文献速递关节外科

非骨水泥3D打印全膝关节假体的持续稳定:一项随机对照试验的五年放射立体分析结果

Continued Stabilization of a Cementless 3D-Printed Total Knee Arthroplasty: Five-Year Results of a Randomized Controlled Trial Using Radiostereometric Analysis.
PubMed 文献 · J Bone Joint Surg Am · 2023-08-31

72例患者随机接受非骨水泥3D打印Triathlon Tritanium或骨水泥Triathlon全膝关节置换,随访5年。5年时两组胫骨假体最大总点运动均值无显著差异(0.66 mm vs 0.53 mm,p=0.09);骨水泥组在2年后仍有迁移进展,非骨水泥组早期迁移后保持稳定。临床评分两组相当。

要点

  • 5年时非骨水泥组与骨水泥组胫骨假体MTPM均值无显著差异(0.66 mm vs 0.53 mm,p=0.09)。
  • 非骨水泥组2至5年MTPM无进展(0.02 mm),骨水泥组有进展(0.07 mm)。
  • 非骨水泥组1例、骨水泥组4例假体持续迁移;临床评分两组全程相当。

核心结论

  • 5年时非骨水泥与骨水泥全膝关节假体的平均迁移无显著差异。
  • 骨水泥假体在术后2年后仍存在胫骨假体迁移进展,而非骨水泥假体在早期迁移后保持稳定。

研究设计与样本

  • 随机对照试验,72例患者随机接受非骨水泥3D打印Triathlon Tritanium后交叉韧带保留型全膝关节置换或骨水泥Triathlon后交叉韧带保留型全膝关节置换。
  • 采用放射立体分析在基线及术后3个月、1年、2年、5年评估假体迁移。
  • 比较5年时胫骨假体最大总点运动,并评估2至5年间的迁移进展;临床评分采用线性混合效应模型分析重复测量。

主要发现

  • 5年时平均MTPM非骨水泥组为0.66 mm(95%CI 0.56至0.78 mm),骨水泥组为0.53 mm(95%CI 0.43至0.64 mm),p=0.09。
  • 2至5年间,非骨水泥组平均MTPM无进展(0.02 mm;95%CI -0.06至0.10 mm),骨水泥组为0.07 mm(95%CI 0.00至0.14 mm)。
  • 非骨水泥组有1例假体持续迁移,骨水泥组有4例持续迁移。
  • 临床评分在整个随访期间两组相当。

关键数值与推荐

  • 假体持续迁移定义为术后2年后MTPM≥0.1 mm/年。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Three-dimensional (3D) printing of highly porous orthopaedic implants aims to promote better osseointegration, thus preventing aseptic loosening. However, short-term radiostereometric analysis (RSA) after total knee arthroplasty (TKA) has shown higher initial migration of cementless 3D-printed tibial components compared with their cemented counterparts. Therefore, critical evaluation of longer-term tibial component migration is needed. We investigated migration of a cementless 3D-printed and a cemented tibial component with otherwise similar TKA design during 5 years of follow-up, particularly the progression in migration beyond 2 years postoperatively. METHODS: Seventy-two patients were randomized to a cementless 3D-printed Triathlon Tritanium (Stryker) cruciate-retaining (CR) TKA or a cemented Triathlon CR (Stryker) TKA implant. Implant migration was evaluated with RSA at baseline and postoperatively at 3 months and at 1, 2, and 5 years. The maximum total point motion (MTPM) of the tibial component was compared between the groups at 5 years, and progression in migration was assessed between 2 and 5 years. Individual implants were classified as continuously migrating if the MTPM was ≥0.1 mm/year beyond 2 years postoperatively. Clinical scores were evaluated, and a linear mixed-effects model was used to analyze repeated measurements. RESULTS: At 5 years, the mean MTPM was 0.66 mm (95% confidence interval [CI], 0.56 to 0.78 mm) for the cementless group and 0.53 mm (95% CI, 0.43 to 0.64 mm) for the cemented group (p = 0.09). Between 2 and 5 years, there was no progression in mean MTPM for the cementless group (0.02 mm; 95% CI, -0.06 to 0.10 mm) versus 0.07 mm (95% CI, 0.00 to 0.14) for the cemented group. One implant was continuously migrating in the cementless group, and 4 were continuously migrating in the cemented group. The clinical scores were comparable between the groups across the entire time of follow-up. CONCLUSIONS: No significant difference in mean migration was found at 5 years between the cementless and cemented TKA implants. Progression of tibial component migration was present beyond 2 years for the cemented implant, whereas the cementless implant remained stable after initial early migration. LEVEL OF EVIDENCE: Therapeutic Level I . See Instructions for Authors for a complete description of levels of evidence.

来源:J Bone Joint Surg Am,查看原文。版权归原作者所有。

引用

  1. van der Lelij TJN, Marang-van de Mheen PJ, Kaptein BL, et al. Continued Stabilization of a Cementless 3D-Printed Total Knee Arthroplasty: Five-Year Results of a Randomized Controlled Trial Using Radiostereometric Analysis. J Bone Joint Surg Am. 2023. PMID: 37651549. doi:10.2106/JBJS.23.00221

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。