模块化头颈适配器系统与陶瓷头在翻修髋关节置换术中的应用:一项基于354个植入物的登记研究
该登记研究评估了翻修髋关节置换术中模块化头颈适配器系统联合Delta陶瓷头的5年和7年生存率、再翻修原因,并与工厂组装的陶瓷钛套头进行对比。354例翻修平均随访5年(范围0至13年),5年和7年生存率分别为87.9%和86.9%,脱位(2.8%)和髋臼杯无菌性松动(4.2%)是最常见的再翻修原因,未发生适配器系统或陶瓷头断裂。两组生存率(P=0.696)和再翻修原因(脱位P=0.983;髋臼杯无菌性松动P=0.296)无差异。
要点
- 模块化头颈适配器联合陶瓷头翻修髋5年生存率87.9%、7年86.9%。
- 再翻修最常见原因为脱位(2.8%)和髋臼杯无菌性松动(4.2%)。
- 未发生适配器系统或陶瓷头断裂,股骨颈失败0.3%。
核心结论
- 模块化头颈适配器系统联合陶瓷头在翻修术中5年和7年 seems 是有效选择,该登记研究中未增加植入物断裂风险。
研究设计与样本
- 利用区域关节置换登记系统,研究翻修髋中模块化适配器系统联合陶瓷头,并与工厂组装的陶瓷钛套头翻修病例作为对照组比较。
- 共纳入354例翻修,平均随访5年(范围0至13年),记录人口学、植入物特征和翻修原因,评估生存率和再翻修原因。
主要发现
- 5年和7年生存率分别为87.9%和86.9%。
- 再翻修最常见原因为脱位(2.8%)和髋臼杯无菌性松动(4.2%)。
- 未发生适配器系统或陶瓷头断裂;股骨颈失败0.3%。
- 植入物特征偏距(P=0.088)或裙状锥度(P=0.870)对失败无影响。
- 两队列生存率无差异(P=0.696),再翻修原因无差异(脱位P=0.983;髋臼杯无菌性松动P=0.296)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Five- to 10-year outcomes of a modular head-neck adapter system with ceramic heads in revision hip arthroplasty on large populations are still lacking. A registry study about modular adapter system with Delta ceramic head in revisions was designed, aiming to assess (1) the survival rates of the device, (2) the reasons for re-revisions of the device, and (3) a comparison to factory assembled titanium sleeve and Delta head cohort in revision hips. METHODS: Using a regional arthroplasty registry, we investigated the modular adapter system with a ceramic head in revision hips. Demographics, implant features, and reasons for revision were recorded. Survival rates and reasons for re-revision were assessed. The modular adapter system with the ceramic ball was compared to a factory-assembled ceramic titanium-sleeved head in revisions, acting as a control group. There were 354 revisions included at a mean follow-up of 5 years (range, 0 to 13). RESULTS: The 5- and 7-year survival rates were 87.9% and 86.9%, respectively. Dislocations (2.8%) and cup aseptic loosening (4.2%) were the 2 most frequent reasons for re-revision. No breakage of the adapter system or the ceramic head occurred. A femoral neck failed (0.3%). No implant features, offset (P = .088) or skirted (P = .870) tapers, impacted on failures. No differences between the 2 cohorts were found regarding survival rates (P = .696) and reasons for re-revision (dislocations, P = .983; cup aseptic loosening, P = .296). CONCLUSION: The modular head neck adapter system with a ceramic head seems to be a valid option in revisions at 5 and 7 years, without additional risk of implant breakage in this registry.
引用
- Pardo F, Castagnini F, Bordini B, et al. A Modular Head-Neck Adapter System and Ceramic Heads in Revision Hip Arthroplasty: A Registry Study on 354 Implants. J Arthroplasty. 2023. PMID: 36764407. doi:10.1016/j.arth.2023.01.055
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。