温哥华B2型假体周围股骨骨折固定的稳定性:植入技术的影响

The Stability of Fixation of Vancouver B2 Periprosthetic Femoral Fractures: Effect of Implantation Technique.
PubMed 文献 · J Arthroplasty · 2023-03-16

研究通过7对尸体股骨复制温哥华B2型骨折,比较「先复位」与「先扩髓」两种技术植入锥形带槽股骨柄联合环扎钢缆的固定稳定性。结果显示两组假体下沉、旋转及骨折块间动度均无显著差异,两种技术均提供足够的假体与骨折稳定性。

要点

  • 在7对尸体股骨上复制温哥华B2型骨折,比较先复位与先扩髓两种技术。
  • 先扩髓组平均柄直径16.1±0.4 mm,先复位组15.4±0.5 mm。
  • 两组假体下沉、旋转及骨折块动度均无显著差异,固定稳定性相当。

核心结论

  • 使用锥形带槽股骨柄联合环扎钢缆治疗温哥华B2型假体周围股骨骨折时,先复位与先扩髓两种技术均显示足够的假体与骨折稳定性。

研究设计与样本

  • 通过回顾30例B2型病例创建一种常见B2型骨折,在7对尸体股骨上复制该骨折并分为两组。
  • I组(先复位)先复位骨折块后植入锥形带槽股骨柄;II组(先扩髓)先在股骨远端植入股骨柄,再复位并固定骨折块。
  • 每例标本在多轴测试架上以行走时峰值载荷的70%加载,用运动捕捉系统追踪股骨柄与骨折块的运动。

主要发现

  • II组平均柄直径16.1±0.4 mm,I组为15.4±0.5 mm。
  • 测试后I组平均假体下沉0.36±0.31 mm,II组0.19±0.14 mm(P = .17)。
  • I组平均旋转1.67±1.30°,II组0.91±1.11°(P = .16)。
  • 与股骨柄相比,骨折块动度更小,且两组间无差异(P > .05)。

临床怎么用

  • 原文结论认为,使用锥形带槽股骨柄联合环扎钢缆时,先复位与先扩髓两种技术均可提供足够的假体与骨折稳定性。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Due to increasing volume of total hip arthroplasties, periprosthetic femoral fractures have become a common complication with increased revision burden and perioperative morbidity. The objective of this study was to evaluate the fixation stability of Vancouver B2 fractures treated with 2 techniques. METHODS: A common B2 fracture was created by reviewing 30 type B2 cases. The fracture was then reproduced in 7 pairs of cadaveric femora. The specimens were divided into 2 groups. In Group I ("reduce-first"), the fragments were reduced first, followed by implantation of a tapered fluted stem. In Group II ("ream-first"), the stem was implanted in the distal femur first, followed by fragment reduction and fixation. Each specimen was loaded in a multiaxial testing frame with 70% of peak load during walking. A motion capture system was used to track the motion of the stem and fragments. RESULTS: The average stem diameter in Group II was 16.1 ± 0.4 mm, versus 15.4 ± 0.5 mm in Group I. The fixation stability was not significantly different in the 2 groups. After the testing, the average stem subsidence was 0.36 ± 0.31 mm and 0.19 ± 0.14 mm (P = .17) and the average rotation was 1.67 ± 1.30° and 0.91 ± 1.11° (P = .16) in Groups I and II, respectively. Compared to the stem, there was less motion of the fragments and there was no difference between the 2 groups (P > .05). CONCLUSIONS: When tapered fluted stems were used in combination with cerclage cables for treatment of Vancouver type B2 periprosthetic femoral fractures, both the "reduce-first" and "ream-first" techniques showed adequate stem and fracture stability.