文献速递关节外科

无导航且功能良好的间隙平衡全膝关节置换术中的假体旋转

Component Rotation in Well-Functioning, Gap Balanced Total Knee Arthroplasty Without Navigation.
PubMed 文献 · J Arthroplasty · 2023-03-23

研究纳入93例功能良好、采用间隙平衡技术且无导航的全膝关节置换术(TKA),术后用CT评估假体旋转对线。多数假体旋转在期望范围内,股骨假体后髁轴相对经上髁轴平均内旋0.78°,胫骨假体相对胫骨结节平均内旋17.2°。天然股骨前倾与股骨假体旋转无相关(r=0.007),天然胫骨扭转与胫骨假体对线弱相关(r=0.24)。

要点

  • 功能良好的间隙平衡TKA中,多数假体旋转对线在期望范围内。
  • 股骨假体后髁轴相对经上髁轴平均内旋0.78°(范围-8.3至4.1°)。
  • 天然股骨前倾与股骨假体旋转无相关(r=0.007),胫骨扭转与胫骨假体对线弱相关(r=0.24)。

核心结论

  • 尽管旋转对线只是间隙平衡技术的次要目标,大多数评估的假体仍在期望的旋转范围内。
  • 假体对线未受天然骨性旋转几何形态的影响。

研究设计与样本

  • 回顾性队列研究(证据等级III),纳入93例功能良好、采用先行伸直间隙的间隙平衡技术且无导航的全膝关节置换术,随访至少2年。
  • 采用计算机断层扫描评估假体旋转对线以及骨性股骨前倾和胫骨扭转。

主要发现

  • 平均膝关节协会评分为185.7±21.7。
  • 平均活动范围为128.5±7.8°。
  • 股骨后髁轴(相对经上髁轴)值为-8.3至4.1°,平均-0.78±2.7°(内旋)。
  • 胫骨旋转相对胫骨结节平均内旋17.2±7.9°。
  • 天然股骨前倾与股骨假体旋转对线无相关性(Pearson相关系数r=0.007)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Malalignment of total knee arthroplasty (TKA) components is a potential cause of clinical failure following TKA. Since the goal of a gap balancing (GB) technique is equal flexion and extension gaps secondary to soft-tissue balancing, and not necessarily component alignment, variation in component placement may exist. Our purpose was (1) to evaluate precision of component alignment in well-functioning GB TKAs performed without the aid of navigation using computed tomographic evaluation and (2) to determine any relationship between femoral version and/or tibial torsion and TKA component positioning. METHODS: There were 93 well-functioning TKAs performed with an extension gap first GB technique with a minimum 2-year follow-up evaluated using computed tomography to assess component rotational alignment, as well as osseous femoral version and tibial torsion. Femoral and tibial rotational alignment was assessed by previously described methods. RESULTS: The mean Knee Society Score was 185.7 ± 21.7. The mean range of motion was 128.5 ± 7.8°. Femoral posterocondylar axis (relative to the transepicondylar axis) values ranged from -8.3 to 4.1° with a mean of -0.78 ± 2.7° (internal rotation). Mean tibial rotation was 17.2 ± 7.9° internal rotation relative to the tibial tubercle. No correlation was found between native femoral version and femoral component rotational alignment (Pearson's correlation coefficient, r, 0.007). Weak correlation was found between native tibial torsion and tibial component alignment (r = 0.24). CONCLUSION: Despite being only a secondary objective with the GB technique, most components evaluated were within the desired range of rotation. Alignment was not influenced by native osseous rotational geometry. LEVEL OF EVIDENCE: Level III, Retrospective Cohort Study.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Elkins JM, Jennings JM, Johnson RM, et al. Component Rotation in Well-Functioning, Gap Balanced Total Knee Arthroplasty Without Navigation. J Arthroplasty. 2023. PMID: 36963529. doi:10.1016/j.arth.2023.03.033

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。