20年经验中2141例腰椎全椎间盘置换装置的取出与翻修
该回顾性研究纳入2141例腰椎全椎间盘置换(TDR)患者,平均随访78.6个月,27例(1.26%)接受装置取出或翻修。取出24例(1.12%),翻修3例(0.14%),37.0%发生在植入后1个月内,40.7%发生在个别术者最初25例TDR中。作者认为20年期间的低取出/翻修率支持这些装置的安全性。
要点
- 2141例腰椎TDR中27例(1.26%)接受取出或翻修,其中取出24例(1.12%)、翻修3例(0.14%)。
- 37.0%的取出/翻修发生在植入后1个月内,40.7%发生在个别术者最初25例TDR中。
- 258例随访≥15年的患者中仅1例接受取出/翻修,该例因创伤取出并发生1例显著血管并发症。
核心结论
- 在这组大型连续病例中,1.26%的TDR被取出或翻修,20年期间的低发生率支持这些装置的安全性。
研究设计与样本
- 回顾性研究,尝试进行前瞻性患者联系以收集当前数据,纳入自2000年首例开始的2141例腰椎TDR患者,平均随访78.6个月。
主要发现
- 2141例患者中27例(1.26%)接受TDR取出或翻修,其中取出24例(1.12%)、翻修3例(0.14%)。
- 24例取出原因:12例为迁移和/或松动,3例创伤后出现问题,2例对装置材料发生淋巴细胞反应,2例持续疼痛,TDR过大、椎体骨折(骨质疏松)、溶骨性病变、装置下沉合并小关节骨关节炎、TDR后146个月胸部感染播散各1例。
- 3例翻修分别为Core重新定位(技术错误)、装置移位后重新定位、因磨损/失效更换Core。
- 37.0%的取出/翻修发生在植入后1个月内,40.7%发生在个别术者最初25例TDR中。
- 1例因创伤取出TDR的患者发生1例显著血管并发症,该例也是258例随访≥15年患者中唯一接受取出/翻修者。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
STUDY DESIGN: This was a retrospective study with prospective patient contact attempted to collect current data. OBJECTIVE: The purpose was to investigate the incidence and reasons for lumbar total disk replacement (TDR) removal or revision. SUMMARY OF BACKGROUND DATA: A concern regarding lumbar TDR was safety, particularly the need for device removal or revision. This may be particularly important considering removal/revision requires repeat anterior exposure with an increased risk of vascular injury. METHODS: Data were collected for a series of 2141 lumbar TDR patients, beginning with the first case experience in 2000. The mean follow-up was 78.6 months. For each case of device removal/revision, the reason, duration from index surgery, and procedure performed were recorded. RESULTS: Of 2141 patients, 27 (1.26%) underwent TDR removal or revision. Device removal was performed in 24 patients (1.12%), while three patients underwent revision (0.14%). Of the 24 removals, 12 were due to migration and/or loosening, three developed problems post-trauma, two developed lymphocytic reaction to device materials, two had ongoing pain, and there was one case of each: TDR was too large, vertebral body fracture (osteoporosis), lytic lesion, device subsidence and facet arthrosis, and infection seeded from a chest infection 146 months post-TDR. The three revisions were for Core repositioning (technique error), device repositioning after displacement, and core replacement due to wear/failure. With respect to timing, 37.0% of removals/revisions occurred within one-month postimplantation. Of note, 40.7% of removals/revisions occurred in the first 25 TDR cases performed by individual surgeons. There was one significant vascular complication occurring in a patient whose TDR was removed due to trauma. This was also the only patient among 258 with ≥15-year follow-up who underwent removal/revision. CONCLUSION: In this large consecutive series, 1.26% of TDRs were removed/revised. The low rate over a 20 year period supports the safety of these devices.
引用
- Guyer RD, Blumenthal SL, Shellock JL, et al. Lumbar Total Disk Replacement Device Removals and Revisions Performed During a 20-Year Experience with 2141 Patients. Spine (Phila Pa 1976). 2024. PMID: 38282440. doi:10.1097/BRS.0000000000004942
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。