基于双平面同时X线图像测量髋臼杯方位新方法的验证

Validation of a Novel Method of Measuring Cup Orientation Using BiPlanar Simultaneous Radiographic Images.
PubMed 文献 · J Arthroplasty · 2023-04-17

研究纳入40例对侧已行全髋关节置换术的患者,在术前计划时同时接受CT和双平面正交X线扫描,用新方法计算髋臼杯的手术倾斜角(OI)和手术前倾角(OA),并与CT测量比较,由2名独立观察者完成测量。结果显示OA平均误差0.5°(标准差1.9°),OI平均误差0.0°(标准差1.7°),观察者间相关系数OA为0.83、OI为0.93,作者认为该新方法准确且可重复。

要点

  • 新方法基于双平面同时X线图像测量髋臼杯方位,与CT测量相比OA平均误差0.5°、OI平均误差0.0°。
  • OA平均绝对误差1.5°、OI平均绝对误差1.2°。
  • 观察者间相关系数OA为0.83、OI为0.93。

核心结论

  • 本研究所用的基于双平面同时X线扫描测量髋臼杯方位的新方法,与CT测量相比准确,且观察者间可重复。

研究设计与样本

  • 连续纳入40例对侧已行全髋关节置换术、因全髋关节置换术术前计划而接受CT和双平面同时正交X线扫描的患者。
  • 由2名独立观察者用新方法基于双平面同时扫描计算髋臼杯的手术倾斜角(OI)和手术前倾角(OA),并与CT上的髋臼杯方位测量比较,计算观察者间相关系数评估可靠性。

主要发现

  • 双平面同时正交X线与CT测量髋臼杯OA的平均误差为0.5°(标准差1.9°,最小-4.0°,最大5.0°)。
  • 双平面同时正交X线与CT测量髋臼杯OI的平均误差为0.0°(标准差1.7°,最小-5.0°,最大4.0°)。
  • OA的平均绝对误差为1.5°,OI的平均绝对误差为1.2°。
  • 观察者间相关系数OA为0.83,OI为0.93。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Accurate acetabular component positioning is paramount to the success of total hip arthroplasty (THA). Two-dimensional imaging alone remains a popular tool for implant position assessment despite known limitations. We investigated the accuracy of a novel method for assessing acetabular component position based upon orthogonal simultaneous biplanar X-ray images. METHODS: There were forty consecutive patients who had a preexisting THA on the contralateral side who underwent both computed tomography (CT) and simultaneous orthogonal biplanar radiographic scans for preoperative planning of THA. The operative inclination (OI) and operative anteversion (OA) of the acetabular cup were calculated by a new measurement method using the biplanar simultaneous scans. Those measurements were compared to measurement of the cup orientation on CT. The measurements were made by 2 independent observers. Interobserver correlation coefficients were calculated between the 2 observers to measure reliability. RESULTS: The mean error in OA measurement of the acetabular cup between simultaneous orthogonal biplanar radiographic and CT imaging was 0.5° (SD: 1.9°, minimum -4.0°, maximum 5.0°), the mean error in OI was 0.0° (SD: 1.7°, minimum -5.0°, maximum 4.0°). The average absolute error was 1.5° for OA and 1.2° for OI. Interobserver correlation coefficient was 0.83 for OA and 0.93 for OI. CONCLUSION: The novel method of measuring cup orientation using simultaneous biplanar radiographic scans utilized in this study was accurate and reproducible between observers compared to CT measurements.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Sun DC, Murphy WS, Amundson AJ, et al. Validation of a Novel Method of Measuring Cup Orientation Using BiPlanar Simultaneous Radiographic Images. J Arthroplasty. 2023. PMID: 37075906. doi:10.1016/j.arth.2023.04.011