上端固定椎使用钩与椎弓根螺钉的体外生物力学比较

Hooks Versus Pedicle Screws at the Upper Instrumented Level: An In Vitro Biomechanical Comparison.
PubMed 文献 · Spine (Phila Pa 1976) · 2023-02-06

该体外生物力学研究将27个胸椎功能脊柱单元随机分为三组(每组9个),比较双侧椎弓根螺钉、双侧横突钩及后部结构损伤的双侧椎弓根螺钉在上端固定椎的固定效果。循环加载后,横突钩组上端固定椎相邻节段屈伸活动度平均增加29.4%,低于椎弓根螺钉组的76.6%(P<0.05);小关节囊损伤导致相邻节段活动度增加49.9%。作者认为椎弓根螺钉刚度更大,使相邻节段形成较软的过渡。

要点

  • 循环加载后,横突钩组上端固定椎相邻节段屈伸活动度平均增加29.4%,椎弓根螺钉组增加76.6%(P<0.05)。
  • 后部结构损伤的双侧椎弓根螺钉组在循环加载前,小关节切开和棘间韧带切断分别使屈伸活动度增加49.9%和62.19%。
  • 作者认为椎弓根螺钉较横突钩刚度更大,在上端固定椎与相邻节段间形成较软的过渡。

核心结论

  • 循环加载后,上端固定椎使用双侧椎弓根螺钉较双侧横突钩导致相邻节段活动度更大。
  • 作者认为这可能源于椎弓根螺钉较横突钩刚度更大,使横突钩构造与相邻节段原生解剖之间形成较软的过渡。
  • 小关节囊损伤导致上端固定椎相邻节段活动度增加49.9%,凸显保留后部韧带复合体的重要性。
  • 作者认为需要开展考虑融合率的临床研究,以确定较软过渡的构造是否在体内减少近端交界性后凸。

研究设计与样本

  • 受控实验室研究,采用27个胸椎功能脊柱单元,随机分为三组(每组9个):双侧椎弓根螺钉组、双侧横突钩组、后部结构损伤的双侧椎弓根螺钉组。
  • 标本在伺服液压材料测试系统上分别于原生状态、置入器械后及循环加载后进行测试。
  • 功能脊柱单元接受屈伸、侧弯和轴向旋转加载。

主要发现

  • 循环测试后,横突钩组上端固定椎相邻节段屈伸活动度平均增加29.4%,双侧椎弓根螺钉组增加76.6%(P<0.05)。
  • 后部结构损伤的双侧椎弓根螺钉组在循环测试前,小关节切开和棘间韧带切断分别使屈伸活动度增加49.9%和62.19%。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 作者认为需要开展考虑融合率的临床研究,以确定较软过渡的构造是否在体内减少近端交界性后凸。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

STUDY DESIGN: Controlled laboratory study. OBJECTIVE: The aim was to compare motions at the upper instrumented vertebra (UIV) and supra-adjacent level (UIV+1) between two fixation techniques in thoracic posterior spinal fusion constructs. We hypothesized there would be greater motion at UIV+1 after cyclic loading across all constructs and bilateral pedicle screws (BPSs) with posterior ligamentous compromise would demonstrate the greatest UIV+1 range of motion. SUMMARY OF BACKGROUND DATA: Proximal junctional kyphosis is a well-recognized complication following long thoracolumbar posterior spinal fusion, however, its mechanism is poorly understood. MATERIALS AND METHODS: Twenty-seven thoracic functional spine units were randomly divided into three UIV fixation groups (n=9): (1) BPS, (2) bilateral transverse process hooks (TPHs), and (3) BPS with compromise of the posterior elements between UIV and UIV+1 (BPS-C). Specimens were tested on a servohydraulic materials testing system in native state, following instrumentation, and after cyclic loading. functional spine units were loaded in flexion-extension (FE), lateral bending, and axial rotation. RESULTS: After cyclic testing, the TPH group had a mean 29.4% increase in FE range of motion at UIV+1 versus 76.6% in the BPS group ( P <0.05). The BPS-C group showed an increased FE of 49.9% and 62.19% with sectioning of the facet joints and interspinous ligament respectively prior to cyclic testing. CONCLUSION: BPSs at the UIV led to greater motion at UIV+1 compared to bilateral TPH after cyclic loading. This is likely due to the increased rigidity of BPS compared to TPH leading to a "softer" transition between the TPH construct and native anatomy at the supra-adjacent level. Facet capsule compromise led to a 49.9% increase in UIV+1 motion, underscoring the importance of preserving the posterior ligamentous complex. Clinical studies that account for fusion rates are warranted to determine if constructs with a "soft transition" result in less proximal junctional kyphosis in vivo .

来源:Spine (Phila Pa 1976),查看原文。版权归原作者所有。