文献速递关节外科

牛津单髁膝关节假体的新设计是否改善了生存率?来自挪威关节置换登记处2012-2021年的研究

Did a New Design of the Oxford Unicompartmental Knee Prosthesis Result in Improved Survival? A Study From the Norwegian Arthroplasty Register 2012-2021.
PubMed 文献 · Clin Orthop Relat Res · 2023-05-30

基于挪威关节置换登记处2012-2021年7549例牛津单髁膝关节置换数据,分析骨水泥型牛津III单钉、骨水泥型牛津Partial双钉和非骨水泥型牛津Partial双钉的5年假体生存率及翻修原因。结果显示5年总体翻修风险无差异,但非骨水泥型牛津Partial感染翻修风险更高,假体周围骨折风险也更高;作者因此不推荐非骨水泥型牛津Partial优于骨水泥型。

要点

  • 5年Kaplan-Meier生存率:骨水泥牛津III为92%,骨水泥牛津Partial为94%,非骨水泥牛津Partial为94%(p=0.03)。
  • 5年总体翻修风险三组间无差异:骨水泥牛津Partial的HR为0.8(95%CI 0.6-1.0;p=0.09),非骨水泥牛津Partial的HR为1.0(95%CI 0.7-1.4;p=0.89)。
  • 非骨水泥牛津Partial相比骨水泥牛津Partial,假体周围骨折翻修风险HR为15(95%CI 4-54;p=0.001),第一年内感染翻修风险HR为3.0(95%CI 1.5-5.7;p=0.001)。

核心结论

  • 考虑到5年内总体翻修风险无差异,但非骨水泥型牛津Partial的感染翻修风险、假体周围骨折风险及假体成本更高,作者目前不推荐使用非骨水泥型牛津Partial优于骨水泥型牛津Partial或骨水泥型牛津III。

研究设计与样本

  • 基于挪威关节置换登记处的登记库观察性研究,该登记处为全国性、强制性政府登记处,报告率高。
  • 2012年至2021年间共施行7549例牛津单髁膝关节置换,排除105例因三种设计组合、外侧间室置换或混合固定者,最终分析908例骨水泥型牛津III单钉(2012-2017年使用)、4715例骨水泥型牛津Partial双钉(2012-2021年使用)和1821例非骨水泥型牛津Partial双钉(2014-2021年使用)。
  • 采用Kaplan-Meier法和Cox回归多变量分析,调整年龄、性别、诊断、美国麻醉医师协会分级和时间段,计算5年假体生存率和翻修风险(风险比)。
  • 翻修定义为任何更换或取出假体部件的操作。

主要发现

  • 5年Kaplan-Meier总体假体生存率(无任何原因翻修)在内侧牛津Partial单髁膝关节中未随时间改善。
  • 5年Kaplan-Meier生存率三组间不同(p=0.03):骨水泥型牛津III为92%(95%CI 90%-94%),骨水泥型牛津Partial为94%(95%CI 93%-95%),非骨水泥型牛津Partial为94%(95%CI 92%-95%)。
  • 前5年总体翻修风险三组间无差异:与骨水泥型牛津III(HR 1)相比,骨水泥型牛津Partial的HR为0.8(95%CI 0.6-1.0;p=0.09),非骨水泥型牛津Partial的HR为1.0(95%CI 0.7-1.4;p=0.89)。
  • 与骨水泥型牛津III相比,非骨水泥型牛津Partial的感染翻修风险更高(HR 3.6;95%CI 1.2-10.5;p=0.02),疼痛翻修风险更低(HR 0.5;95%CI 0.2-1.0;p=0.045),不稳翻修风险更低(HR 0.3;95%CI 0.1-0.9;p=0.03)。
  • 与骨水泥型牛津III相比,骨水泥型牛津Partial的无菌性股骨松动翻修风险更低(HR 0.3;95%CI 0.1-1.0;p=0.04)。

关键数值与推荐

  • 5年Kaplan-Meier生存率:骨水泥型牛津III为92%(95%CI 90%-94%),骨水泥型牛津Partial为94%(95%CI 93%-95%),非骨水泥型牛津Partial为94%(95%CI 92%-95%),组间p=0.03。
  • 与骨水泥型牛津III相比,骨水泥型牛津Partial总体翻修HR为0.8(95%CI 0.6-1.0;p=0.09),非骨水泥型牛津Partial总体翻修HR为1.0(95%CI 0.7-1.4;p=0.89)。
  • 与骨水泥型牛津III相比,非骨水泥型牛津Partial感染翻修HR为3.6(95%CI 1.2-10.5;p=0.02),疼痛翻修HR为0.5(95%CI 0.2-1.0;p=0.045),不稳翻修HR为0.3(95%CI 0.1-0.9;p=0.03)。
  • 与骨水泥型牛津III相比,骨水泥型牛津Partial无菌性股骨松动翻修HR为0.3(95%CI 0.1-1.0;p=0.04)。
  • 与骨水泥型牛津Partial相比,非骨水泥型牛津Partial假体周围骨折翻修HR为15(95%CI 4-54;p=0.001),第一年内感染翻修HR为3.0(95%CI 1.5-5.7;p=0.001)。

临床怎么用

  • 作者目前不推荐使用非骨水泥型牛津Partial优于骨水泥型牛津Partial或骨水泥型牛津III。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Unicompartmental knee arthroplasty (UKA) has generally shown higher revision rates than TKA, and this is particularly true for the femoral component. A twin-peg femoral component (Oxford Partial) has replaced the single-peg version (Oxford Phase III) of the widely used Oxford medial UKA, with the aim of improving femoral component fixation. The introduction of the Oxford Partial Knee also included a fully uncemented option. However, there has been relatively little evidence regarding the effect of these changes on implant survival and revision diagnoses from groups not associated with the implant design. QUESTIONS/PURPOSES: Using data from the Norwegian Arthroplasty Register, we asked: (1) Has the 5-year implant survival (free from revision for any cause) improved with the medial Oxford unicompartmental knee after the introduction of new designs? (2) Did the causes of revision change between the old and new designs? (3) Is there a difference in risk for specific revision causes between the uncemented and cemented versions of the new design? METHODS: We performed a registry-based observational study using data from the Norwegian Arthroplasty Register, a nationwide, mandatory and governmental registry with a high reporting rate. Between 2012 and 2021, 7549 Oxford UKAs were performed, and 105 were excluded due to combinations of the three designs, lateral compartment replacement, or hybrid fixation, leaving 908 cemented Oxford Phase III single-peg (used from 2012 to 2017), 4715 cemented Oxford Partial twin-peg (used from 2012 to 2021), and 1821 uncemented Oxford Partial twin-peg (used from 2014 to 2021), UKAs available for the analysis. The Kaplan-Meier method and Cox regression multivariate analysis were used to find the 5-year implant survival and the risk of revision (hazard ratio), when adjusting for age, gender, diagnosis, American Society of Anesthesiologists grade, and time period. The risk of revision for any cause and the risk of revision for specific causes were compared, first for the older with the two new designs, and second for the cemented with the uncemented version of the new design. Revision was defined as any operation exchanging or removing implant parts. RESULTS: The 5-year Kaplan-Meier overall implant survival (free from revision for any cause) for the medial Oxford Partial unicompartmental knee did not improve over the study period. The 5-year Kaplan-Meier survival was different (p = 0.03) between the groups: it was 92% (95% confidence interval [CI] 90% to 94%) for the cemented Oxford III, 94% (95% CI 93% to 95%) for the cemented Oxford Partial, and 94% (95% CI 92% to 95%) for the uncemented Oxford Partial. However, the overall risk of revision during the first 5 years was not different between the groups (Cox regression HR 0.8 [95% CI 0.6 to 1.0]; p = 0.09 and 1.0 [95% CI 0.7 to 1.4]; p = 0.89 for the cemented Oxford Partial and the uncemented Oxford Partial, respectively, compared with cemented Oxford III [HR 1]). The uncemented Oxford Partial had a higher risk of revision for infection (HR 3.6 [95% CI 1.2 to 10.5]; p = 0.02) compared with the cemented Oxford III. The uncemented Oxford Partial had a lower risk of revision for pain (HR 0.5 [95% CI 0.2 to 1.0]; p = 0.045) and instability (HR 0.3 [95% CI 0.1 to 0.9]; p = 0.03) compared with the cemented Oxford III. The cemented Oxford Partial had a lower risk of revision for aseptic femoral loosening (HR 0.3 [95% CI 0.1 to 1.0]; p = 0.04) compared with the cemented Oxford III. When comparing the uncemented and cemented versions of the new design, the uncemented Oxford Partial had a higher risk of revision for periprosthetic fracture (HR 15 [95% CI 4 to 54]; p = 0.001) and infection within the first year (HR 3.0 [95% CI 1.5 to 5.7]; p = 0.001) than the cemented Oxford Partial. CONCLUSION: Considering that we found no difference in overall risk of revision during the first 5 years but we found a higher risk of revision for infection, periprosthetic fracture, and higher per implant cost, we currently would recommend against the use of uncemented Oxford Partial over the cemented Oxford Partial or the cemented Oxford III. LEVEL OF EVIDENCE: Level III, therapeutic study.

来源:Clin Orthop Relat Res,查看原文。版权归原作者所有。

引用

  1. Skåden Ø, Furnes ON, Låstad Lygre SH, et al. Did a New Design of the Oxford Unicompartmental Knee Prosthesis Result in Improved Survival? A Study From the Norwegian Arthroplasty Register 2012-2021. Clin Orthop Relat Res. 2023. PMID: 37140943. doi:10.1097/CORR.0000000000002671

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。