哪些患者与植入物因素影响锥部腐蚀严重程度?664 根股骨柄的植入物回收分析
该植入物回收研究对 664 个髋关节置换组配式股骨柄锥部进行数字显微镜下腐蚀评分,用回归分析评估患者与器械因素与腐蚀等级的关系。结果显示植入时间更长(P = .005)和球头尺寸更大(P < .001)与腐蚀等级升高相关,而初次手术年龄较大(P = .035)、不锈钢柄材质(P = .022)、因骨或假体周围骨折翻修(P = .017)、感染(P = .018)及某些较大锥度几何形状与腐蚀等级降低相关。作者认为上述因素与多数既往较小样本回收研究一致,并建议外科医生在选择植入物及诊断有症状髋关节置换患者锥部腐蚀时了解这些危险因素。
要点
- 植入时间更长(P = .005)和球头尺寸更大(P < .001)与股骨柄锥部腐蚀等级升高相关。
- 初次手术年龄较大(P = .035)、不锈钢柄材质(P = .022)与腐蚀等级降低相关。
- 因骨或假体周围骨折翻修(P = .017)、感染(P = .018)及某些较大锥度几何形状与腐蚀等级降低相关。
核心结论
- 与更严重腐蚀风险升高或降低相关的因素与多数既往较小样本回收研究一致。
- 作者认为外科医生在选择植入物及诊断有症状髋关节置换患者锥部腐蚀时应了解这些危险因素。
研究设计与样本
- 对 664 个髋关节置换组配式股骨柄组件进行植入物回收研究,收集患者与器械信息。
- 锥部在数字显微镜下成像并用分级系统进行腐蚀损伤评分,采用回归分析将损伤与患者及器械因素关联。
主要发现
- 植入时间更长(P = .005)与腐蚀等级升高相关。
- 球头尺寸更大(P < .001)与腐蚀等级升高相关。
- 初次手术年龄较大(P = .035)与腐蚀等级降低相关。
- 不锈钢柄材质(P = .022)与腐蚀等级降低相关。
- 因骨或假体周围骨折翻修(P = .017)、感染(P = .018)及某些较大锥度几何形状与腐蚀等级降低相关。
临床怎么用
- 作者认为外科医生在选择植入物及诊断有症状髋关节置换患者锥部腐蚀时应了解上述危险因素。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Corrosion at the modular head-neck taper interface of total and hemiarthroplasty hip implants (trunnionosis) is a cause of implant failure and thus a clinical concern. Patient and device factors contributing to the occurrence of trunnionosis have been investigated in prior implant retrieval studies but generally with limited sample sizes and a narrow range of models. The purpose of the present investigation was to determine which patient and device factors were associated with corrosion damage on the femoral stem taper across a large collection of different implant models retrieved following revision hip arthroplasty. METHODS: A retrieval study of 664 hip arthroplasty modular stem components was performed. Patient and device information was collected. Trunnions were imaged under digital microscopy and scored for corrosion damage using a scaling system. Damage was related to patient and device factors using regression analyses. RESULTS: Greater duration of implantation (P = .005) and larger head size (P < .001) were associated with an elevated corrosion class. Older age at index surgery (P = .035), stainless steel stem material (P = .022), indication for revision as bone or periprosthetic fracture (P = .017), and infection (P = .018) and certain larger taper geometries were associated with a decreased corrosion class. CONCLUSION: Factors identified as contributing to a higher or lower risk of more severe corrosion are consistent with most prior smaller retrieval studies. Surgeons should be aware of these risk factors when selecting implants for their patients and when diagnosing trunnionosis in symptomatic hip arthroplasty patients.
引用
- Codirenzi AM, Lanting BA, Teeter MG, et al. What Patient and Implant Factors Affect Trunnionosis Severity? An Implant Retrieval Analysis of 664 Femoral Stems. J Arthroplasty. 2022. PMID: 36084756. doi:10.1016/j.arth.2022.08.023
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。