新型关节测量仪用于全膝关节置换术后膝关节三平面体内松弛度的量化评估
本研究验证一种新型多平面关节测量仪在全膝关节置换术后患者中评估膝关节松弛度的安全性与可靠性。20例患者(平均年龄65岁)在术后3个月或1年时接受测试,测试中疼痛平均0.7分(满分10分)。组内信度在所有加载方向上均大于0.77,组间信度在内外翻、内外旋和前后方向上分别为0.85、0.67和0.54。
要点
- 新型关节测量仪可安全评估全膝关节置换术后前后、内外翻和内外旋三方向松弛度。
- 测试中疼痛平均0.7分(范围0-2.5,满分10分)。
- 组内信度在所有加载方向均大于0.77,组间信度内外翻0.85、内外旋0.67、前后0.54。
核心结论
- 新型关节测量仪用于评估全膝关节置换术后患者的前后、内外翻和内外旋松弛度是安全的。
- 该设备可用于检查松弛度与患者对膝关节不稳定性感知之间的关系。
研究设计与样本
- 研究使用五自由度仪器化连杆的关节测量仪。
- 两名检查者分别对20例全膝关节置换术后患者的患腿各进行两次测试,患者平均年龄65岁(范围53-75岁),其中9名男性、11名女性。
- 术后3个月测试9例不同患者,术后1年测试11例不同患者。
主要发现
- 所有受试者均成功完成测试。
- 测试中疼痛平均0.7分(满分10分;范围0-2.5)。
- 组内信度在所有加载方向和检查者均大于0.77。
- 组间信度及95%置信区间在内外翻方向为0.85(0.66-0.94),在内外旋方向为0.67(0.35-0.85),在前后方向为0.54(0.16-0.79)。
关键数值与推荐
- 对每名受试者的置换膝关节施加前后方向力-10至30牛顿、内外翻力矩±3牛顿·米、内外旋力矩±2.5牛顿·米。
- 膝关节疼痛严重程度和位置采用视觉模拟量表评估。
- 组内和组间信度采用组内相关系数表征。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Knee instability is a leading cause of dissatisfaction following total knee arthroplasty (TKA). Instability can involve abnormal laxity in multiple directions including varus-valgus (VV) angulation, anterior-posterior (AP) translation, and internal-external rotation (IER). No existing arthrometer objectively quantifies knee laxity in all three of these directions. The study objectives were to verify the safety and assess reliability of a novel multiplanar arthrometer. METHODS: The arthrometer utilized a five degree-of-freedom instrumented linkage. Two examiners each conducted two tests on the leg that had received a TKA of 20 patients (mean age 65 years (range, 53-75); 9 men, 11 women), with nine and eleven distinct patients tested at 3-month and 1-year postoperative time points, respectively. AP forces from -10 to 30 Newtons, VV moments of ±3 Newton-meters, and IER moments of ±2.5 Newton-meters were applied to each subject's replaced knee. Severity and location of knee pain during testing were assessed using a visual analog scale. Intraexaminer and interexaminer reliabilities were characterized using intraclass correlation coefficients. RESULTS: All subjects successfully completed testing. Pain during testing averaged 0.7 (out of possible 10; range, 0-2.5). Intraexaminer reliability was >0.77 for all loading directions and examiners. Interexaminer reliability and 95% confidence intervals were 0.85 (0.66-0.94), 0.67 (0.35-0.85), and 0.54 (0.16-0.79) in the VV, IER, and AP directions, respectively. CONCLUSION: The novel arthrometer was safe for evaluating AP, VV, and IER laxities in subjects who had received TKA. This device could be used to examine relationships between laxity and patient perceptions of knee instability.