股骨头材料、术者经验与组装技术对模拟全髋关节置换头颈撞击力的影响
本研究用体外装置测量31名术者(11名主治、14名住院、6名研究员)以各自惯用击打方式组装金属与陶瓷股骨头时的动态三维撞击力,发现股骨头材料间无显著差异,主治医师施力更沿轴线,多击首次击打各指标显著低于单次击打(偏轴角除外),多击末次击打合力冲量显著低于单次击打。
要点
- 全体术者平均峰值轴向力为6.92 ± 2.11 kN,金属与陶瓷股骨头间无显著差异。
- 主治医师施力较住院医师和研究员更沿轴线(on-axis)。
- 多击试验首次击打除偏轴角外各指标显著低于单次击打,末次击打合力冲量显著低于单次击打。
核心结论
- 术者经验水平及单次与多次击打试验之间在施加的三维力-时间曲线动态特征上存在差异。
- 股骨头材料之间未发现显著差异。
研究设计与样本
- 体外研究,使用仪器化装置测量31名术者(11名主治医师、14名住院医师、6名研究员)以各自偏好技术(击打次数或槌击次数)经股骨柄锥度施加的动态三维力。
主要发现
- 全体术者平均峰值轴向力为6.92 ± 2.11 kN。
- 股骨头材料之间未发现显著差异。
- 主治医师施力较住院医师和研究员均更沿轴线。
- 9名术者1次击打组装股骨头,3名2次,14名3次,2名4次,3名≥5次。
- 多击试验首次击打除偏轴角外所有结局指标均显著低于单次击打。
关键数值与推荐
- 结局变量包括峰值轴向力、峰值合力、合力冲量、合力加载速率和偏轴角。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: There is no standard method for assembling the femoral head onto the femoral stem during total hip arthroplasty (THA). This study aimed to measure and record dynamic 3-dimensional (3D) THA head-neck assembly loads from residents, fellows, and attending surgeons, for metal and ceramic femoral heads. METHODS: An instrumented apparatus measured dynamic 3D forces applied through the femoral stem taper in vitro for 31 surgeons (11 attendings, 14 residents, 6 fellows) using their preferred technique (ie, number of hits or mallet strikes). Outcome variables included peak axial force, peak resultant force, impulse of the resultant force, loading rate of the resultant force, and off-axis angle. They were compared between femoral head material, surgeon experience level, and the number of hits per trial. RESULTS: Average peak axial force was 6.92 ± 2.11kN for all surgeons. No significant differences were found between femoral head material. Attendings applied forces more "on-axis" as compared to both residents and fellows. Nine surgeons assembled the head with 1 hit, 3 with 2 hits, 14 with 3 hits, 2 with 4 hits, and 3 with ≥5 hits. The first hit of multihit trials was significantly lower than single-hit trials for all outcome measures except the off-axis angle. The last hit of multihit trials had a significantly lower impulse of resultant force than single-hit trials. CONCLUSION: Differences in applied 3D force-time curve dynamic characteristics were found between surgeon experience level and single and multihit trials. No significant differences were found between femoral head material.
引用
- Godoy M, Sipek K, Gustafson JA, et al. Effect of Femoral Head Material, Surgeon Experience, and Assembly Technique on Simulated Head-Neck Total Hip Arthroplasty Impaction Forces. J Arthroplasty. 2023. PMID: 37598779. doi:10.1016/j.arth.2023.08.049