单块陶瓷髋臼杯翻修率更低且稳定性更佳
基于澳大利亚骨科协会国家关节置换登记处2001年1月1日至2021年12月31日486,946例初次常规全髋关节置换术数据,其中4,033例使用DeltaMotion单块陶瓷杯、482,913例使用模块化设计。DeltaMotion杯在所有诊断中全因翻修累积百分比最低,假体脱位翻修率显著更低,无因异响翻修,陶瓷破损1例而模块化组175例。
要点
- DeltaMotion杯全因翻修累积百分比在所有时间点均低于模块化轴承。
- DeltaMotion杯假体脱位翻修率显著低于模块化轴承,且无因异响翻修。
- DeltaMotion组陶瓷破损1例,模块化轴承组记录175例陶瓷破损。
核心结论
- DeltaMotion杯的全因翻修率低,因脱位、陶瓷破损和异响的翻修率也低。
- 尽管该杯已不再生产,对更新型单块陶瓷杯的持续随访将确定其是否适合年轻和更活跃的患者。
研究设计与样本
- 分析澳大利亚骨科协会国家关节置换登记处(AOANJRR)2001年1月1日至2021年12月31日期间所有接受初次常规全髋关节置换术的患者数据。
- 仅纳入陶瓷/陶瓷、陶瓷/交联聚乙烯(XLPE)、金属/XLPE或陶瓷金属/XLPE轴承表面的假体。
- 主要结局指标为全因翻修累积百分比,次要结局指标为脱位/不稳、陶瓷破损或异响的翻修,并进行杯尺寸的亚组分析。
- 共486,946例初次常规全髋关节置换术,其中4,033例使用DeltaMotion杯,482,913例使用模块化设计。
主要发现
- DeltaMotion杯在所有诊断中全因翻修累积百分比在所有时间点均低于模块化轴承。
- DeltaMotion杯假体脱位翻修率显著低于其他模块化轴承,且与其他模块化轴承相比无因异响翻修。
- 模块化轴承组记录175例陶瓷破损,DeltaMotion组记录1例陶瓷破损。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Monoblock ceramic cups are designed to accommodate large-diameter femoral heads. This has the potential to offer the advantages of an increased range of motion and enhanced joint stability. These features could benefit younger and high-demand patients in need of total hip arthroplasty. The aim of this study was to assess the survival rate and the reasons for revision of the DeltaMotion cup. METHODS: Data from the AOANJRR were analyzed for all patients who had undergone a primary conventional THA performed between January 1, 2001 and December 31, 2021. Only prostheses with ceramic/ceramic, ceramic/XLPE, metal/XLPE, or CM/XLPE bearing surfaces were included. The primary outcome measure was the cumulative percent revision for all causes. Secondary outcome measures were revision for dislocation/instability, ceramic breakage, or noise. A subanalysis for cup size was also performed. RESULTS: There were 486,946 primary conventional THA procedures undertaken for any reason. Of these, 4,033 used the DeltaMotion cup and 482,913 were modular designs. The DeltaMotion cup had the lowest CPR for all diagnoses compared to the modular bearings at all time points, had a significantly lower revision rate for prosthesis dislocation and no revisions for squeaking compared to other modular bearings. There were 175 ceramic breakages recorded in the modular bearing group and 1 ceramic breakage in the DeltaMotion group. CONCLUSIONS: The DeltaMotion cup had a low rate of all-cause revision, and for dislocation, ceramic breakage, and noise. Although this cup is no longer manufactured, ongoing follow-up of newer monoblock ceramic cups will determine their suitability for younger and more active patients.
引用
- Vercruysse LYG, Milne LP, Harries DTC, et al. Lower Revision Rates and Improved Stability With a Monoblock Ceramic Acetabular Cup. J Arthroplasty. 2023. PMID: 37871861. doi:10.1016/j.arth.2023.10.032
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。