the potentially confounding variables of age, sex, geographic region, osteoporosis or osteopenia diagnosis, hospital volume, and the competing risk of death, we found that compared with gradual taper/metadiaphyseal-filling stems, single-wedge designs were associated with a greater risk of periprosthetic femur fracture (HR 2.9 [95% confidence interval (CI) 2.2 to 3.9]; p < 0. 001). Compared with gradual taper/metadiaphyseal-filling stems, double-wedge designs showed an increased risk of periprosthetic femur fracture (HR 3.0 [95% CI 2.2 to 4.0]; p < 0. 001). Collarless stems showed an increased risk of periprosthetic fracture compared with collared stems (HR 7.8 [95% CI 4.1 to 15]; p < 0. 001). CONCLUSION: If cementless femoral fixation is used for THA in patients 65 years or older, surgeons should consider using gradual taper/metadiaphyseal-filling and collared stem designs because they are associated with a lower risk of periprosthetic femur fracture. Future investigations should compare gradual taper/metadiaphyseal-filling and collared cementless designs with cemented fixation in this population. LEVEL OF EVIDENCE: Level III, therapeutic study.