定制假体重建Paprosky III型缺损:能否获得正确位置?短期影像学结果

Reconstruction of Paprosky III defects with custom-made implants: do we get them in the correct position? : short-term radiological results.
PubMed 文献 · Bone Joint J · 2022-01-01

本研究连续分析45例(2014至2019年)因Paprosky III型缺损行翻修全髋关节置换术并采用定制部分骨盆假体的患者,比较术前规划CT与术后CT及X线片的假体位置偏差。结果显示计划与实现的髋臼前倾和外展平均偏差分别为4.5°和4°,5髋(10.6%)出现假体位置不良(偏差>10°)。作者认为,在短期随访中,通过细致规划可实现高度精确的定位和解剖重建,定制部分骨盆假体为伴大量髋臼骨缺损的翻修全髋关节置换术提供了可行解决方案。

要点

  • 定制部分骨盆假体用于Paprosky III型髋臼缺损翻修全髋关节置换术,计划与实现的髋臼前倾平均偏差4.5°、外展平均偏差4°。
  • 5髋(10.6%)出现假体位置不良(偏差>10°),股骨柄保留、手术入路或固定方式对位置不良无显著影响。
  • 42例(93%)计划重建原生旋转中心,平均三维偏差向量为15.5 mm。

核心结论

  • 在短期随访中,定制部分骨盆假体为伴大量髋臼骨缺损的翻修全髋关节置换术提供了可行解决方案,通过细致规划可实现高度精确的定位和 anatomical 重建,假体安放精度有助于减少并发症,螺钉固定未损伤重要结构。

研究设计与样本

  • 连续纳入2014至2019年因Paprosky III型缺损行翻修全髋关节置换术的45例定制部分骨盆假体患者,比较术前规划CT与术后CT及X线片。
  • 平均随访30个月(标准差19;范围6至74),4例失访。

主要发现

  • 计划与实现的髋臼前倾平均CT偏差为4.5°(标准差3°;范围0°至12°),外展平均CT偏差为4°(标准差3.5°;范围1°至12°)。
  • 5髋(10.6%)出现假体位置不良(偏差>10°)。
  • 42例(93%)计划重建原生旋转中心,平均三维偏差向量为15.5 mm(标准差8.5;范围4至35)。
  • 股骨柄保留、手术入路或固定方式对假体位置不良无显著影响。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIMS: The aim of this study was to examine the implant accuracy of custom-made partial pelvis replacements (PPRs) in revision total hip arthroplasty (rTHA). Custom-made implants offer an option to achieve a reconstruction in cases with severe acetabular bone loss. By analyzing implant deviation in CT and radiograph imaging and correlating early clinical complications, we aimed to optimize the usage of custom-made implants. METHODS: A consecutive series of 45 (2014 to 2019) PPRs for Paprosky III defects at rTHA were analyzed comparing the preoperative planning CT scans used to manufacture the implants with postoperative CT scans and radiographs. The anteversion (AV), inclination (IC), deviation from the preoperatively planned implant position, and deviation of the centre of rotation (COR) were explored. Early postoperative complications were recorded, and factors for malpositioning were sought. The mean follow-up was 30 months (SD 19; 6 to 74), with four patients lost to follow-up. RESULTS: Mean CT defined discrepancy (Δ) between planned and achieved AV and IC was 4.5° (SD 3°; 0° to 12°) and 4° (SD 3.5°; 1° to 12°), respectively. Malpositioning (Δ > 10°) occurred in five hips (10.6%). Native COR reconstruction was planned in 42 cases (93%), and the mean 3D deviation vector was 15.5 mm (SD 8.5; 4 to 35). There was no significant influence in malpositioning found for femoral stem retention, surgical approach, or fixation method. CONCLUSION: At short-term follow-up, we found that PPR offers a viable solution for rTHA in cases with massive acetabular bone loss, as highly accurate positioning can be accomplished with meticulous planning, achieving anatomical reconstruction. Accuracy of achieved placement contributed to reduced complications with no injury to vital structures by screw fixation.Cite this article: Bone Joint J 2022;104-B(10):1110-1117.