脊柱手术中的机器人导航辅助
本研究评估了使用Mazor X Stealth系统行机器人导航辅助(RNA)椎弓根螺钉置入的准确性、术中因素、辐射暴露和并发症率,纳入2019年6月至2022年3月期间196例成人胸腰椎退行性疾病患者,共置入1123枚螺钉,其中机器人置入的螺钉中94%为Gertzbein and Robbins A级,90天随访无不良临床后遗症。
要点
- 机器人置入的1001枚螺钉中,942枚(94%)为Gertzbein and Robbins A级,中位偏差0.8 mm(IQR 0.4至1.6)。
- 63枚(6%)螺钉从机器人置入转为徒手技术,59枚(5%)计划徒手置入。
- 20枚螺钉(1.8%)出现滑移事件,90天随访无不良临床后遗症。
核心结论
- 机器人导航辅助(RNA)具有高度准确性和可靠性,掌握后放弃率低。
- 使用最新一代RNA置入大量椎弓根螺钉后无不良临床后遗症发生。
- 了解患者特定解剖特征和术中实时识别次优螺钉置入的危险因素可能进一步提高准确性。
研究设计与样本
- 研究回顾了2019年6月至2022年3月期间前瞻性收集的196例成人胸腰椎退行性疾病患者数据,患者接受Mazor X Stealth系统机器人导航辅助(RNA)椎弓根螺钉置入。
- 所有螺钉由一位经验丰富的脊柱外科医生在学习期完成后置入,准确性通过术中3D透视确定。
主要发现
- 共置入1123枚椎弓根螺钉:1001枚(89%)为机器人置入,63枚(6%)从机器人置入转为徒手技术,59枚(5%)计划徒手置入。
- 机器人置入的螺钉中,942枚(94%)为Gertzbein and Robbins A级,中位偏差0.8 mm(四分位距0.4至1.6)。
- 20枚椎弓根螺钉(1.8%)出现滑移事件。
- 90天随访中未观察到不良临床后遗症。
- 每枚螺钉平均透视暴露时间为4.9秒(标准差3.8)。
关键数值与推荐
- 机器人置入螺钉的Gertzbein and Robbins A级比例为94%(942/1001),中位偏差0.8 mm(IQR 0.4至1.6)。
- 滑移事件发生率为1.8%(20/1123枚螺钉)。
- 每枚螺钉平均透视暴露时间4.9秒(SD 3.8)。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
The aim of this study was to assess the accuracy of pedicle screw placement, as well as intraoperative factors, radiation exposure, and complication rates in adult patients with degenerative disorders of the thoracic and lumbar spines who have undergone robotic-navigated spinal surgery using a contemporary system. The authors reviewed the prospectively collected data on 196 adult patients who had pedicle screws implanted with robot-navigated assistance (RNA) using the Mazor X Stealth system between June 2019 and March 2022. Pedicle screws were implanted by one experienced spinal surgeon after completion of a learning period. The accuracy of pedicle screw placement was determined using intraoperative 3D fluoroscopy. A total of 1,123 pedicle screws were implanted: 1,001 screws (89%) were placed robotically, 63 (6%) were converted from robotic placement to a freehand technique, and 59 (5%) were planned to be implanted freehand. Of the robotically placed screws, 942 screws (94%) were determined to be Gertzbein and Robbins grade A with median deviation of 0.8 mm (interquartile range 0.4 to 1.6). Skive events were noted with 20 pedicle screws (1.8%). No adverse clinical sequelae were noted in the 90-day follow-up. The mean fluoroscopic exposure per screw was 4.9 seconds (SD 3.8). RNA is highly accurate and reliable, with a low rate of abandonment once mastered. No adverse clinical sequelae occurred after implanting a large series of pedicle screws using the latest generation of RNA. Understanding of patient-specific anatomical features and the real-time intraoperative identification of risk factors for suboptimal screw placement have the potential to improve accuracy further.
引用
- Abel F, Avrumova F, Goldman SN, et al. Robotic-navigated assistance in spine surgery. Bone Joint J. 2023. PMID: 37121590. doi:10.1302/0301-620X.105B5.BJJ-2022-0810.R3
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。