部分肩胛盂关节置换术与Latarjet手术治疗肩胛盂前部骨缺损的尸体生物力学研究

Cadaveric Biomechanical Study of Partial Glenoid Arthroplasty Versus the Latarjet Procedure for Anterior Glenoid Bone Loss.
PubMed 文献 · Am J Sports Med · 2023-09-16

研究在14对匹配的尸体肩关节上比较新型3D打印部分肩胛盂关节置换假体(PGA)与经典Latarjet手术。PGA术后关节面面积更大、台阶更小,肩胛骨平面外旋和总旋转活动度更接近完整状态,肱骨头顶点轨迹更接近完整状态。PGA在抵抗前移方面与Latarjet无显著差异,但在2mm结构位移时线性刚度和载荷显著更高。

要点

  • PGA术后关节面面积显著大于Latarjet组(P=.006),关节台阶显著更小(P=.030)。
  • PGA在肩胛骨平面外旋(P=.006)和总旋转活动度(P=.019)上更接近完整状态。
  • PGA组在2mm结构位移时线性刚度(P=.031)和载荷(P=.002)显著高于Latarjet组。

核心结论

  • 在尸体模型中,PGA较Latarjet手术更好地接近完整盂肱关节运动学,关节台阶更小;PGA在抵抗前移方面与Latarjet相当,在2mm结构位移时抗载荷显著更强;作者认为需要进一步临床研究来验证这一新术式。

研究设计与样本

  • 控制性实验室研究,共测试14对匹配的尸体肩关节;PGA假体基于术前CT用钛3D打印;测试完整状态、25%肩胛盂前部骨缺损状态和术后状态,在肩胛骨平面和冠状面进行。

主要发现

  • 两组肩胛盂关节面基线尺寸相当。
  • PGA术后关节面面积显著更大(P=.006),关节台阶显著更小(P=.030)。
  • PGA在肩胛骨平面外旋(P=.006)和总旋转活动度(P=.019)上更好地接近完整状态。
  • PGA术后肱骨头顶点轨迹更好地跟随完整状态(P<.001)。
  • PGA术后抵抗前移与Latarjet术后无显著差异;PGA组在2mm结构位移时线性刚度(P=.031)和载荷(P=.002)更大。

临床怎么用

  • 作者认为,3D打印PGA假体可能为肩关节不稳所致严重肩胛盂骨缺损提供一种替代治疗选择,克服Latarjet手术既往的运动学改变、固定失败和硬件问题等缺点。

局限性

  • 作者指出需要进一步临床研究来验证这一新术式。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: For severe anterior glenoid bone loss due to recurrent shoulder instability, the Latarjet procedure offers a dynamic sling effect in addition to bone augmentation. Yet, it heavily alters the surrounding anatomy, while fixation and graft union issues are also common. PURPOSE/HYPOTHESIS: The purpose of this study was to compare a novel printed 3-dimensional (3D) partial glenoid arthroplasty (PGA) implant with the classic Latarjet procedure. It was hypothesized that by replicating the original glenoid geometry and preserving soft tissue anatomy, PGA may better reproduce normal joint kinematics. In addition, the locking screw construct may offer stronger fixation. STUDY DESIGN: Controlled laboratory study. METHODS: A total of 14 matched cadaveric shoulders were tested. The PGA implant was 3D printed in titanium based on preoperative computed tomography. The intact, 25% anterior glenoid bone loss, and postoperative states were tested in the scapular and coronal planes. The following parameters were measured: articular surface area and stepoff, rotational range of motion and the humeral head apex position during rotation, and load and linear stiffness at 25% anterior translation and at 2-mm construct displacement. RESULTS: The baseline dimensions of the glenoid articular surface were comparable between the groups. The articular surface area after PGA was significantly larger (P = .006) with less articular stepoff (P = .030). PGA better approximated the intact state's external (P = .006) and total (P = .019) rotational range of motion in the scapular plane. The course of the humeral head apex after PGA better followed that of the intact state (P < .001). Resistance against anterior translation after PGA was not significantly different compared with after the Latarjet procedure. Greater linear stiffness (P = .031) and loading (P = .002) at 2-mm construct displacement were demonstrated in the PGA group. CONCLUSION: In addressing anterior glenoid bone loss, PGA better approximated intact glenohumeral joint kinematics compared with the Latarjet procedure with less articular stepoff in a cadaveric model. PGA was comparable in resisting anterior translation while being significantly stronger against loading at 2-mm construct displacement. Further clinical studies are warranted to validate this novel procedure. CLINICAL RELEVANCE: A 3D-printed PGA implant may offer an alternative treatment option for severe glenoid bone loss due to shoulder instability, overcoming the previous drawbacks of the Latarjet procedure, including altered kinematics, fixation failure, and hardware issues.

来源:Am J Sports Med,查看原文。版权归原作者所有。