文献速递关节外科

后稳定型全膝关节置换术中聚乙烯柱骨折的临床与生物力学特征:回收分析

Clinical and Biomechanical Characteristics of Posterior-Stabilized Polyethylene Post Fractures in Total Knee Arthroplasty: A Retrieval Analysis.
PubMed 文献 · J Arthroplasty · 2023-02-21

该回收分析评估了2015至2022年间33例因聚乙烯柱骨折而翻修的后稳定型全膝关节置换术(TKA)聚乙烯部件。结果显示,超高分子量聚乙烯(UHMWPE)组的总表面损伤评分显著高于高交联聚乙烯(XLPE)组(57.3 vs 44.2,P=0.003)。XLPE植入物的骨折发生于更短的植入时长后且表面损伤更轻,扫描电镜(SEM)评估提示更脆性的骨折模式。

要点

  • UHMWPE组总表面损伤评分显著高于XLPE组(57.3 vs 44.2,P=0.003)。
  • SEM显示13例中10例骨折起始于柱的后缘。
  • XLPE柱骨折发生于更短植入时长后且表面损伤更轻,提示更脆性骨折模式。

核心结论

  • XLPE与UHMWPE植入物的后稳定柱骨折特征不同,XLPE骨折发生于更短植入时长后且表面损伤更轻,SEM评估提示更脆性骨折模式。

研究设计与样本

  • 回顾性回收分析,纳入2015至2022年间因聚乙烯柱骨折而翻修的33例初次后稳定型全膝关节置换术聚乙烯部件。

主要发现

  • UHMWPE组总表面损伤评分显著高于XLPE组(57.3 vs 44.2,P=0.003)。
  • SEM显示13例中10例骨折起始于柱的后缘。
  • UHMWPE骨折面具有更多簇状、不规则蛤壳特征,而XLPE柱具有更精确的蛤壳标记及急性最终骨折区域的钻石图案。
  • 患者平均初次手术年龄55岁(范围35至69),平均体重指数29.5(范围18.5至37.2),平均植入时长10.0年(范围4至26)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Fractures of the polyethylene post are a rare but known complication after posterior-stabilized (PS) total knee arthroplasty (TKA). We evaluated the polyethylene and patient characteristics for 33 primary PS polyethylene components that were revised with fractured posts. METHODS: We identified 33 PS inserts revised between 2015 and 2022. Patient characteristics collected included age at index TKA, sex, body mass index, length of implantation (LOI), and patient-reported details on events surrounding the post fracture. Implant characteristics recorded were manufacturer, cross-linking properties (highly cross-linked polyethylene [XLPE] versus ultra-high molecular weight polyethylene [UHMWPE]), wear characteristics based on subjective scoring of the articular surfaces and scanning electron microscopy (SEM) of fracture surfaces. Mean age at index surgery was 55 years (range, 35 to 69), mean body mass index was 29.5 (range, 18.5 to 37.2), and mean LOI was 10.0 (range, 4 to 26). RESULTS: Total surface damage scores were significantly higher in the UHMWPE group versus the XLPE group (57.3 versus 44.2, P = .003). SEM demonstrated fracture initiation at the posterior edge of the post in 10 of 13 cases. UHMWPE fracture surfaces posts had more tufted, irregularly clamshell features, while XLPE posts had more precise clamshell marking and a diamond pattern in the region of acute, final fracture. CONCLUSION: Characteristics of PS post fracture differed between XLPE and UHMWPE implants, with fractures occurring in the XLPE with less generalized surface damage, after a shorter LOI, and with SEM evaluation indicative of a more brittle fracture pattern.

来源:J Arthroplasty,查看原文。版权归原作者所有。