全膝关节置换术使用增强现实能否提升手术表现?一项76例连续患者的单中心研究
该单中心研究纳入76例连续患者,在全膝关节置换术中采用增强现实(AR)辅助进行胫骨截骨,记录AR工具使用时间及与术前计划的胫骨内翻角和后倾角差异。AR工具平均使用时间为5±1分钟,内翻角平均差异0.59±0.55°、后倾角平均差异0.70±0.75°,内翻角差异<1°的病例占96%、后倾角差异<1°的病例占89%。作者认为截骨准确性优异且手术时间增加有限,该技术有潜力成为有效的手术辅助选择。
要点
- AR工具平均使用时间为5±1分钟。
- 内翻角平均差异0.59±0.55°,后倾角平均差异0.70±0.75°。
- 内翻角差异<1°的病例占96%,后倾角差异<1°的病例占89%。
核心结论
- 作者认为结果显示了手术截骨优异的准确性及手术时长有限增加,突出了该新技术作为有效手术辅助选择的潜力。
研究设计与样本
- 单中心研究,共纳入76例连续患者,术前按AR方案规划并设定目标内翻角和后倾角以指导设备,术中由设备引导胫骨截骨。
- 手术从胫骨截骨开始,记录校准、配准和固定截骨块所需时间,并记录达到的内翻角和后倾角与术前计划比较,计算差异的均值和标准差。
主要发现
- AR工具平均使用时间为5±1分钟。
- 内翻角平均差异为0.59±0.55°,后倾角平均差异为0.70±0.75°。
- 内翻角差异<1°的病例占96%,后倾角差异<1°的病例占89%。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Augmented reality (AR) is a powerful multipurpose tool. With a dedicated visor, AR allows the visualization of a series of information and/or images superimposed on the user's field of vision. For this reason, it was recently introduced as a surgical assistant tool. This single-center study aimed to evaluate the intraoperative outcomes of total knee arthroplasties performed with AR assistance in terms of time required and the difference between preplanned and achieved implant positioning (in terms of tibial cut varus and slope angles). METHODS: A total of 76 consecutive patients was selected. Preplanning was performed according to the AR protocol, and the target varus and slope angles were defined to instruct the device, which subsequently guided the tibial cuts intraoperatively. Surgeries were performed starting from the tibial cut, and the time required to perform the calibration, registration, and fixation of the resection block was recorded. The varus and slope angles achieved were recorded to compare with the preplanned ones to determine the means and SDs of the differences. RESULTS: The mean usage time of the AR tool was 5 ± 1 minutes. Results showed a mean difference of 0.59 ± 0.55° for varus angles and 0.70 ± 0.75° for the slope. For varus angles, the differences were <1° for 96% of the cases. Concerning the slope, 89% of the cases were <1°. CONCLUSION: The results showed excellent accuracy of the surgical cuts and a limited increase in surgery duration. Therefore, these outcomes highlighted the potential of this new technology as a valid option for surgical assistance.
引用
- Castellarin G, Bori E, Barbieux E, et al. Is Total Knee Arthroplasty Surgical Performance Enhanced Using Augmented Reality? A Single-Center Study on 76 Consecutive Patients. J Arthroplasty. 2023. PMID: 37572726. doi:10.1016/j.arth.2023.08.013
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。