65岁及以上患者全膝关节置换术中聚乙烯轴承设计的翻修风险:来自美国关节置换登记处的分析
基于美国关节置换登记处(AJRR)2012至2019年数据,分析305,279例初次全膝关节置换术,比较后稳定型与最小稳定型聚乙烯轴承的翻修风险。后稳定型轴承的全因翻修风险(HR 1.25,95%CI 1.2至1.3,p<0.0001)和感染翻修风险(HR 1.18,95%CI 1.0至1.4,p=0.02)均高于最小稳定型轴承。作者认为外科医生在选择初次全膝关节置换术轴承时应考虑这一发现。
要点
- 后稳定型轴承全因翻修HR 1.25(95%CI 1.2至1.3,p<0.0001),高于最小稳定型轴承。
- 后稳定型轴承感染翻修HR 1.18(95%CI 1.0至1.4,p=0.02),高于最小稳定型轴承。
- 最小稳定型轴承全因翻修1,693例(1.18%),后稳定型2,406例(1.49%)。
核心结论
- 与美国以外国际登记处的研究相似,在美国全膝关节置换术中使用后稳定型轴承与全因翻修和感染翻修风险增加相关。
- 作者认为外科医生在选择初次全膝关节置换术轴承时应考虑这一发现。
研究设计与样本
- 对美国关节置换登记处(AJRR)2012至2019年数据的分析,识别研究期间所有初次全膝关节置换术,并在有补充数据时链接至美国医疗保险和医疗补助服务中心(CMS)数据。
- 共识别305,279例报告最小稳定型或后稳定型假体特征的病例,其中后稳定型轴承161,486例(52.9%),最小稳定型轴承143,793例(47.1%)。
主要发现
- 最小稳定型轴承中全因翻修1,693例(1.18%),感染翻修334例(0.23%)。
- 后稳定型轴承中全因翻修2,406例(1.49%),感染翻修446例(0.28%)。
- 后稳定型轴承全因翻修HR 1.25(95%CI 1.2至1.3,p<0.0001),感染翻修HR 1.18(95%CI 1.0至1.4,p=0.02)。
关键数值与推荐
- 最小稳定型假体包括交叉韧带保留型、前稳定型或旋转平台设计;后稳定型假体为对照。
- 采用累积发生率函数(CIF)曲线和Cox比例风险比(HR)分析全因翻修和感染翻修。
临床怎么用
- 作者认为外科医生在选择初次全膝关节置换术轴承时应考虑这一发现。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Polyethylene bearing design may influence the risk of revision after total knee arthroplasty (TKA). Previous investigations from outside of the United States have suggested the potential for an increased risk of all-cause revision and revision for infection with the use of posterior-stabilized bearings. We examine the risk of revision based on the polyethylene bearing design selection for primary TKA. METHODS: An analysis of the American Joint Replacement Registry (AJRR) data from 2012 to 2019 was performed. We identified all primary TKAs over the study period and linked cases to supplemental U.S. Centers for Medicare & Medicaid Services (CMS) data where available. Patient demographic characteristics, polyethylene characteristics, procedure dates, and cause for revision were recorded. Analysis was performed to compare minimally stabilized implants (cruciate-retaining, anterior-stabilized, or pivot bearing designs) with posterior-stabilized implants. Cumulative incidence function (CIF) curves and Cox proportional hazard ratios (HRs) were created for all-cause revision and revision for infection in each group. RESULTS: We identified 305,279 cases with reported minimally stabilized or posterior-stabilized implant characteristics. There were 161,486 cases (52.9%) that utilized posterior-stabilized bearings compared with 143,793 cases (47.1%) with minimally stabilized bearings. For minimally stabilized bearings, 1,693 cases (1.18%) had all-cause revision and 334 cases (0.23%) had revision for infection. For posterior-stabilized bearings, 2,406 cases (1.49%) had all-cause revision and 446 cases (0.28%) had revision for infection. The use of posterior-stabilized bearings had HRs of 1.25 (95% confidence interval [CI], 1.2 to 1.3; p < 0.0001) for all-cause revision and 1.18 (95% CI, 1.0 to 1.4; p = 0.02) for revision for infection. CONCLUSIONS: Similar to investigations from international registries, we found an increased risk of all-cause revision and revision for infection when using posterior-stabilized bearings in TKA in the United States. Surgeons should consider this finding when considering bearing selection for primary TKA. LEVEL OF EVIDENCE: Therapeutic Level III . See Instructions for Authors for a complete description of levels of evidence.