全膝关节置换术中胫骨假体旋转的参考轴:基于1351例胫骨的计算机断层扫描研究

Reference Axes for Tibial Component Rotation in Total Knee Arthroplasty: Computed Tomography-Based Study of 1,351 Tibiae.
PubMed 文献 · J Bone Joint Surg Am · 2022-10-26

基于SOMA数据库1351例胫骨CT扫描,以Berger法为参考标准,比较多种胫骨旋转参考轴的变异性与内旋误差风险。Mayo法与Insall线变异性低(分别为7.8°±1.0°和5.1°±2.2°),内旋误差风险低(分别为0.7%和1.8%)。亚洲人较白人胫骨内旋更大(TCL位置平均差异+4.5°内旋,p<0.001)。

要点

  • Mayo法与Insall线均参考胫骨结节内中三分之一,变异性低且内旋误差风险低。
  • 基于远端标志(胫骨干及以远)设定旋转可能导致明显旋转对线不良。
  • 亚洲人较白人胫骨内旋更大(TCL位置平均差异+4.5°内旋,p<0.001)。

核心结论

  • Mayo法与Insall线(均参考胫骨结节内中三分之一)在准确性、低变异性及降低内旋误差风险之间提供理想平衡。
  • 基于远端标志(胫骨干及以远)设定旋转可能使术者面临明显旋转对线不良错误,尤其考虑到本研究中发现的族群间胫骨扭转差异。

研究设计与样本

  • 基于SOMA(Stryker Orthopaedic Modeling and Analytics)数据库识别1351例完整胫骨CT扫描。
  • 根据已发表指南构建多种胫骨参考轴(包括Mayo轴、Akagi线、Insall线、前髁轴[ACA]、后髁轴[PCA]、外侧胫骨皮质[LTC]、Cobb轴、胫骨嵴线[TCL]及经踝轴[TMA])。
  • 以Berger法作为参考标准。

主要发现

  • Mayo法(连接胫骨结节内中三分之一与胫骨几何中心的线)相对Berger法变异性低(7.8°±1.0°),内旋误差可能性低(0.7%)。
  • Insall线(连接后交叉韧带[PCL]止点与胫骨结节中内三分之一交点的线)相对Berger法变异性低(5.1°±2.2°),内旋误差可能性低(1.8%)。
  • 未发现临床显著的性别差异(所有均<0.7°)。
  • 族群方面除亚洲人较白人胫骨内旋一致更大外(TCL位置平均差异+4.5°内旋,p<0.001),其余结果相同。

关键数值与推荐

  • Mayo法相对Berger法变异性为7.8°±1.0°,内旋误差可能性为0.7%。
  • Insall线相对Berger法变异性为5.1°±2.2°,内旋误差可能性为1.8%。
  • 性别差异均<0.7°。
  • 亚洲人较白人TCL位置平均差异+4.5°内旋,p<0.001。

临床怎么用

  • Mayo法与Insall线(均参考胫骨结节内中三分之一)在准确性、低变异性及降低内旋误差风险之间提供理想平衡。
  • 基于远端标志(胫骨干及以远)设定旋转可能使术者面临明显旋转对线不良错误。

局限性

  • 证据等级为治疗性IV级。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Many anatomic landmarks have been described for setting tibial component rotation intraoperatively. There is no consensus as to which axis is best for reducing outliers and preventing malrotation. METHODS: The SOMA (Stryker Orthopaedic Modeling and Analytics) database (Stryker) was used to identify 1,351 computed tomography (CT) scans of the entire tibia. Several reference axes for the tibia (including the Mayo axis, Akagi line, Insall line, anterior condylar axis [ACA], posterior condylar axis [PCA], lateral tibial cortex [LTC], Cobb axis, tibial crest line [TCL], and transmalleolar axis [TMA]) were constructed according to published guidelines. The Berger method served as the reference standard. RESULTS: The Mayo method (involving a line connecting the medial and middle one-thirds of the tibial tubercle and the geometric center of the tibia) and the Insall line (involving a line connecting the posterior cruciate ligament [PCL] insertion and the intersection of the middle and medial one-thirds of the tibial tubercle) both had low variability relative to the Berger method (7.8° ± 1.0° and 5.1° ± 2.2°, respectively) and a low likelihood of internal rotation errors (0.7% and 1.8%, respectively). No clinically significant gender-based differences were found (<0.7° for all). The same was true for ethnicity, with the exception of consistently greater tibial intorsion in Asian versus Caucasian individuals (mean difference in TCL position, +4.5° intorsion for Asian individuals; p < 0.001). CONCLUSIONS: This CT-based study of 1,351 tibiae (which we believe to be the largest study of its kind) showed that the Mayo and Insall methods (both of which reference the medial and middle one-thirds of the tibial tubercle) offer an ideal balance of accuracy, low variability, and a reduced likelihood of internal rotation errors. Setting rotation on the basis of distal landmarks (tibial shaft and beyond) may predispose surgeons to substantial malrotation errors, especially given the differences in tibial torsion found between ethnic groups in this study. LEVEL OF EVIDENCE: Therapeutic Level IV . See Instructions for Authors for a complete description of levels of evidence.