陶瓷股骨头相比钴铬股骨头表现出更低的磨损率:一项Meta分析
该Meta分析纳入2003至2020年间36项体内研究(钴铬组1657例、陶瓷组659例),比较第四代陶瓷与钴铬股骨头在高交联聚乙烯内衬上的线性磨损率。钴铬组加权平均磨损率为0.063 mm/年,陶瓷组为0.047 mm/年(P<.01);4项直接对比研究的Meta分析显示钴铬头磨损比陶瓷头多0.029 mm/年(95%CI 0.026-0.059,P=.306)。结论为第四代陶瓷股骨头的磨损率显著低于钴铬头。
要点
- 钴铬组加权平均磨损率为0.063 mm/年,陶瓷组为0.047 mm/年(P<.01)。
- 4项直接对比研究的Meta分析中,钴铬头磨损比陶瓷头多0.029 mm/年(95%CI 0.026-0.059,P=.306)。
- 32 mm头陶瓷磨损显著更高(P<.01),36 mm头钴铬磨损显著更高(P<.01)。
核心结论
- 第四代陶瓷股骨头的磨损率显著低于钴铬股骨头。
- 与既往研究不同,本Meta分析仅纳入体内研究且使用同一代高交联聚乙烯内衬。
研究设计与样本
- 遵循PRISMA指南的系统综述与Meta分析,检索2003至2020年间所有评估第四代陶瓷或钴铬股骨头在高交联聚乙烯内衬上体内磨损率的研究。
- 共纳入36项研究,钴铬组1657例、陶瓷组659例。
- 采用加权均值分析和随机效应Meta分析。
主要发现
- 钴铬组加权平均磨损率为0.063 mm/年(标准差0.061,95%CI 0.049-0.077),陶瓷组为0.047 mm/年(标准差0.057,95%CI 0.033-0.062;P<.01)。
- 4项直接对比陶瓷与钴铬研究的Meta分析显示,钴铬头磨损比陶瓷头多0.029 mm/年(95%CI 0.026-0.059,P=.306)。
- 32 mm头的平均磨损陶瓷显著更高(P<.01),36 mm头的平均磨损钴铬显著更高(P<.01)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Wear between the femoral head and acetabular liners continues to limit the longevity of total hip arthroplasty implants despite advances in implant materials. The purpose of this meta-analysis was to compare linear wear rates of cobalt-chromium (CoCr) and fourth-generation ceramic femoral heads on highly cross-linked polyethylene (XLPE) liners. METHODS: A systematic review following Preferred Reporting Items for Systematic Reviews and Meta-Analyses guidelines was conducted to identify all studies between 2003 and 2020 that examined in vivo wear rates of either fourth-generation ceramics or CoCr femoral heads on XLPE liners. Studies were analyzed in a weighted means analysis of wear rates and a random effects meta-analysis. RESULTS: A total of 36 studies met inclusion criteria (1,657 CoCr and 659 ceramic patients). The pooled, weighted mean wear rate was 0.063 mm/year (standard deviation [SD]: 0.061, confidence interval [CI]: 0.049-0.077) for CoCr and 0.047 mm/year (SD: 0.057, CI: 0.033-0.062; P < .01) for ceramic (P < .01). A meta-analysis of 4 studies directly comparing ceramic and CoCr found that CoCr heads demonstrated 0.029 mm/year more wear than ceramic heads (95% CI: 0.026-0.059, P = .306). Mean wear for 32-mm heads was significantly higher for ceramic (P < .01), while mean wear for 36-mm heads was significantly higher for CoCr (P < .01). CONCLUSION: Fourth-generation ceramic femoral heads were found to have significantly lower wear rates than CoCr heads. Unlike previous studies, this meta-analysis included only in vivo studies and those with the same generation of highly XLPE liners.