全髋关节置换术中维生素E稳定化聚乙烯植入物的体内性能:综述
本文系统综述了41项研究,评估维生素E稳定化高交联聚乙烯(VEPE)在全髋关节置换术中的体内性能。与不含维生素E的高交联聚乙烯及传统超高分子量聚乙烯相比,VEPE的结局等效或更优,髋臼内衬磨损率普遍低于0.1毫米/年,多数低于0.05毫米/年,无因骨溶解或维生素E相关不良结局而翻修的组件。
要点
- VEPE在THA中临床有效,磨损率低于或等效于对照聚乙烯。
- 髋臼内衬磨损率普遍<0.1毫米/年,多数<0.05毫米/年。
- 无VEPE组件因骨溶解或维生素E相关不良结局而翻修。
核心结论
- 文献表明VEPE在全髋关节置换术中临床有效,多数研究显示临床结局积极且磨损率低于或等效于对照聚乙烯。
- 有早期组件骨折的报道,但存在多种潜在原因。
- 混合法与扩散法制造的VEPE组件比较存在文献空白,维生素E引入方式的体内影响尚待明确。
研究设计与样本
- 遵循PRISMA指南的系统文献综述,检索PubMed和Embase,纳入报告VEPE在THA中体内行为的研究,共复习41项研究。
主要发现
- 所有比较含与不含维生素E稳定化聚乙烯的研究中,VEPE结局等效或优于对照组(分别针对不含维生素E的高交联聚乙烯和传统超高分子量聚乙烯对照)。
- 髋臼内衬磨损率普遍低于0.1毫米/年,多数低于0.05毫米/年。
- 无VEPE组件因骨溶解或维生素E引入相关不良结局而翻修。
关键数值与推荐
- 髋臼内衬磨损率普遍<0.1毫米/年,多数<0.05毫米/年。
临床怎么用
- 原文未据此给出推荐。
局限性
- 有早期组件骨折的报道,但存在多种潜在原因。
- 混合法与扩散法制造的VEPE组件比较存在文献空白,维生素E引入方式的体内影响尚待明确。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Vitamin E stabilization was introduced to improve the oxidative stability, wear resistance, and mechanical properties of highly crosslinked polyethylene (HXLPE). In this literature review, we asked: (1) How has vitamin E-stabilized HXLPE (VEPE) performed in vivo for total hip arthroplasty (THA) and how does it compare with conventional ultra-high molecular weight polyethylene (UHMWPE) and HXLPE without vitamin E; and (2) Is there an apparent difference in the clinical performance of VEPE created by blending versus diffusion? METHODS: We performed a systematic search of the literature according to the Preferred Reporting Items for Systematic Review and Meta-Analyses (PRISMA) guidelines using PubMed and Embase. Included studies reported the in vivo behavior of VEPE in THA. We reviewed 41 studies. RESULTS: For all studies that compared polyethylene with and without VE stabilization, outcomes for VEPE were either equivalent or superior to the control group (for HXLPE without VE and conventional UHMWPE controls, respectively). Hip insert wear rates were generally less than 0.1 mm/year and in most cases were less than 0.05 mm/year. No VEPE components were revised for osteolysis or adverse outcomes specific to VE incorporation. CONCLUSION: Across the literature, we found that VEPE was reported to be clinically effective for THA applications, with much of the research indicating positive clinical outcomes and lower or equivalent wear rates compared to conventional UHMWPE and HXLPE controls without VE. Instances of early component fracture were reported, but have multiple potential causes. There is a gap in the literature for comparison of blended and diffused components, so the in vivo impact of VE incorporation method remains to be seen. Overall, this study provides a comprehensive summary of VEPE clinical performance for THA and may serve as a resource for future investigations.
引用
- Spece H, Yarbrough RV, Kurtz SM, et al. In Vivo Performance of Vitamin E Stabilized Polyethylene Implants for Total Hip Arthroplasty: A Review. J Arthroplasty. 2022. PMID: 36481286. doi:10.1016/j.arth.2022.11.010
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。